Prema Rassiah1, Fan-Chi Frances Su1, Y Jessica Huang1, Dan Spitznagel2, Vikren Sarkar1, Martin W Szegedi1, Hui Zhao1, Adam B Paxton1, Geoff Nelson1, Bill J Salter1.
Abstract
PURPOSE: To apply failure mode and effect analysis (FMEA) to generate an effective and efficient initial physics plan checklist.Entities:
Keywords: FMEA; initial plan check; quality assurance
Mesh:
Year: 2020 PMID: 32583912 PMCID: PMC7484852 DOI: 10.1002/acm2.12918
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Fig. 1Treatment plan generation flow.
Fig. 2Frequency of recorded errors caught during the second check during 2017–2018. A total number of 210 errors were detected and an approximate of 5000 s checks were done during this 2‐yr period.
Summary of identified process steps and failure mode. RPN scores or severity score that exceed the set value are in italics. Items that had a high score but was addressed via change of process or automatically checked are highlighted in gray.
| Process | # | Process step | Failure mode | Average | |||
|---|---|---|---|---|---|---|---|
| RPN | S | O | D | ||||
| Simulation | 1 |
| Immobilization device not sufficient for the accuracy of required for treatment. | 18 | 3.3 | 2.8 | 2.3 |
| 2 |
|
| 9 | 3.0 | 2.5 | 1.2 | |
|
| 18 | 2.0 | 2.0 | 3.8 | |||
|
| 5 | 3.0 | 2.5 | 1.3 | |||
| 3 | Place isocenter | Did not place isocenter based on standard departmental protocol | 14 | 3.3 | 1.2 | 1.5 | |
| 4 | Mark isocenter on patient | Isocenter tattoo marked wrongly in the inf/sup direction, forgot (‐) sign | 10 | 3.0 | 2.8 | 1.5 | |
| Lasers are not in their zero positions when marking CT "0" | 9 | 2.5 | 2.0 | 1.7 | |||
| 5 | Write setup instructions in Mosaiq | Wrong or incomplete setup instructions. That is, did not put in remove dentures, wrong shift to isocenter |
| 4.0 | 2.7 | 1.3 | |
| Preplan preparation | 6 | Transfer CT to TPS | Incomplete dataset transferred | 3 | 1.8 | 2.7 | 2.5 |
| Wrong dataset transferred ex 2nd scan, contrast, |
| 3.4 | 2.7 | 1.5 | |||
| 7 | Place beams on isocenter (s) | Plan generated on wrong isocenter | 8 | 3.0 | 2.7 | 1.8 | |
| Plan generation and review | 8 | Generate plan | Plan does not adhere to constraints |
| 4.2 | 1.8 | 1.2 |
| Complicated plan generated, i.e., 14 fields when a simpler one is possible | 10 | 1.8 | 2.8 | 2.8 | |||
| Previous RT tx dose not accounted for |
| 6.4 | 2.8 | 1.2 | |||
| Couch not accounted for in plan |
| 3.8 | 3.0 |
| |||
| Wrong prescription/ fractionation in plan (did not adhere to physician's order) | 17 | 6.2 | 2.2 | 3.2 | |||
| Wrong algorithm, wrong CT to ED table, wrong grid size, wrong fusion, bolus not attached, | 19 | 4.4 | 1.2 |
| |||
| Export and preparation in OIS | 9 | Export plan pdf to Mosaiq | Wrong plan approved and exported to Mosaiq (multiple plans same patient) |
| 6.4 | 3.6 | 1.8 |
| Wrong shift instruction (primary to bst) | 13 | 3.6 | 1.4 | 2.0 | |||
| Plan incomplete (surface rendering, no DVH etc.) | 15 | 2.2 | 1.4 | 3.0 | |||
| 10 | Export treatment fields to Mosaiq | Wrong field ID ( bolus/non bolus) as per dept. protocol | 17 | 2.6 | 2.8 |
| |
| 11 | Export DRR to Mosaiq | Isocenter placed wrongly on DRR |
| 4.8 | 1.8 |
| |
| 12 | Export DRR to Mosaiq |
| 7 | 2.2 | 2.4 | 1.4 | |
| 13 | Export ref CT to Mosaiq | Wrong reference CT exported | 5 | 2.4 | 3.6 | 1.8 | |
| No ref CT exported to Mosaiq | 5 | 2.0 | 3.6 | 1.6 | |||
| 14 | Export reference to IGRT systems | Wrong isocenter exported to OIS, U/S , Exactrac, Calypso | 17 | 5.0 | 2.3 | 3.5 | |
| 15 | Export structure set to Mosaiq | Contours not exported as per physician ins. | 14 | 3.0 | 2.4 | 1.4 | |
| 16 | Import tx fields into Mosaiq | Incomplete import of treatment fields | 6 | 4.2 | 2.0 | 1.0 | |
| Wrong field ID (wrong number, missing B, etc) | 17 | 2.8 | 2.4 | 1.0 | |||
| Wrong dose rate | 13 | 1.6 | 2.8 | 1.8 | |||
| Wrong energy (energy changed in MQ) | 7 | 5.2 | 2.6 | 2.4 | |||
| Wrong MU (mismatched MU on hand calculation) | 12 | 4.2 | 1.4 | 1.2 | |||
| Wrong/missing FDA code for electron fields. Wrong wedge | 11 | 2.6 | 3.8 | 1.4 | |||
| Imaging field parameter wrong or missing (wrong name, gantry angle, field opening size, SSD, etc.) | 9 | 2.4 | 2.8 | 2.6 | |||
| Wrong/missing SSD | 14 | 3.4 | 1.0 | 1.2 | |||
| Possible collision (lack clearance) |
| 6.8 | 2.0 | 1.4 | |||
| 18 | Attached DRR to field | DRR not attached | 8 | 2.2 | 2.8 | 1.2 | |
| DRR attached wrongly |
| 3.4 | 3.2 | 1.2 | |||
| 19 | Attach tolerance table | Wrong tolerance table attached | 14 | 2.8 | 2.8 | 1.4 | |
| Missing tolerance table | 4 | 2.4 | 2.6 | 1.8 | |||
| 20 | Prep site setup in OIS | Missing site setup for electron fields | 6 | 2.2 | 3.4 | 1.2 | |
| 21 | Associate plan(s) with prescription(s) | Tx fields associated to wrong prescription | 9 | 3.0 | 2.8 | 2.2 | |
| 22 | Augment setup instructions if needed | Postsimulation isocenter shift is missing or isocenter added postsimulation | 18 | 3.4 | 2.6 | 1.6 | |
| 23 | Setup treatment calendar | Tx calendar setup wrongly (every other day tx/bolus, concurrent treatment, number of fractions, dose accumulation) |
| 4.0 | 1.8 | 1.0 | |
| 24 | Prescription approval | Prescription not complete (missing energy, technique, etc.) | 9 | 2.6 | 2.4 | 1.0 | |
| 25 | Imaging instruction | Sub optimal imaging instructions, i.e., frequency or imaging type not adequate |
| 4.0 | 1.6 | 1.8 | |
| Instruction for alignment not clear, overly complicated | 10 | 3.0 | 2.2 | 2.0 | |||
| Inappropriate imaging instructions (Imaging modality not available in vault, etc) | 10 | 2.4 | 3.0 | 2.0 | |||
| 26 | Special instruction | Missing info. Ex “need diode measurement” | 18 | 3.4 | 2.2 | 1.6 | |
| Verification | 27 | IMRT QA/MUcheck | Failed IMRT QA |
| 2 | 3.6 | 1.8 |
Second check Checklist as per the FMEA process.
| Checklist category | # | Checklist item | What to check |
|---|---|---|---|
| Simulation | 1 | Setup instructions complete and correct | Complete, correct shift/table top, SSD, bolus thickness. Postsimulation isocenter shifts? |
| Plan generation | 2 | Planning dataset | Correct dataset used for planning (contrast/empty/full bladder). |
| 3 | Previous treatment | If previously treated, is the previous treatment accounted for. | |
| 4 | Prescription | Is the prescription complete (Energy, technique and fractionation, dose)? Does prescription match the narrative summary and plan | |
| 5 | Clearcheck | Dose constraints adhered to, Couch inserted correctly, correct algorithm, grid size, clearance | |
| Export and preparation in OIS | 6 | Correct plan exported to OIS | Was the plan that was approved by the physician in the planning system exported to the OIS? |
| 7 | Export DRR to Mosaiq | Does the isocenter on the DRR look reasonable with plan? | |
| 8 | Export reference to IGRT systems | Is the isocenter exported to the imaging system correct (U/S, Exactrac, CBCT, Calypso etc.) and match plan? | |
| 9 | Treatment field | Do the treatment fields in OIS match plan | |
| 10 | Setup treatment calendar | Is the treatment calendar setup per prescription? Every other day tx/bolus, total fraction and dose | |
| 11 | Imaging instruction | Does imaging instruction seem reasonable (frequency/modality)? | |
| Verification | 12 | IMRT QA/MU check | Did IMRT QA pass per dept. policy? Modulation OK? MU < 5X prescribed dose |
Shows the checklist used prior to the FMEA process.
| Checklist category | # | Checklist item |
|---|---|---|
| Prescription | 1 | Prescription complete |
| 2 | Prescription approved by AU | |
| 3 | Narrative summary match prescribed treatment | |
| Plan | 4 | Tx plan matches prescription |
| 5 | Hetero and bulk density correction | |
| 6 | Does the dose distribution look reasonable? | |
| 7 | Isocenter coordinates matches plan | |
| 8 | Does the DVH look reasonable | |
| 9 | Are the MUs reasonable as per prescription? | |
| 10 | Are the MUs checked? | |
| 11 | Clearance | |
| OIS | 12 | Appropriate tolerance table used? |
| 13 | Correct beam description and # | |
| 14 | DRRs correctly labelled | |
| 15 | Site setup approved | |
| 16 | Does MU match plan | |
| 17 | Tx parameter match plan | |
| 18 | Are set up instructions clear? | |
| 19 | Is the treatment calendar set up correctly? | |
| 20 | Wkly/Final/Protocol QCL available | |
| 21 | Special instructions clear | |
| 22 | Plan billing done | |
| QA | 23 | Physics consult fulfilled? |
| 24 | If IMRT, was the IMRT QA done prior to treatment |