Literature DB >> 26130341

Five-year comparison of wear using oxidised zirconium and cobalt-chrome femoral heads in total hip arthroplasty: a multicentre randomised controlled trial.

S S Jassim1, S Patel2, N Wardle3, J Tahmassebi4, R Middleton5, D L Shardlow6, A Stephen7, J Hutchinson7, F S Haddad2.   

Abstract

Oxidised zirconium (OxZi) has been developed as an alternative bearing surface for femoral heads in total hip arthroplasty (THA). This study has investigated polyethylene wear, functional outcomes and complications, comparing OxZi and cobalt-chrome (CoCr) as part of a three-arm, multicentre randomised controlled trial. Patients undergoing THA from four institutions were prospectively randomised into three groups. Group A received a CoCr femoral head and highly cross-linked polyethylene (XLPE) liner; Group B received an OxZi femoral head and XLPE liner; Group C received an OxZi femoral head and ultra-high molecular weight polyethylene (UHMWPE) liner. At five years, 368 patients had no statistically significant differences in short-form-36 (p = 0.176 mental, p = 0.756 physical), Western Ontario and McMaster Universities Osteoarthritis Index (p = 0.847), pain scores (p = 0.458) or complications. The mean rate of linear wear was 0.028 mm/year (standard deviation (SD) 0.010) for Group A, 0.023 mm/year (SD 0.010) for Group B, and 0.09 mm/year (SD 0.045) for Group C. Penetration was significantly higher in the UHMWPE liner group compared with both XLPE liner groups (p < 0.001) but no significant difference was noted between CoCr and OxZi when articulating with XLPE (p = 0.153). In this, the largest randomised study of this bearing surface, it appears that using a XLPE acetabular liner is more important in reducing THA component wear than the choice of femoral head bearing, at mid-term follow-up. There is a non-significant trend towards lower wear, coupling OxZi rather than CoCr with XLPE but long-term analysis is required to see if this observation changes with time and becomes significant. ©2015 The British Editorial Society of Bone & Joint Surgery.

Entities:  

Keywords:  Hip arthroplasty alternative bearing surfaces; highly cross-linked polyethyelene; oxidized zirconium; total hip arthroplasty; ultra-high molecular weight polyethylene

Mesh:

Substances:

Year:  2015        PMID: 26130341     DOI: 10.1302/0301-620X.97B7.35285

Source DB:  PubMed          Journal:  Bone Joint J        ISSN: 2049-4394            Impact factor:   5.082


  11 in total

1.  CORR Insights(®): oxidized zirconium head on crosslinked polyethylene liner in total hip arthroplasty: a 7- to 12-year in vivo comparative wear study.

Authors:  Alexander Jaime Grübl
Journal:  Clin Orthop Relat Res       Date:  2015-09-10       Impact factor: 4.176

2.  Wear performance of cobalt chromium, ceramic, and oxidized zirconium on highly crosslinked polyethylene at mid-term follow-up.

Authors:  Matthew G Teeter; Colin J MacLean; Lyndsay E Somerville; James L Howard; Richard W McCalden; Brent A Lanting; Edward M Vasarhelyi
Journal:  J Orthop       Date:  2018-05-07

Review 3.  New alternate bearing surfaces in total hip arthroplasty: A review of the current literature.

Authors:  Preston W Grieco; Scott Pascal; Jared M Newman; Neil V Shah; Sarah G Stroud; Neil P Sheth; Aditya V Maheshwari
Journal:  J Clin Orthop Trauma       Date:  2017-10-27

4.  Editor's Spotlight/Take 5: Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.

Authors:  Seth S Leopold
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

5.  CORR Insights®: Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.

Authors:  Matthew G Teeter
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

6.  Polyethylene Wear Increases in Liners Articulating With Scratched Oxidized Zirconium Femoral Heads.

Authors:  Alberto Carli; Chelsea N Koch; Christina I Esposito; Timothy M Wright; Douglas E Padgett
Journal:  Clin Orthop Relat Res       Date:  2018-02       Impact factor: 4.176

7.  Validation of a new 2-D technique for radiographic wear measurement of cemented, highly cross-linked polyethylene acetabular cups.

Authors:  Brian Derbyshire; Bilal Barkatali
Journal:  Med Eng Phys       Date:  2017-07-03       Impact factor: 2.242

8.  Surgical trial design - learning curve and surgeon volume: Determining whether inferior results are due to the procedure itself, or delivery of the procedure by the surgeon.

Authors:  A H R W Simpson; C R Howie; J Norrie
Journal:  Bone Joint Res       Date:  2017-04       Impact factor: 5.853

9.  The effect of bearing type on the outcome of total hip arthroplasty.

Authors:  Rinne M Peters; Liza N Van Steenbergen; Martin Stevens; Paul C Rijk; Sjoerd K Bulstra; Wierd P Zijlstra
Journal:  Acta Orthop       Date:  2017-11-21       Impact factor: 3.717

10.  Recent update on crosslinked polyethylene in total hip arthroplasty.

Authors:  Jean Langlois; Moussa Hamadouche
Journal:  SICOT J       Date:  2020-05-15
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.