| Literature DB >> 26167491 |
Lin Wang1, Stanley Sai-Chuen Hui2.
Abstract
The aim of the study is to examine the validity in predicting body fat percentage (%BF) of different bioelectrical impedance (BIA) devices among Chinese children and adolescents. A total of 255 Chinese children and adolescents aged 9-19 years old participated in the study. %BF was assessed by BIA scales, namely, Biodynamics-310 (Model A), Tanita TBF-543 (Model B), Tanita BC-545 (Model C), and InBody 520 (Model D). Dual-energy X-ray absorptiometry (DXA) was used as the criterion measurement. Lin's concordance correlation coefficients of estimated %BF between Model A, Model B, Model C, and DXA showed poor agreements for both genders. Moderate agreements for %BF were found between DXA and Model D measurements. In boys, differences in %BF were found between DXA and Model B and Model C. No significant %BF differences were found between Model A, Model D, and DXA. However, the two BIA analyzers showed a significant positive correlation between the bias and average %BF between BIA and DXA. In girls, differences in %BF were observed between Model B, Model C, Model D, and DXA. Model A and DXA showed no significant differences of %BF; however, the bias and the average %BF between the BIA and DXA had a significant positive correlation. Using embedded equations in BIA devices should be validated in assessing the %BF of Chinese children and adolescents.Entities:
Mesh:
Year: 2015 PMID: 26167491 PMCID: PMC4475745 DOI: 10.1155/2015/614858
Source DB: PubMed Journal: Biomed Res Int Impact factor: 3.411
Physical characteristics and body compositions of the study population.
| Boys ( | Girls ( | |||||
|---|---|---|---|---|---|---|
| Mean (SD) | Range | Mean (SD) | Range | |||
| Lower | Upper | Lower | Upper | |||
| Age (years) | 13.8 (2.9) | 9.0 | 19.5 | 13.7 (2.8) | 9.0 | 19.4 |
| Height (cm) | 161.8 (14.0) | 130.5 | 186.0 | 154.3 (9.8) | 126.6 | 172.0 |
| Weight (kg) | 55.9 (17.0) | 27.5 | 108.1 | 47.4 (11.3) | 23.5 | 97.1 |
| BMI (kg2/m) | 21.0 (4.2) | 13.6 | 37.1 | 19.7 (3.2) | 13.3 | 34.8 |
| LTM-DXA (kg) | 41.0 (11.2) | 21.0 | 65.3 | 31.0 (5.5) | 18.7 | 54.3 |
| BMC-DXA (kg) | 2.0 (0.7) | 0.9 | 3.8 | 1.7 (0.5) | 0.8 | 2.7 |
| FM-DXA (kg) | 12.7 (9.5) | 3.3 | 53.1 | 14.3 (6.8) | 2.7 | 40.6 |
| FFM-DXA (kg) | 43.0 (11.8) | 22.1 | 69.0 | 32.8 (5.9) | 19.5 | 56.8 |
| FFM-Model A (kg) | 43.8 (12.7) | 21.8 | 77.4 | 33.9 (6.5) | 18.6 | 61.2 |
| FFM-Model B (kg) | 43.2 (11.7) | 20.8 | 73.2 | 34.7 (5.5) | 20.6 | 51.7 |
| FFM-Model C (kg) | 44.5 (12.0) | 22.6 | 73.2 | 34.5 (5.5) | 20.6 | 51.6 |
| FFM-Model D (kg) | 44.3 (12.2) | 22.8 | 72.0 | 34.0 (5.9) | 20.0 | 55.3 |
| %BF-DXA (%) | 20.8 (9.9) | 6.1 | 49.1 | 27.9 (7.7) | 11.8 | 46.5 |
| %BF-Model A (%) | 21.4 (7.0) | 4.9 | 39.6 | 27.4 (6.0) | 13.6 | 42.1 |
| %BF-Model B (%) | 21.9 (6.9) | 10.1 | 45.3 | 25.3 (7.8) | 9.3 | 49.4 |
| %BF-Model C (%) | 19.0 (9.0) | 5.0 | 54.7 | 25.8 (7.7) | 10.3 | 47.6 |
| %BF-Model D (%) | 20.2 (8.5) | 6.2 | 47.3 | 26.3 (8.1) | 6.4 | 44.8 |
Note: data are presented as mean (SD); BMI: body mass index; LTM-DXA: lean tissue mass measured by dual-energy X-ray absorptiometry; BMC-DXA: bone mineral content measured by dual-energy X-ray absorptiometry; FM-DXA: fat mass measured by dual-energy X-ray absorptiometry; %BF-DXA: body fat percentage measured by dual-energy X-ray absorptiometry; FFM-DXA: fat-free mass calculated as the sum of LTM and BMC; Model A: Biodynamics-310; Model B: Tanita TBF 543; Model C: Tanita BC-545; Model D: InBody 520.
Lin's concordance correlation coefficients between parameters assessed by different BIA and DXA measurement.
| BIA devices | |||||
|---|---|---|---|---|---|
| Model A | Model B | Model C | Model D | ||
| Boys ( | %BF-DXA | 0.765 | 0.750 | 0.855 | 0.926 |
| FFM-DXA | 0.953 | 0.953 | 0.951 | 0.979 | |
|
| |||||
| Girls ( | %BF-DXA | 0.747 | 0.863 | 0.881 | 0.912 |
| FFM-DXA | 0.909 | 0.897 | 0.910 | 0.957 | |
Note: BMI: body mass index; %BF-DXA: body fat percentage measured by dual-energy X-ray absorptiometry; FFM-DXA: fat-free mass measured by dual-energy X-ray absorptiometry.
Model A: Biodynamics-310; Model B: Tanita TBF 543; Model C: Tanita BC-545; Model D: InBody 520.
Cross-validation of four BIA devices (embedded equations) for the perdition of %BF/FFM in Chinese children and adolescents.
| Gender | BIA devices | %BF-DXA | %BF-BIA | FFM-DXA | FFM-BIA |
| Bias (SD) | TE |
| 95% |
|---|---|---|---|---|---|---|---|---|---|---|
|
Boys ( | Model A | 20.77% | 21.34% | — | — | 0.268 | 0.58 (5.86)% | 5.86% | 0.513 | −10.96% to 12.12% |
| Model B | 20.77% | 21.92% | — | — | 0.031 | 1.15 (5.95)% | 6.04% | 0.528 | −10.51% to 12.81% | |
| Model C | 20.77% | 19.04% | — | — | <0.001 | −1.67 (4.80)% | 5.06% | 0.162 | −11.08% to 7.74% | |
| Model D | 20.77% | 20.22% | — | — | 0.246 | −0.37 (3.51)% | 3.51% | 0.365 | −7.25% to 6.51% | |
| Model A | — | — | 43.02 kg | 43.80 kg | 0.019 | 0.78 (3.70) kg | 3.77 kg | 0.261 | −6.47 kg to 8.03 kg | |
| Model B | — | — | 43.02 kg | 43.18 kg | 0.608 | 0.16 (3.60) kg | 3.60 kg | −0.027 | −6.90 kg to 7.22 kg | |
| Model C | — | — | 43.02 kg | 44.50 kg | <0.001 | 1.54 (3.35) kg | 3.60 kg | 0.024 | −5.03 kg to 8.11 kg | |
| Model D | — | — | 43.02 kg | 44.31 kg | <0.001 | 1.02 (2.17) kg | 2.39 kg | 0.197 | −3.23 kg to 5.27 kg | |
|
| ||||||||||
|
Girls ( | Model A | 27.85% | 27.37% | — | — | 0.266 | −0.48 (4.88)% | 4.88% | 0.359 | −10.04% to 9.08% |
| Model B | 27.85% | 25.22% | — | — | <0.001 | −2.63 (3.23)% | 4.15% | −0.049 | −8.96% to 3.70% | |
| Model C | 27.85% | 25.81% | — | — | <0.001 | −2.05 (3.21)% | 3.79% | 0.012 | −8.34% to 4.24% | |
| Model D | 27.85% | 26.34% | — | — | <0.001 | −1.47 (2.99)% | 3.31% | −0.147 | −7.33% to 4.39% | |
| Model A | — | — | 32.77 kg | 33.89 kg | <0.001 | 1.11 (2.42) kg | 2.66 kg | 0.245 | −3.63 kg to 5.85 kg | |
| Model B | — | — | 32.77 kg | 34.74 kg | <0.001 | 1.97 (1.77) kg | 2.64 kg | −0.255 | −1.50 kg to 5.44 kg | |
| Model C | — | — | 32.77 kg | 34.49 kg | <0.001 | 1.72 (1.77) kg | 2.47 kg | −0.219 | −1.75 kg to 5.19 kg | |
| Model D | — | — | 32.77 kg | 34.03 kg | <0.001 | 1.25 (1.22) kg | 1.68 kg | −0.022 | −1.14 kg to 3.64 kg | |
Note: %BF: body fat percentage; FFM: fat-free mass; DXA: dual-energy X-ray absorptiometry; BIA: bioelectrical impedance analysis.
Bias: value obtained from BIA measurement minus DXA measurement; SD: standard deviation.
r , Pearson's correlation coefficient for the relationship between averaged %BF/FFM [(value of DXA + value of BIA)/2] and the bias.
p < 0.05; p < 0.01.
Model A: Biodynamics-310; Model B: Tanita TBF 543; Model C: Tanita BC-545; Model D: Inbody 520.