| Literature DB >> 26167136 |
J R Anderson1, J J H Ackerman2, J R Garbow3.
Abstract
Two semipermeable, hollow fiber phantoms for the validation of perfusion-sensitive magnetic resonance methods and signal models are described. Semipermeable hollow fibers harvested from a standard commercial hemodialysis cartridge serve to mimic tissue capillary function. Flow of aqueous media through the fiber lumen is achieved with a laboratory-grade peristaltic pump. Diffusion of water and solute species (e.g., Gd-based contrast agent) occurs across the fiber wall, allowing exchange between the lumen and the extralumenal space. Phantom design attributes include: i) small physical size, ii) easy and low-cost construction, iii) definable compartment volumes, and iv) experimental control over media content and flow rate.Entities:
Keywords: MR phantom design; dialyzer; hollow fiber bioreactor; perfusion; semipermeable hollow fiber
Year: 2011 PMID: 26167136 PMCID: PMC4497530 DOI: 10.1002/cmr.b.20202
Source DB: PubMed Journal: Concepts Magn Reson Part B Magn Reson Eng ISSN: 1552-5031 Impact factor: 1.176