Comparison of 3 marking techniques in preoperative assessment of toric intraocular lenses using a wavefront aberrometer.

Young Jae Woo1, Hun Lee1, Hong Seok Kim1, Eung Kweon Kim1, Kyoung Yul Seo1, Tae-im Kim2.   

Abstract

PURPOSE: To assess the accuracy of 3 corneal horizontal meridian marking techniques using the iTrace Surgical Workstation in preoperative assessment for toric intraocular lens (IOL) implantation.
SETTING: Department of Ophthalmology, Yonsei University College of Medicine, Seoul, South Korea.
DESIGN: Prospective comparative observational study.
METHODS: Patients were allocated into 3 groups. Three techniques-marking with a nonpendular marker with a surgeon's direct visualization, slitlamp-assisted marking with a pendulum-attached marker, and slitlamp-assisted marking with a horizontal slit beam-were used to perform the horizontal meridian marking. The marks were immediately documented and aligned with the Zaldivar toric caliper transparent topographic map. Using a horizontal reference meridian determined by the toric caliper, the accuracy of corneal horizontal marking techniques was quantitatively evaluated by determining the rotational deviation and vertical misalignment.
RESULTS: For rotational deviation, the pendular marking showed the fewest errors to the horizontal reference meridian (mean error -0.66 degrees; mean absolute error [MAE] 3.77 degrees). There was a significant difference between the pendular marking and the surgeon's direct visual marking. For the MAE, the pendular marking showed less deviation than the surgeon's direct visual marking and the horizontal slit-beam marking. With vertical misalignment, there was no significant difference between the 3 different corneal horizontal meridian marking techniques.
CONCLUSION: The slitlamp-assisted technique with the pendulum-attached marker was more precise than marking with a nonpendular marker with a surgeon's direct visualization and marking with a horizontal slit beam in preoperative assessment for toric IOLs. FINANCIAL DISCLOSURE: No author has a financial or proprietary interest in any material or method mentioned.
Copyright © 2015 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2015        PMID: 26092366     DOI: 10.1016/j.jcrs.2014.09.045

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  12 in total

1.  Toric IOL positioning with a no-touch head-up display axis alignment.

Authors:  J Luebke; D Boehringer; P Maier; T Reinhard; P Eberwein
Journal:  Int Ophthalmol       Date:  2019-11-23       Impact factor: 2.031

2.  Evaluation of axis alignment and refractive results of toric phakic IOL using image-guided system.

Authors:  Sherif Emerah
Journal:  Int J Ophthalmol       Date:  2020-04-18       Impact factor: 1.779

3.  A comparison of three different corneal marking methods used to determine cyclotorsion in the horizontal meridian.

Authors:  Hung-Yuan Lin; Yi-Ting Fang; Ya-Jung Chuang; Justin N Karlin; Hsin-Yang Chen; Szu-Yuan Lin; Pi-Jung Lin; Ming Chen
Journal:  Clin Ophthalmol       Date:  2017-02-08

Review 4.  Optimizing outcomes with toric intraocular lenses.

Authors:  Manpreet Kaur; Farin Shaikh; Ruchita Falera; Jeewan S Titiyal
Journal:  Indian J Ophthalmol       Date:  2017-12       Impact factor: 1.848

5.  Comparative evaluation of toric intraocular lens alignment and visual quality with image-guided surgery and conventional three-step manual marking.

Authors:  Jeewan S Titiyal; Manpreet Kaur; Cijin P Jose; Ruchita Falera; Ashutosh Kinkar; Lalit Ms Bageshwar
Journal:  Clin Ophthalmol       Date:  2018-04-24

6.  Novel method for preventing cyclorotation in Ziemer Femto LDV Z8 femtosecond laser-assisted cataract surgery with Verion image-guided system.

Authors:  Hung-Yuan Lin; Ya-Jung Chuang; Pi-Jung Lin; Yi-Ju Ho
Journal:  Clin Ophthalmol       Date:  2019-02-20

7.  Early visual quality outcomes after small-incision lenticule extraction surgery for correcting high myopic astigmatism.

Authors:  Xiangtao Hou; Kaixuan Du; Dan Wen; Shengfa Hu; Tu Hu; Chenling Li; Yanhui Tang; Xiaoying Wu
Journal:  BMC Ophthalmol       Date:  2021-01-19       Impact factor: 2.209

8.  Three-Random-Point Marking Method for Toric Intraocular Lens Alignment Using the iTrace Aberrometer.

Authors:  Yongyi Niu; Hongliang Lin; Yongjie Qin; Cheng Yang; Yanlei Chen; Jin Zeng; Hongyang Zhang
Journal:  J Ophthalmol       Date:  2021-04-08       Impact factor: 1.909

9.  Evaluation of Scheimpflug imaging system as an added tool in improving the accuracy of reference marking (as compared to the slit lamp marking system) for toric intraocular lens implantation.

Authors:  Ajay Sharma; Akanksha Batra
Journal:  Indian J Ophthalmol       Date:  2020-04       Impact factor: 1.848

10.  Comparison Between Digital and Manual Marking for Toric Intraocular Lenses: A Randomized Trial.

Authors:  Abdel Hamid Elhofi; Hany Ahmed Helaly
Journal:  Medicine (Baltimore)       Date:  2015-09       Impact factor: 1.817

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.