| Literature DB >> 20635080 |
Paul J Trim1, Marie-Claude Djidja, Sally J Atkinson, Keith Oakes, Laura M Cole, David M G Anderson, Philippa J Hart, Simona Francese, Malcolm R Clench.
Abstract
A commercial hybrid quadrupole time-of-flight mass spectrometer has been modified for high-speed matrix-assisted laser desorption ionisation (MALDI) imaging using a short-pulse optical technology Nd:YVO(4) laser. The laser operating in frequency-tripled mode (lambda = 355 nm) is capable of delivering 1.5-ns pulses of energy at up to 8 microJ at 5-10 kHz and 3 microJ at 20 kHz. Experiments to improve beam homogeneity and reduce laser speckle by mechanical vibration of the fibre-optic laser delivery system are reported along with data from trial and tissue imaging experiments using the modified instrument. The laser appeared to yield best results for MALDI-MS imaging experiments when operating at repetition rates 5-10 kHz. Combining this with raster imaging allowed images of rat brain sections to be recorded in 37 min. Similarly, images of the distribution of peptides in "on-tissue" digest experiments from tumour tissues were recorded in 1 h and 30 min rather than the 8-h acquisition time previously used. A brief investigation of targeted protein analysis/imaging by multiple reaction monitoring experiments "on-tissue" is reported. A total of 26 transitions were recorded over a 3-s cycle time and images of abundant proteins were successfully recorded.Entities:
Mesh:
Substances:
Year: 2010 PMID: 20635080 DOI: 10.1007/s00216-010-3874-6
Source DB: PubMed Journal: Anal Bioanal Chem ISSN: 1618-2642 Impact factor: 4.142