Using failure mode and effects analysis (FMEA) to generate an initial plan check checklist for improved safety in radiation treatment.

Prema Rassiah1, Fan-Chi Frances Su1, Y Jessica Huang1, Dan Spitznagel2, Vikren Sarkar1, Martin W Szegedi1, Hui Zhao1, Adam B Paxton1, Geoff Nelson1, Bill J Salter1.   

Abstract

PURPOSE: To apply failure mode and effect analysis (FMEA) to generate an effective and efficient initial physics plan checklist.
METHODS: A team of physicists, dosimetrists, and therapists was setup to reconstruct the workflow processes involved in the generation of a treatment plan beginning from simulation. The team then identified possible failure modes in each of the processes. For each failure mode, the severity (S), frequency of occurrence (O), and the probability of detection (D) was assigned a value and the risk priority number (RPN) was calculated. The values assigned were based on TG 100. Prior to assigning a value, the team discussed the values in the scoring system to minimize randomness in scoring. A local database of errors was used to help guide the scoring of frequency.
RESULTS: Twenty-seven process steps and 50 possible failure modes were identified starting from simulation to the final approved plan ready for treatment at the machine. Any failure mode that scored an average RPN value of 20 or greater was deemed "eligible" to be placed on the second checklist. In addition, any failure mode with a severity score value of 4 or greater was also considered for inclusion in the checklist. As a by-product of this procedure, safety improvement methods such as automation and standardization of certain processes (e.g., dose constraint checking, check tools), removal of manual transcription of treatment-related information as well as staff education were implemented, although this was not the team's original objective. Prior to the implementation of the new FMEA-based checklist, an in-service for all the second checkers was organized to ensure further standardization of the process.
CONCLUSION: The FMEA proved to be a valuable tool for identifying vulnerabilities in our workflow and processes in generating a treatment plan and subsequently a new, more effective initial plan checklist was created.
© 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals, Inc. on behalf of American Association of Physicists in Medicine.

Entities:  

Keywords:  FMEA; initial plan check; quality assurance

Mesh:

Year:  2020        PMID: 32583912      PMCID: PMC7484852          DOI: 10.1002/acm2.12918

Source DB:  PubMed          Journal:  J Appl Clin Med Phys        ISSN: 1526-9914            Impact factor:   2.102


INTRODUCTION

Quality assurance (QA) is an integral and routine practice of radiation oncology departments. Over the years, QA has become more comprehensive to include not just equipment but also procedures, workflows, and communication. , , The initial plan check or second check done at the end of the treatment planning process is one such QA procedure implemented to ensure correct procedure and communication between various systems and teams (radiation oncologists, physicists, dosimetrists, therapists etc.) take place when a treatment plan is generated. Ford et al determined from studying the effectiveness of 15 different quality control (QC) tools that the initial plan check was the foremost effective tool in detecting potential error. The Medical Physics Practice Guideline (MPPG) for plan and chart review defines the purpose of the initial plan review as “to ensure compliance with prescription, no clinically significant deviations and that all information needed for the therapist to deliver the treatment is provided”. The American College of Radiology — American Association of Physicists in Medicine (ACRAAPM) technical standards for performance of radiation oncology physics for external beam defines the role of the qualified medical physicist in performing chart reviews and details essential items that need to be reviewed. These technical standards and guidelines provide recommendations on what needs to be checked, but are not department specific and may not be able to address vulnerabilities related to department‐specific workflows or procedures. The use of checklists has been shown to reduce medical errors , , and has been advocated by the World Health organization (WHO). A practice guideline on the development, implementation, use, and maintenance of safety checklist was published by AAPM in 2015 in recognition of the importance of a well‐structured checklist. The purpose of a checklist in the initial plan review is to serve as an aid to ensure crucial steps in the process are not forgotten. However, the success of the checklist depends on it being relevant and effective enough that the users of the checklist do not succumb to checklist fatigue. Recently in our department, a treatment planning system (TPS) upgrade and introduction of supporting software have led to workflow and procedural changes. Several items had to be added to the initial plan checklist to address incidents that occurred and may occur due to these changes, leading to a significantly lengthened checklist. In order to avoid checklist fatigue while keeping the list’s effectiveness, we decided to review the checklist as a whole. Failure mode and effects analysis (FMEA) is a widely used tool to analyze work processes methodically to identify vulnerabilities and their impact on safety. , , , Based on FMEA, resources can be concentrated on the most significant effect. In fact, the American Association of Physicists in Medicine’s (AAPM) Task Group (TG) 275 recommends that each practice should assess local processes and identify key high‐risk failure modes. We, therefore, decided to use FMEA to develop an initial plan checklist to ensure that effort is spent on checking high impact items, thereby making the process efficient and effective while avoiding checklist fatigue. To our knowledge, FMEA thus far has been used to review processes and workflows as a means to reduce risk , , , , , but has not been used for the purpose of creating a checklist. Here we report our experience on using FMEA to create an initial plan checklist specific to our department while adhering to ACR requirements, MPPG guidelines, and taking into account TG275 recommendations.

MATERIALS AND METHODS

Hardware and software systems

Two computed tomography (CT) simulators (GE lightspeed 16, GE Healthcare, Chicago, IL, USA and Somatom Definition, Siemens Healthcare, Erlangen, Germany) equipped with moveable lasers (LAP GmbH, Germany) are located within our department. Both scanners are used for simulation and with no specific patients directed to either scanners. Eclipse (version 15.5, Varian Medical Systems, Palo Alto, CA, USA) is used as both a virtual simulator and TPS. The algorithms used for dose calculation are Acuros version 15.5 for photon and electron Monte Carlo (EMC) version 15.5 for electrons. The TPS is complemented by a software package (MIM Vista, version 2.6, MIM, OH, USA), which is capable of performing functions such as image fusion, contouring, and dose summation. An Oncology Information System (OIS), (Mosaiq, version 2.6, Elekta, CA, USA) is used to transfer plan parameters required for treatment delivery to six linear accelerators: four Truebeams, one clinac iX (Varian Medical Systems, Palo Alto, CA, USA), and one Artiste (Siemens Healthcare, Erlangen, Germany). All linear accelerators are equipped with conebeam CT and the Artiste is equipped with CT on rails.

Clinical team

In the recent 5 yr, our clinical team in radiation oncology has almost doubled. Currently, our department consists of 10 attending physicians, 9 radiation oncology residents, 11 physicists, 2 physics residents, 1 medical physicist assistant, 9 dosimetrists, 42 therapists, 7 nurses, and 3 medical assistants.

Treatment planning process overview

Immobilization devices and scanning technique (such as the use of contrast, 4DCT, etc.) are determined during simulation based on written simulation orders. The treatment isocenter is placed during simulation with the dosimetrist and radiation oncologist present. Setup instructions, which include photos, are created by the simulation therapist during simulation and placed in the OIS. Treatment plans are generated by the dosimetrist and reviewed in the TPS by the radiation oncologist. The dosimetrist then exports the treatment fields, reference CT, digitally reconstructed radiographs (DRRs), and a pdf of the plan to the OIS. The plan and prescription are then approved by the radiation oncologist in the OIS. All plans are second checked by a physicist prior to treatment who then signs off on the plan. Currently, there is no formal prereview prior to the second check in our workflow. The OIS is configured such that the system will prevent treatment if the prescription is not approved by the radiation oncologist or if the plan is not approved by both the radiation oncologist and physicist. A summary of the overall workflow is shown in Fig. 1.
Fig. 1

Treatment plan generation flow.

Treatment plan generation flow.

FMEA process

A team of physicists, dosimetrists, and therapists worked together to reconstruct the workflow processes involved in generation of a treatment plan beginning from simulation. The team then identified possible failure modes in each of these processes. For each failure mode, the severity (S), frequency of occurrence (O), and the probability of detection (D) were assigned a value and the risk priority number (RPN) was calculated as the product of these values. The values assigned were based on the report of AAPM task group 100 (TG 100). The range of scores for the S, O, and D was 1–10, leading to RPN values ranging from 1–1000. Prior to each team member assigning a value, the team discussed the scoring system as laid out by TG 100 to minimize the randomness in scoring. The consequence/s of each failure mode was discussed so that all team members understand the severity or repercussion of each failure mode. The resulting averages were also discussed as a group. A local database of errors that were identified and caught during the second check between 2017 and 2018, as shown in Fig. 2, was used to help guide the scoring of frequency. The team members were shown the error frequency so that they could use it as a guide when scoring. No formal correlation was made between the frequency score and the error frequency. There was a total of 210 errors detected and an approximate of 5000 s checks were done during this 2‐yr period. The scoring was done based on the relative frequency of these errors.
Fig. 2

Frequency of recorded errors caught during the second check during 2017–2018. A total number of 210 errors were detected and an approximate of 5000 s checks were done during this 2‐yr period.

Frequency of recorded errors caught during the second check during 2017–2018. A total number of 210 errors were detected and an approximate of 5000 s checks were done during this 2‐yr period.

RESULTS

FMEA and RPN determination

Twenty‐seven process steps and 50 possible failure modes were identified starting from simulation to the final approved plan which is ready for treatment. A summary of the identified failure modes and the associated RPN scores is shown in Table 1. Any failure mode that scored an average RPN value of 20 or greater was deemed ‘eligible’ to be placed on the second checklist. This cutoff value was chosen somewhat arbitrarily and was aimed at balancing between the workload and the ability to catch errors. In addition, any event with a severity score value of four or greater was also placed on the checklist. Based on TG 100, a severity score of fou would lead to a minor dosimetric error. Any value less than 4 means an error that does not have a dosimetric effect but may cause inconvenience to team members or patients. Prior to placing an item on the checklist, two things were considered: firstly, whether there is a simple workflow change that would change the likelihood of occurrence (decrease O); and secondly, if the failure cannot be lessened, whether software can automate the check to increase the detectability (decrease D) and remove the human element.
Table 1

Summary of identified process steps and failure mode. RPN scores or severity score that exceed the set value are in italics. Items that had a high score but was addressed via change of process or automatically checked are highlighted in gray.

Process#Process stepFailure modeAverage
RPNSOD
Simulation1 Make immobilization device Immobilization device not sufficient for the accuracy of required for treatment.183.32.82.3
2 Scan patient Did not scan the entire length of treatment. 93.02.51.2
Wrong scan protocol used 182.02.03.8
Wrong site 53.02.51.3
3Place isocenterDid not place isocenter based on standard departmental protocol143.31.21.5
4Mark isocenter on patientIsocenter tattoo marked wrongly in the inf/sup direction, forgot (‐) sign103.02.81.5
Lasers are not in their zero positions when marking CT "0"92.52.01.7
5Write setup instructions in MosaiqWrong or incomplete setup instructions. That is, did not put in remove dentures, wrong shift to isocenter 25 4.02.71.3
Preplan preparation6Transfer CT to TPSIncomplete dataset transferred31.82.72.5
Wrong dataset transferred ex 2nd scan, contrast, 20 3.42.71.5
7Place beams on isocenter (s)Plan generated on wrong isocenter83.02.71.8
Plan generation and review8Generate planPlan does not adhere to constraints 66 4.21.81.2
Complicated plan generated, i.e., 14 fields when a simpler one is possible101.82.82.8
Previous RT tx dose not accounted for 44 6.42.81.2
Couch not accounted for in plan 25 3.83.0 6.4
Wrong prescription/ fractionation in plan (did not adhere to physician's order)176.22.23.2
Wrong algorithm, wrong CT to ED table, wrong grid size, wrong fusion, bolus not attached,194.41.2 5.6
Export and preparation in OIS9Export plan pdf to MosaiqWrong plan approved and exported to Mosaiq (multiple plans same patient) 74 6.43.61.8
Wrong shift instruction (primary to bst)133.61.42.0
Plan incomplete (surface rendering, no DVH etc.)152.21.43.0
10Export treatment fields to MosaiqWrong field ID ( bolus/non bolus) as per dept. protocol172.62.8 4.4
11Export DRR to MosaiqIsocenter placed wrongly on DRR 31 4.81.8 6.4
12Export DRR to Mosaiq Suboptimal (image quality of) DRRs exported to Mosaiq 72.22.41.4
13Export ref CT to MosaiqWrong reference CT exported52.43.61.8
No ref CT exported to Mosaiq52.03.61.6
14Export reference to IGRT systemsWrong isocenter exported to OIS, U/S , Exactrac, Calypso175.02.33.5
15Export structure set to MosaiqContours not exported as per physician ins.143.02.41.4
16Import tx fields into MosaiqIncomplete import of treatment fields64.22.01.0
Wrong field ID (wrong number, missing B, etc)172.82.41.0
Wrong dose rate131.62.81.8
Wrong energy (energy changed in MQ)75.22.62.4
Wrong MU (mismatched MU on hand calculation)124.21.41.2
Wrong/missing FDA code for electron fields. Wrong wedge112.63.81.4
Imaging field parameter wrong or missing (wrong name, gantry angle, field opening size, SSD, etc.)92.42.82.6
Wrong/missing SSD143.41.01.2
Possible collision (lack clearance) 26 6.82.01.4
18Attached DRR to fieldDRR not attached82.22.81.2
DRR attached wrongly 22 3.43.21.2
19Attach tolerance tableWrong tolerance table attached142.82.81.4
Missing tolerance table42.42.61.8
20Prep site setup in OISMissing site setup for electron fields62.23.41.2
21Associate plan(s) with prescription(s)Tx fields associated to wrong prescription93.02.82.2
22Augment setup instructions if neededPostsimulation isocenter shift is missing or isocenter added postsimulation183.42.61.6
23Setup treatment calendarTx calendar setup wrongly (every other day tx/bolus, concurrent treatment, number of fractions, dose accumulation) 28 4.01.81.0
24Prescription approvalPrescription not complete (missing energy, technique, etc.)92.62.41.0
25Imaging instructionSub optimal imaging instructions, i.e., frequency or imaging type not adequate 27 4.01.61.8
Instruction for alignment not clear, overly complicated103.02.22.0
Inappropriate imaging instructions (Imaging modality not available in vault, etc)102.43.02.0
26Special instructionMissing info. Ex “need diode measurement”183.42.21.6
Verification27IMRT QA/MUcheckFailed IMRT QA 25 23.61.8
Summary of identified process steps and failure mode. RPN scores or severity score that exceed the set value are in italics. Items that had a high score but was addressed via change of process or automatically checked are highlighted in gray. The severity scoring in our institution may differ from that of another institution depending on the work process. For example, the score of 3.0 may seem low for “wrong site scanned” but in our workflow, the simulation therapist scans the site to be planned based on written instructions by the physician given prior to scanning. The physician and the dosimetrist are present during simulation and confirm the site and isocenter location before the patient is taken of the table. Therefore, in our case the worst case scenario is that the patient is rescanned, resulting in a low severity score.

Checklist creation

Initially, 18 items were deemed “eligible” to be placed on the checklist. One of these was “wrong MU calculation,” which is mainly caused by how the MUs are calculated for electrons. While the dose distribution is calculated using Eclipse’s implementation of EMC, the final MU setting is calculated using a hand calculation, a manual process. This was addressed by devoting physics resources into fully commissioning the EMC algorithm in Eclipse. The electron plan MU is now calculated by the TPS, eliminating a manual calculation and therefore the failure mode “wrong MU calculation” was removed from the checklist. Four other failure modes namely “dose constraints not being adhered to,” “treatment couch not accounted for,” “wrong calculation parameter,” and “clearance” can be checked by an automated software, Clearcheck (Radformation, NY, USA), and was placed under a single item in the checklist. The planner is required to run the automated software prior to the second check. The second checker will then review the report of the program as part of the check. Table 2 shows the new second checklist and the old checklist is shown in Table 3 for comparison.
Table 2

Second check Checklist as per the FMEA process.

Checklist category#Checklist itemWhat to check
Simulation1Setup instructions complete and correctComplete, correct shift/table top, SSD, bolus thickness. Postsimulation isocenter shifts?
Plan generation2Planning datasetCorrect dataset used for planning (contrast/empty/full bladder).
3Previous treatmentIf previously treated, is the previous treatment accounted for.
4PrescriptionIs the prescription complete (Energy, technique and fractionation, dose)? Does prescription match the narrative summary and plan
5ClearcheckDose constraints adhered to, Couch inserted correctly, correct algorithm, grid size, clearance
Export and preparation in OIS6Correct plan exported to OISWas the plan that was approved by the physician in the planning system exported to the OIS?
7Export DRR to MosaiqDoes the isocenter on the DRR look reasonable with plan?
8Export reference to IGRT systemsIs the isocenter exported to the imaging system correct (U/S, Exactrac, CBCT, Calypso etc.) and match plan?
9Treatment fieldDo the treatment fields in OIS match plan
10Setup treatment calendarIs the treatment calendar setup per prescription? Every other day tx/bolus, total fraction and dose
11Imaging instructionDoes imaging instruction seem reasonable (frequency/modality)?
Verification12IMRT QA/MU checkDid IMRT QA pass per dept. policy? Modulation OK? MU < 5X prescribed dose
Table 3

Shows the checklist used prior to the FMEA process.

Checklist category#Checklist item
Prescription1Prescription complete
2Prescription approved by AU
3Narrative summary match prescribed treatment
Plan4Tx plan matches prescription
5Hetero and bulk density correction
6Does the dose distribution look reasonable?
7Isocenter coordinates matches plan
8Does the DVH look reasonable
9Are the MUs reasonable as per prescription?
10Are the MUs checked?
11Clearance
OIS12Appropriate tolerance table used?
13Correct beam description and #
14DRRs correctly labelled
15Site setup approved
16Does MU match plan
17Tx parameter match plan
18Are set up instructions clear?
19Is the treatment calendar set up correctly?
20Wkly/Final/Protocol QCL available
21Special instructions clear
22Plan billing done
QA23Physics consult fulfilled?
24If IMRT, was the IMRT QA done prior to treatment
Second check Checklist as per the FMEA process. Shows the checklist used prior to the FMEA process. Significant additions to the second checklist generated by the FMEA process are item #3: previous treatment, item#8: iso “plan coordinates in IGRT systems” and item#11: “imaging instruction.” These are all items that were not checked with the previous checklist but the team deemed these items to be potentially unsafe for treatment if an error was made. Several items were consolidated in item #5: clearcheck. The items that were omitted that were previously checked in Table 3 are item #20: “weekly/final QCL” and #22: “plan billing.” Item “weekly/final QCL” is now automated, whereas “plan billing” was revealed to be checked by the billing department during the FMEA process, meaning it was a redundant check that could be dropped. With the exception of item #22: “Plan billing done” and #23: “physics consult fulfilled” in the old checklist, all other items and more are checked in the new checklist. Despite this, the number of items in the new checklist is only 12 compared to the old checklist which has 24 items. This reduction in number was achieved by using clearcheck and consolidating several items that can be checked simultaneously into one item. The amount of time the second checker spends checking a plan with the new checklist may remain the same but the processes and the method of checking is more efficient than it used to be.

Additional mitigation strategies

The process of systematically analyzing the treatment plan generation workflow and processes via FMEA resulted in the identification of other safety improvements, such as automation of certain processes, removal of manual transcription of treatment‐related information, and standardization of plan printouts in terms of content and organization. The purpose of standardizing the plan printout is to ensure that the required plan information is always available, should certain information be needed during treatment for patient setup, without the need for scrolling of the plan document back and forth in search of specific information. This standardization also facilitates a more efficient plan review by both physician and physicist. Prior to the implementation of the new checklist, an in‐service for all the second checkers was organized to ensure further standardization of the second check process. The “assessment’ module in Mosaiq is utilized to create a checklist with explanations on what needs to be checked. The second checker is required to complete the checklist by checking off each item electronically in Mosaiq.

DISCUSSION

Failure mode and effects analysis proved to be a very effective, systematic method of reviewing workflows and processes to identify vulnerabilities. Although not the initial objective of the team, workflow changes were implemented, certain checks were automated, and the standardization of certain processes was initiated. Standardization plays an important role in error prevention. While the order and content of the plan printout has been standardized, the process of putting much of the report together is still manual. For example, the planner prints a PDF document of the treatment summary (for therapist time out), plan summary, dose distribution, dose constraints adherence etc. separately and then combines them in a given order before importing the report into Mosaiq. A script is currently being developed within Eclipse to generate the whole report with a single click which removes the manual step involved with report generation. As shown in Fig. 2, errors made in the treatment field parameters of the OIS are one of the most frequently observed events. Considerable effort has been put into creating a more seamless export (from Eclipse) and import (into Mosaiq), where the planner does not have to interact with the treatment field once it is imported into the OIS. For example, block codes were setup in Eclipse, dose rates for certain treatments were defaulted in Eclipse, and initial treatment couch position (vertical, longitudinal and lateral) was setup as a default within Mosaiq so that the planner does not have to manually enter these values. Clearcheck is a software module integrated within the Eclipse environment that is capable of automatically verifying several things. Although some of these items are not individually listed in the checklist, they are verified in Clearcheck and the report is attached to the plan in the OIS system for the second checker to review. The utilization of third party software (Clearcheck), standardization of parameter transferred between Eclipse and Mosaiq, and treatment printouts are all efforts to transition from a somewhat manual workflow to an automated workflow. Automation has been shown to be one of the most effective ways of preventing errors in the radiation planning and treatment workflow. , , , The other by‐product of this process is the improved communication among the physicists, dosimetrists, and therapists and better understanding of each other’s work processes. While there are many similarities between the second checklist generated in this study and the example second checklist given in TG275, there are items that are on our list which we think are specific to our department. For example, item #6 on our checklist “Was the plan that was approved by the physician in the planning system exported to the OIS?” In our workflow, the radiation oncologist chooses a plan in the TPS (if more than one plan is presented to the physician). The radiation oncologist then labels the plan that he or she likes with his or her initials. The dosimetrist then exports the plan with the label. There is a possibility for the wrong plan to be exported to the OIS. This may not be the case for a department where the TPS is an integral part of the OIS. There are also very explicit items in our checklist to ensure physician intent and plan is translated to the OIS which are not listed in TG275 (items # 7, #9, and #10). The TG275 second checklist explicitly spells out every item in detail, whereas our checklist consolidates several checks in a single item. For example, TG275 spells out prescription energy, modality, technique, etc. as separate items, whereas we chose to consolidate this item as prescription (see #4 in Table 2). The use of clearcheck also enabled us to consolidate a number of items such as grid size, couch insertion, heterogeneity check, etc. into a single item. The one item that is on the TG275 list but not on ours is “Special Considerations for radiotherapy (e.g., pacemakers, ICDs, pumps, etc.).” In our department, there is a separate process and workflow to deal with implanted devices which the second checker does not have to deal with and hence it is not listed in our second checklist. Our current study has several limitations; one is that the scoring system is subjective. We have tried to address this by discussing each failure mode and the consequent severity and the ability to detect each failure amongst the team members. In order to reduce the randomness of the “frequency of occurrence” scoring, all team members were shown the frequencies of incidents in the local database. The second limitation is the local database in itself. The accuracy of frequency is dependent on the ability of our department to capture all incidents in this database. The third limitation is the somewhat arbitrary choice of the cutoff score value (20) chosen for an item to be listed in the checklist. If the value is too large we may omit some important items and if it is too small then, there may be too many items on the checklist. We tried to mitigate this concern by adding another criteria, which is, any item with a severity score of 4 or greater is also placed on the checklist.

Future work

One of the key recommendations of TG275 is to incorporate a formal physics review of critical data earlier in the treatment planning and not rely solely on the second check which is done at the end of planning. This will certainly be helpful in catching errors earlier on in the process and avoid a last minute replan or addressing of errors. We are assessing the feasibility of doing this in terms of man hours required, the work process, and the timeliness of getting this done without slowing down the planning process. We plan to evaluate the effectiveness of the new checklist in error detection by comparing the error detection rate pre‐ and postimplementation of the new checklist as items that are not on the list may not be checked.

CONCLUSION

Failure Mode and Effects Analysis proved to be a valuable tool in identifying vulnerabilities within our workflow and processes in generating a treatment plan, subsequently leading to the creation of a new initial plan checklist that may be both more efficient and effective. As radiation treatment technology advances, the changes in the clinical workflow is constant. Therefore, even after the initial second checklist is established, it is equally important to maintain a systematic periodic review of the second check process to ensure that the checklist is kept efficient and effective. It will also be beneficial to review the list, when new software, upgrades, programs, or technologies are introduced into the clinic.

CONFLICT OF INTEREST

None.
  24 in total

1.  How safe is safe? Risk in radiotherapy.

Authors:  Eric C Ford; Stephanie Terezakis
Journal:  Int J Radiat Oncol Biol Phys       Date:  2010-10-01       Impact factor: 7.038

2.  Linking quality assurance to performance improvement to produce a high reliability organization.

Authors:  Andrea B Silvey; Louise H Warrick
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008       Impact factor: 7.038

3.  Failure mode and effect analysis oriented to risk-reduction interventions in intraoperative electron radiation therapy: the specific impact of patient transportation, automation, and treatment planning availability.

Authors:  Juan López-Tarjuelo; Ana Bouché-Babiloni; Agustín Santos-Serra; Virginia Morillo-Macías; Felipe A Calvo; Yuri Kubyshin; Carlos Ferrer-Albiach
Journal:  Radiother Oncol       Date:  2014-11-24       Impact factor: 6.280

4.  Practical implementation of failure mode and effects analysis for safety and efficiency in stereotactic radiosurgery.

Authors:  Kelly Cooper Younge; Yizhen Wang; John Thompson; Julia Giovinazzo; Marisa Finlay; Raxa Sankreacha
Journal:  Int J Radiat Oncol Biol Phys       Date:  2015-02-07       Impact factor: 7.038

5.  Failure modes and effects analysis in image-guided high-dose-rate brachytherapy: Quality control optimization to reduce errors in treatment volume.

Authors:  Shada Wadi-Ramahi; Waleed Alnajjar; Rana Mahmood; Noha Jastaniyah; Belal Moftah
Journal:  Brachytherapy       Date:  2016-07-27       Impact factor: 2.362

6.  Evaluation of safety in a radiation oncology setting using failure mode and effects analysis.

Authors:  Eric C Ford; Ray Gaudette; Lee Myers; Bruce Vanderver; Lilly Engineer; Richard Zellars; Danny Y Song; John Wong; Theodore L Deweese
Journal:  Int J Radiat Oncol Biol Phys       Date:  2009-05-04       Impact factor: 7.038

7.  The impact of treatment complexity and computer-control delivery technology on treatment delivery errors.

Authors:  B A Fraass; K L Lash; G M Matrone; S K Volkman; D L McShan; M L Kessler; A S Lichter
Journal:  Int J Radiat Oncol Biol Phys       Date:  1998-10-01       Impact factor: 7.038

8.  A method for evaluating quality assurance needs in radiation therapy.

Authors:  M Saiful Huq; Benedick A Fraass; Peter B Dunscombe; John P Gibbons; Geoffrey S Ibbott; Paul M Medin; Arno Mundt; Sassa Mutic; Jatinder R Palta; Bruce R Thomadsen; Jeffrey F Williamson; Ellen D Yorke
Journal:  Int J Radiat Oncol Biol Phys       Date:  2008       Impact factor: 7.038

9.  Strategies for effective physics plan and chart review in radiation therapy: Report of AAPM Task Group 275.

Authors:  Eric Ford; Leigh Conroy; Lei Dong; Luis Fong de Los Santos; Anne Greener; Grace Gwe-Ya Kim; Jennifer Johnson; Perry Johnson; James G Mechalakos; Brian Napolitano; Stephanie Parker; Deborah Schofield; Koren Smith; Ellen Yorke; Michelle Wells
Journal:  Med Phys       Date:  2020-04-15       Impact factor: 4.071

10.  AutoLock: a semiautomated system for radiotherapy treatment plan quality control.

Authors:  Joseph M Dewhurst; Matthew Lowe; Mark J Hardy; Christopher J Boylan; Philip Whitehurst; Carl G Rowbottom
Journal:  J Appl Clin Med Phys       Date:  2015-05-08       Impact factor: 2.102

View more
  2 in total

1.  Financial and Safety Impact of Simulation-based Clinical Systems Testing on Pediatric Trauma Center Transitions.

Authors:  Sacha A Williams; Katie Fitzpatrick; Nicole M Chandler; Jennifer L Arnold; Christopher W Snyder
Journal:  Pediatr Qual Saf       Date:  2022-08-26

2.  Failure modes and effects analysis for surface-guided DIBH breast radiotherapy.

Authors:  Megan Bright; Ryan D Foster; Carnell J Hampton; Justin Ruiz; Benjamin Moeller
Journal:  J Appl Clin Med Phys       Date:  2022-02-02       Impact factor: 2.102

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.