Literature DB >> 26122108

Midterm Results of Delta Ceramic-on-Ceramic Total Hip Arthroplasty.

William G Hamilton1, James P McAuley2, Thomas J Blumenfeld3, James P Lesko4, Sam E Himden4, Douglas A Dennis5.   

Abstract

This study reports mid-term results of Delta ceramic on ceramic (COC) in total hip arthroplasty (THA). Subjects received Delta COC THA in a prospective multi-center study with either 28 mm (n=177) or 36 mm (n=168) articulations. Annual clinical and radiographic evaluations were performed, and patients were asked about hip noises. At mean 5.3-year follow-up there were 3 (0.9%) post-operative liner fractures. Nine revisions were performed (2 liner fracture, 4 stem loosening, 3 deep infection). Kaplan-Meier survivorship at 6 years was 96.9% (94.0-98.4). Twenty-six (7.5%) subjects reported squeaking, of whom none were revised. One (0.3%) subject could reproduce a sound in clinic. More patients reported squeaking with a 36 mm bearing (28 mm: 7/177, 36 mm: 19/168, P=0.013).
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  bearing surfaces; ceramic; outcome; total hip arthroplasty; total hip replacement

Mesh:

Year:  2015        PMID: 26122108     DOI: 10.1016/j.arth.2015.02.047

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  19 in total

1.  Hard-on-Hard Bearings Are Associated With Increased Noise Generation in Young Patients Undergoing Hip Arthroplasty.

Authors:  Denis Nam; Toby Barrack; Staci R Johnson; Ryan M Nunley; Robert L Barrack
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

2.  Design Considerations for the Next Generation Hip Resurfacing Implant: Commentary.

Authors:  Edwin P Su
Journal:  HSS J       Date:  2016-12-19

3.  Third generation delta ceramic-on-ceramic bearing for total hip arthroplasty at mid-term follow-up.

Authors:  Alessandro Bistolfi; Riccardo Ferracini; Alessandro Aprato; Alessandro Massè; Walter Daghino; Sara Lea; Stefano Artiaco; Gwo-Chin Lee
Journal:  J Orthop       Date:  2020-08-28

4.  Third and fourth generation ceramic-on-ceramic total hip arthroplasty at a minimum of ten years.

Authors:  Cathal J McCarthy; John Mahon; Gerard A Sheridan; Adanna Welch-Phillips; John M O'Byrne; Paddy J Kenny
Journal:  J Clin Orthop Trauma       Date:  2022-07-06

5.  Сomparative analysis of large diameter bearing ceramic monoblock acetabular components.

Authors:  Valery Yu Murylev; Pavel M Elizarov; Aleksei V Muzychenkov; Yaroslav A Rukin; Grigory A Kukovenko; Semen S Alekseev; Michail P Elizarov
Journal:  J Orthop       Date:  2022-08-11

6.  A case-driven hypothesis for multi-stage crack growth mechanism in fourth-generation ceramic head fracture.

Authors:  Stefano Lucchini; Massimiliano Baleani; Federico Giardina; Andrea Martelli; Francesco Castagnini; Barbara Bordini; Francesco Traina
Journal:  J Orthop Surg Res       Date:  2022-06-03       Impact factor: 2.677

7.  Do the Reasons for Ceramic-on-ceramic Revisions Differ From Other Bearings in Total Hip Arthroplasty?

Authors:  Henri Migaud; Sophie Putman; Grégory Kern; Ronald Isida; Julien Girard; Nassima Ramdane; Christian P Delaunay; Moussa Hamadouche
Journal:  Clin Orthop Relat Res       Date:  2016-10       Impact factor: 4.176

8.  Hydroxyapatite reinforced Ti6Al4V composites for load-bearing implants.

Authors:  Jose D Avila; Kevin Stenberg; Susmita Bose; Amit Bandyopadhyay
Journal:  Acta Biomater       Date:  2021-01-12       Impact factor: 8.947

9.  Zirconia phase transformation in retrieved, wear simulated, and artificially aged ceramic femoral heads.

Authors:  Maria Parkes; Kathryn Sayer; Markus Goldhofer; Philippa Cann; William L Walter; Jonathan Jeffers
Journal:  J Orthop Res       Date:  2017-06-20       Impact factor: 3.494

10.  Increased migration of uncemented acetabular cups in female total hip arthroplasty patients with low systemic bone mineral density. A 2-year RSA and 8-year radiographic follow-up study of 34 patients.

Authors:  Sami Finnilä; Niko Moritz; Erkki SvedströM; Jessica J Alm; Hannu T Aro
Journal:  Acta Orthop       Date:  2015-11-16       Impact factor: 3.717

View more