Monitoring the decontamination efficacy of the novel Poseidon-S disinfectant system in dental unit water lines.

Mari Fujita1, Izumi Mashima1, Futoshi Nakazawa2.   

Abstract

BACKGROUND: Contaminated dental unit waterlines (DUWLs) are a known source of specific health care-acquired infections because of the difficulty in keeping them clean during routine dental practice. Recently, an electrolysis apparatus that uses only the chlorine normally present in municipal water, the Poseidon-S system, was developed as a novel additive-free disinfectant system to control microbial contamination in DUWLs.
METHODS: The microbiological quality of water samples collected from DUWLs was assessed before and after installation of the Poseidon-S system in terms of the total viable counts (TVCs) of microorganisms. The microbicidal effects of the electrolyzed water against oral organisms and its cytotoxicity against human oral-derived cell lines were also examined.
RESULTS: Water samples from the DUWLs initially had average microbial TVCs of 103-106 colony-forming units (CFU)/mL. After installation of the Poseidon-S system, the number of microorganisms in the water samples decreased to less than 1 × 102 CFU/mL. The electrolyzed water also exhibited remarkable microbicidal effects on the microorganisms present in the DUWLs as well as microorganisms commonly isolated from human oral cavities, but showed low cytotoxicity towards human oral-derived cells.
CONCLUSION: This study demonstrated that routine use of the Poseidon-S system can effectively maintain low microbial levels in DUWLs.
Copyright © 2015. Published by Elsevier B.V.

Entities:  

Keywords:  cytotoxicity; dental unit water lines (DUWLs); disinfection system; electrolyzed water; microbial contamination

Mesh:

Substances:

Year:  2015        PMID: 26066542     DOI: 10.1016/j.jmii.2015.05.006

Source DB:  PubMed          Journal:  J Microbiol Immunol Infect        ISSN: 1684-1182            Impact factor:   4.399


  6 in total

1.  Evaluation of the simulator with automatic irrigation control system designed for countermeasures of internal contamination in dental unit water lines.

Authors:  Keisuke Okubo; Takashi Ito; Kentaro Okamoto; Ichiro Yamamoto; Hajime Mizutani; Yusuke Kawata; Yasuyoshi Shiota; Masahiro Ito; Shin Nakamura; Masako Tai; Tadashi Yamamoto; Shogo Takashiba
Journal:  Heliyon       Date:  2020-06-10

2.  Prevention and Control of Legionella and Pseudomonas spp. Colonization in Dental Units.

Authors:  Benedetta Tuvo; Michele Totaro; Maria Luisa Cristina; Anna Maria Spagnolo; David Di Cave; Sara Profeti; Angelo Baggiani; Gaetano Privitera; Beatrice Casini
Journal:  Pathogens       Date:  2020-04-21

3.  Evaluation of Microbiological and Free-Living Protozoa Contamination in Dental Unit Waterlines.

Authors:  Anna Maria Spagnolo; Marina Sartini; David Di Cave; Beatrice Casini; Benedetta Tuvo; Maria Luisa Cristina
Journal:  Int J Environ Res Public Health       Date:  2019-07-24       Impact factor: 3.390

4.  Effectiveness of a Water Disinfection Method Based on Osmosis and Chlorine Dioxide for the Prevention of Microbial Contamination in Dental Practices.

Authors:  Michele Totaro; Federica Badalucco; Francesca Papini; Niccolò Grassi; Marina Mannocci; Matteo Baggiani; Benedetta Tuvo; Beatrice Casini; Giovanni Battista Menchini Fabris; Angelo Baggiani
Journal:  Int J Environ Res Public Health       Date:  2022-08-24       Impact factor: 4.614

5.  Distinct Microbial Community of Accumulated Biofilm in Dental Unit Waterlines of Different Specialties.

Authors:  Cancan Fan; Haijing Gu; Limin Liu; Haiwei Zhu; Juan Yan; Yongbiao Huo
Journal:  Front Cell Infect Microbiol       Date:  2021-06-17       Impact factor: 5.293

6.  Influence of the acidity of electrolyzed water on the microhardness of inner layer dentin.

Authors:  Ker-Kong Chen; Ju-Hui Wu; Shin-I Wei; Je-Kang Du
Journal:  J Dent Sci       Date:  2019-11-07       Impact factor: 2.080

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.