Design and performance evaluation of a master controller for endovascular catheterization.

Jin Guo1, Shuxiang Guo2,3, Takashi Tamiya4, Hideyuki Hirata5, Hidenori Ishihara5.   

Abstract

PURPOSE: It is difficult to manipulate a flexible catheter to target a position within a patient's complicated and delicate vessels. However, few researchers focused on the controller designs with much consideration of the natural catheter manipulation skills obtained from manual catheterization. Also, the existing catheter motion measurement methods probably lead to the difficulties in designing the force feedback device. Additionally, the commercially available systems are too expensive which makes them cost prohibitive to most hospitals. This paper presents a simple and cost-effective master controller for endovascular catheterization that can allow the interventionalists to apply the conventional pull, push and twist of the catheter used in current practice.
METHODS: A catheter-sensing unit (used to measure the motion of the catheter) and a force feedback unit (used to provide a sense of resistance force) are both presented. A camera was used to allow a contactless measurement avoiding additional friction, and the force feedback in the axial direction was provided by the magnetic force generated between the permanent magnets and the powered coil.
RESULTS: Performance evaluation of the controller was evaluated by first conducting comparison experiments to quantify the accuracy of the catheter-sensing unit, and then conducting several experiments to evaluate the force feedback unit. From the experimental results, the minimum and the maximum errors of translational displacement were 0.003 mm (0.01 %) and 0.425 mm (1.06 %), respectively. The average error was 0.113 mm (0.28 %). In terms of rotational angles, the minimum and the maximum errors were 0.39°(0.33 %) and 7.2°(6 %), respectively. The average error was 3.61°(3.01 %). The force resolution was approximately 25 mN and a maximum current of 3A generated an approximately 1.5 N force.
CONCLUSION: Based on analysis of requirements and state-of-the-art computer-assisted and robot-assisted training systems for endovascular catheterization, a new master controller with force feedback interface was proposed to maintain the natural endovascular catheterization skills of the interventionalists.

Entities:  

Keywords:  A force feedback device; A master controller; Contactless measurement; Endovascular catheterization; Training

Mesh:

Year:  2015        PMID: 26067289     DOI: 10.1007/s11548-015-1211-4

Source DB:  PubMed          Journal:  Int J Comput Assist Radiol Surg        ISSN: 1861-6410            Impact factor:   2.924


  13 in total

Review 1.  Simulation-based training: the next revolution in radiology education?

Authors:  Terry S Desser
Journal:  J Am Coll Radiol       Date:  2007-11       Impact factor: 5.532

2.  Initial experience with a telerobotic system to remotely navigate and automatically reposition standard steerable EP catheters.

Authors:  Laura Cercenelli; Emanuela Marcelli; Gianni Plicchi
Journal:  ASAIO J       Date:  2007 Sep-Oct       Impact factor: 2.872

3.  Design and performance evaluation of a remote catheter navigation system.

Authors:  Yogesh Thakur; Jeffrey S Bax; David W Holdsworth; Maria Drangova
Journal:  IEEE Trans Biomed Eng       Date:  2009-03-27       Impact factor: 4.538

Review 4.  Catheter ablation - new developments in robotics.

Authors:  K R Julian Chun; Boris Schmidt; Bülent Köktürk; Roland Tilz; Alexander Fürnkranz; Melanie Konstantinidou; Erik Wissner; Andreas Metzner; Feifan Ouyang; Karl-Heinz Kuck
Journal:  Herz       Date:  2008-12       Impact factor: 1.443

5.  Intravascular catheter navigation using path planning and virtual visual feedback for oral cancer treatment.

Authors:  Junchen Wang; Takashi Ohya; Hongen Liao; Ichiro Sakuma; Tianmiao Wang; Iwai Tohnai; Toshinori Iwai
Journal:  Int J Med Robot       Date:  2011-04-28       Impact factor: 2.547

6.  Vessel bifurcation localization based on intraoperative three-dimensional ultrasound and catheter path for image-guided catheter intervention of oral cancers.

Authors:  Kuan Luan; Takashi Ohya; Hongen Liao; Etsuko Kobayashi; Ichiro Sakuma
Journal:  Comput Med Imaging Graph       Date:  2013-02-21       Impact factor: 4.790

7.  Principal considerations for the contemporary high-fidelity endovascular simulator design used in training and evaluation.

Authors:  Benjamin A Eslahpazir; Jerry Goldstone; Matthew T Allemang; John C Wang; Vikram S Kashyap
Journal:  J Vasc Surg       Date:  2014-01-11       Impact factor: 4.268

Review 8.  Clinical applications of robotic technology in vascular and endovascular surgery.

Authors:  George A Antoniou; Celia V Riga; Erik K Mayer; Nicholas J W Cheshire; Colin D Bicknell
Journal:  J Vasc Surg       Date:  2011-02       Impact factor: 4.268

9.  Design and fabrication of a robotic mechanism for remote steering and positioning of interventional devices.

Authors:  Govindarajan Srimathveeravalli; Thenkurussi Kesavadas; Xinyan Li
Journal:  Int J Med Robot       Date:  2010-06       Impact factor: 2.547

10.  Atrial fibrillation ablation using a robotic catheter remote control system: initial human experience and long-term follow-up results.

Authors:  Walid Saliba; Vivek Y Reddy; Oussama Wazni; Jennifer E Cummings; J David Burkhardt; Michel Haissaguerre; Josef Kautzner; Petr Peichl; Petr Neuzil; Volker Schibgilla; Georg Noelker; Johannes Brachmann; Luigi Di Biase; Conor Barrett; Pierre Jais; Andrea Natale
Journal:  J Am Coll Cardiol       Date:  2008-06-24       Impact factor: 24.094

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.