| Literature DB >> 12968572 |
Churn K Poh1, Peter A Hardy, Zhijie Liao, Zhongping Huang, William R Clark, Dayong Gao.
Abstract
We used an innovative, nonintrusive MRI technique called the two-dimensional (2D) Phase-Contrast (2DPC) velocity-imaging technique to investigate the effect of flow baffles on the dialysate-side flow distribution in two different hollow-fiber hemodialyzers (A and B); each with flow rates between 200 and 1000 mL/min (3.33 x 10(-6) and 1.67 x 10(-5) m3/s). Our experimental results show that (1) the dialysate-side flow distribution was nonuniform with channeling flow occurred at the peripheral cross section of these hollow-fiber hemodialyzers, and (2) the existing designs of flow baffles failed to promote uniform dialysate-side flow distribution for all flow rates studies.Mesh:
Substances:
Year: 2003 PMID: 12968572 DOI: 10.1115/1.1590355
Source DB: PubMed Journal: J Biomech Eng ISSN: 0148-0731 Impact factor: 2.097