Metal Artifact Reduction in Pelvic Computed Tomography With Hip Prostheses: Comparison of Virtual Monoenergetic Extrapolations From Dual-Energy Computed Tomography and an Iterative Metal Artifact Reduction Algorithm in a Phantom Study.

Kai Higashigaito1, Florian Angst, Val M Runge, Hatem Alkadhi, Olivio F Donati.   

Abstract

OBJECTIVE: The aim of this study was to directly compare metal artifact reduction (MAR) of virtual monoenergetic extrapolations (VMEs) from dual-energy computed tomography (CT) with iterative MAR (iMAR) from single energy in pelvic CT with hip prostheses.
MATERIALS AND METHODS: A human pelvis phantom with unilateral or bilateral metal inserts of different material (steel and titanium) was scanned with third-generation dual-source CT using single (120 kVp) and dual-energy (100/150 kVp) at similar radiation dose (CT dose index, 7.15 mGy). Three image series for each phantom configuration were reconstructed: uncorrected, VME, and iMAR. Two independent, blinded radiologists assessed image quality quantitatively (noise and attenuation) and subjectively (5-point Likert scale). Intraclass correlation coefficients (ICCs) and Cohen κ were calculated to evaluate interreader agreements. Repeated measures analysis of variance and Friedman test were used to compare quantitative and qualitative image quality. Post hoc testing was performed using a corrected (Bonferroni) P < 0.017.
RESULTS: Agreements between readers were high for noise (all, ICC ≥ 0.975) and attenuation (all, ICC ≥ 0.986); agreements for qualitative assessment were good to perfect (all, κ ≥ 0.678). Compared with uncorrected images, VME showed significant noise reduction in the phantom with titanium only (P < 0.017), and iMAR showed significantly lower noise in all regions and phantom configurations (all, P < 0.017). In all phantom configurations, deviations of attenuation were smallest in images reconstructed with iMAR. For VME, there was a tendency toward higher subjective image quality in phantoms with titanium compared with uncorrected images, however, without reaching statistical significance (P > 0.017). Subjective image quality was rated significantly higher for images reconstructed with iMAR than for uncorrected images in all phantom configurations (all, P < 0.017).
CONCLUSIONS: Iterative MAR showed better MAR capabilities than VME in settings with bilateral hip prosthesis or unilateral steel prosthesis. In settings with unilateral hip prosthesis made of titanium, VME and iMAR performed similarly well.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 26171565     DOI: 10.1097/RLI.0000000000000191

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  19 in total

Review 1.  Dual energy computed tomography virtual monoenergetic imaging: technique and clinical applications.

Authors:  Tommaso D'Angelo; Giuseppe Cicero; Silvio Mazziotti; Giorgio Ascenti; Moritz H Albrecht; Simon S Martin; Ahmed E Othman; Thomas J Vogl; Julian L Wichmann
Journal:  Br J Radiol       Date:  2019-04-09       Impact factor: 3.039

2.  Follow-up CT and CT angiography after intracranial aneurysm clipping and coiling-improved image quality by iterative metal artifact reduction.

Authors:  Georg Bier; Malte Niklas Bongers; Johann-Martin Hempel; Anja Örgel; Till-Karsten Hauser; Ulrike Ernemann; Florian Hennersdorf
Journal:  Neuroradiology       Date:  2017-06-03       Impact factor: 2.804

3.  Iterative algorithms for metal artifact reduction in children with orthopedic prostheses: preliminary results.

Authors:  Seema Toso; Meryle Laurent; Elise Dupuis Lozeron; Pauline Brindel; Marirosa Cristallo Lacalamita; Sylviane Hanquinet
Journal:  Pediatr Radiol       Date:  2018-07-28

4.  Automatic tube current modulation for whole-body polytrauma CT with immobilization devices: is there an increase in radiation dose and degradation of image quality?

Authors:  André Euler; Bram Stieltjes; Sebastian T Schindera
Journal:  Emerg Radiol       Date:  2016-09-10

5.  Combining monoenergetic extrapolations from dual-energy CT with iterative reconstructions: reduction of coil and clip artifacts from intracranial aneurysm therapy.

Authors:  Sebastian Winklhofer; Ricarda Hinzpeter; Daniel Stocker; Gerasimos Baltsavias; Lars Michels; Jan-Karl Burkhardt; Luca Regli; Antonios Valavanis; Hatem Alkadhi
Journal:  Neuroradiology       Date:  2018-01-22       Impact factor: 2.804

6.  Clinical evaluation of the iterative metal artefact reduction algorithm for post-operative CT examination after maxillofacial surgery.

Authors:  Arsany Hakim; Johannes Slotboom; Olivier Lieger; Fabian Schlittler; Roland Giger; Chantal Michel; Roland Wiest; Franca Wagner
Journal:  Dentomaxillofac Radiol       Date:  2017-03-07       Impact factor: 2.419

7.  Metal implants on abdominal CT: does split-filter dual-energy CT provide additional value over iterative metal artifact reduction?

Authors:  Hildegard M Wichtmann; Kai R Laukamp; Sebastian Manneck; Konrad Appelt; Bram Stieltjes; Daniel T Boll; Matthias R Benz; Markus M Obmann
Journal:  Abdom Radiol (NY)       Date:  2022-09-30

8.  Value of CT iterative metal artifact reduction in PET/CT-clinical evaluation in 100 patients.

Authors:  Christian Philipp Reinert; Christian la Fougère; Konstantin Nikolaou; Christina Pfannenberg; Sergios Gatidis
Journal:  Br J Radiol       Date:  2019-01-14       Impact factor: 3.039

9.  CT artifacts after contrast media injection in chest imaging: evaluation of post-processing algorithms, virtual monoenergetic images and their combination for artifact reduction.

Authors:  Amit Gupta; Verena Carola Obmann; Michelle Jordan; Simon Lennartz; Markus Michael Obmann; Nils Große Hokamp; David Zopfs; Lenhard Pennig; Gina Fürtjes; Nikhil Ramaiya; Robert Gilkeson; Kai Roman Laukamp
Journal:  Quant Imaging Med Surg       Date:  2021-01

10.  Utility of an automatic adaptive iterative metal artifact reduction AiMAR algorithm in improving CT imaging of patients with hip prostheses evaluated for suspected bladder malignancy.

Authors:  Payam Mohammadinejad; Ashish Khandelwal; Akitoshi Inoue; Hiroaki Takahashi; Mariana Yalon; Zaiyang Long; Ahmed F Halaweish; Shuai Leng; Lifeng Yu; Yong S Lee; Cynthia H McCollough; Joel G Fletcher
Journal:  Abdom Radiol (NY)       Date:  2022-03-23
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.