| Literature DB >> 26103498 |
Joseph M Dewhurst1, Matthew Lowe, Mark J Hardy, Christopher J Boylan, Philip Whitehurst, Carl G Rowbottom.
Abstract
A semiautomated system for radiotherapy treatment plan quality control (QC), named AutoLock, is presented. AutoLock is designed to augment treatment plan QC by automatically checking aspects of treatment plans that are well suited to computational evaluation, whilst summarizing more subjective aspects in the form of a checklist. The treatment plan must pass all automated checks and all checklist items must be acknowledged by the planner as correct before the plan is finalized. Thus AutoLock uniquely integrates automated treatment plan QC, an electronic checklist, and plan finalization. In addition to reducing the potential for the propagation of errors, the integration of AutoLock into the plan finalization workflow has improved efficiency at our center. Detailed audit data are presented, demonstrating that the treatment plan QC rejection rate fell by around a third following the clinical introduction of AutoLock.Entities:
Mesh:
Year: 2015 PMID: 26103498 PMCID: PMC5690112 DOI: 10.1120/jacmp.v16i3.5396
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1Simplified treatment planning workflow at our center and the role of AutoLock within it.
Figure 2AutoLock Review module workflow.
Figure 3AutoLock LockPlan module workflow.
AutoLock check result taxonomy and meanings.
|
|
|
|---|---|
| Pass | The check has passed |
| Fail | A problem has been detected with the treatment plan |
| For review | No specific problems have been detected but the result must be reviewed and acknowledged as correct by the user |
| Not checked | The check is not required for the current treatment plan |
| Error | A programming error has occurred |
Figure 4Screenshot of the “For review” tab in the AutoLock graphical user interface; each checklist item must be acknowledged by the user by completing a tick box.
Figure 5Extract from the Plan Lock Sheet, which shows all results of the plan review, split by result type.
AutoLock general plan checks.
|
|
|
|
|---|---|---|
| 1.1 | Single trial named final | Plan has only one trial and is labeled correctly |
| 1.2 | Couch removed | Couch is removed (does not check couch removal position) |
| 1.3 | Dose grid voxel size 0.3 cm or less | Dose grid voxel size is set appropriately |
| 1.4 | Air threshold is | Patient–air threshold is set appropriately |
| 1.5 | Origin and isocenter POI check | Origin POI matches localizer, origin and isocenter POIs are correctly formatted |
| 1.6 | Overrides check | Overrides are set appropriately; a list of overrides present is displayed for the user to review |
| 1.7 | Plan name length | Plan name length is less than required for DICOM export |
| 1.8 | ROIs inside dose grid | All ROIs are covered entirely by dose grid (exceptions can be set in the configuration file) |
| 1.9 | Has the MU checker run? | For 3D conformal plans, prompts the planner to check the plan runs through the independent monitor unit checking software |
| 1.10 | Plan comment contains final | Plan is labeled correctly |
| 1.11 | Correct version of Pinnacle | Current clinical version of Pinnacle is in use |
AutoLock prescription checks.
|
|
|
|
|---|---|---|
| 2.1 | Prescription dose | Prescription options are set correctly; the prescription dose and number of fractions are displayed for the user to review |
| 2.2 | Prescription method | Correctly prescribed to point dose or ROI mean (and not monitor units); the prescription point or ROI is displayed for the user to review |
| 2.3 | Weights proportional to | Beam weights are correctly set proportional to monitor units or point dose |
| 2.4 | Isodoses | Isodose lines set appropriately; a list of isodoses present is displayed for the user to review |
| 2.5 | Max point dose less than or equal to 110% | Flag for potential hotspot issues |
AutoLock beam checks.
|
|
|
|
|---|---|---|
| 3.1 | Consistent machine and timestamp | All beams are set to the same beam model and timestamp matches the quality system; the beam model in use is displayed for the user to review |
| 3.2 | Consistent beam types | All beams are set to the same type (e.g., conventional, IMRT or VMAT) |
| 3.3 | All beams set to AC or CCC | All beams are set to either adaptive convolve or collapsed cone convolution |
| 3.4 | Consistent heterogeneity | All beams are set to heterogeneous or homogeneous and appropriate CT‐density table is in use |
| 3.5 | All beams set to consistent isocenter | All beams set to the same isocenter |
| 3.6 | All beams set to consistent prescription | All beams set to the same prescription unless multiple prescriptions are present, in which case the user must review |
| 3.7 | All beams have a field ID entered | All beams have a field ID entered to enable unique identification in the R&V system, regardless of beam name |
| 3.8 | Beam labels check | Beam labels are correct |
| 3.9 | In vivo points check |
|
AutoLock control point checks.
|
|
|
|
|---|---|---|
| 4.1 | No control points with less than 3 MUs | No control points have less than the minimum 3 monitor units |
| 4.2 | Y jaw backup | Y jaws are backed up correctly |
| 4.3 | X jaw backup | X jaws are backed up correctly |
| 4.4 | Elekta breast IMRT back edge check | Back edge jaw position is consistent for Elekta breast IMRT plans |
| 4.5 | No blocks present for plans not using MLC | No blocks are present for plan not using the MLC |
| 4.6 | Varian CAX matchlines | No matchlines present near the central axis for Varian step‐and‐shoot IMRT plans |
Thirteen‐category treatment plan QC rejection coding system.
|
|
|
|
|---|---|---|
| 1 | Contouring | ROI delineation and expansion |
| 2 | Density correction | Density overrides |
| 3 | Documentation | Missing or incorrect information in plan printout |
| 4 | Dose calculation | Beam model, dose calculation, voxel grid size, couch removal |
| 5 | InVivo points |
|
| 6 | Jaw/MLC | Jaw backup, control point constraints |
| 7 | Modification of treatment by clinician | Patient setup, ill‐fitting shell, changes to patient, PTV, etc. |
| 8 | Optimization | Coverage, OAR doses, beam energy, shielding |
| 9 | Origin | Localization |
| 10 | Prescription | Prescription and method |
| 11 | Wedge angle | Wedge angle not permitted in quality system |
| 12 | Wrong linac | Beam model does not match linac booking |
| 13 | Other | Reason not described in system; recurrent reasons can be reviewed for inclusion via an additional category |
Figure 6AutoLock compliance rate and action rate as a function of time, showing usage increased to 100% over the audit period and an upwards trend in action rate.
Figure 7AutoLock action rates for individual checks.
Figure 8Treatment plan QC rejection codes before and after the introduction of AutoLock.