| Literature DB >> 26149115 |
Boone M Prentice1,2, Chad W Chumbley3,2, Richard M Caprioli1,3,4,2.
Abstract
A matrix-assisted laser desorption/ionization time of flight/time of flight tandem mass spectrometer (MALDI TOF/TOF) has been used for high-speed precursor/fragment ion transition image acquisition. High-throughput analysis is facilitated by an Nd:YLF solid state laser capable of pulse repetition rates up to 5 kHz, a high digitizer acquisition rate (up to 50 pixels/s), and continuous laser raster sampling. MS/MS experiments are enabled through the use of a precision timed ion selector, second source acceleration, and a dedicated collision cell. Continuous raster sampling is shown here to facilitate rapid MS/MS ion image acquisition from thin tissue sections for the drug rifampicin and for a common kidney lipid, SM4s(d18:1/24:1). The ability to confirm the structural identity of an analyte as part of the MS/MS imaging experiment is an essential part of the analysis. Additionally, the increase in sensitivity and specificity afforded by an MS/MS approach is highly advantageous, especially when interrogating complex chemical environments such as those in biological tissues. Herein, we report continuous laser raster sampling TOF/TOF imaging methodologies which demonstrate 8 to 14-fold increases in throughput compared with existing MS/MS instrumentation, an important advantage when imaging large areas on tissues.Entities:
Keywords: MALDI; MS/MS; TOF/TOF; continuous raster; drug imaging; imaging; lipid imaging
Mesh:
Substances:
Year: 2015 PMID: 26149115 PMCID: PMC4498415 DOI: 10.1002/jms.3579
Source DB: PubMed Journal: J Mass Spectrom ISSN: 1076-5174 Impact factor: 1.982