| Literature DB >> 25993627 |
Soraia M Silva1, Fernanda I Corrêa1, Paula F C Silva1, Daniela F T Silva1, Paulo R G Lucareli1, João C F Corrêa1.
Abstract
BACKGROUND: Handgrip strength is currently considered a predictor of overall muscle strength and functional capacity. Therefore, it is important to find reliable and affordable instruments for this analysis, such as the modified sphygmomanometer test (MST).Entities:
Mesh:
Year: 2015 PMID: 25993627 PMCID: PMC4481834 DOI: 10.1590/bjpt-rbf.2014.0081
Source DB: PubMed Journal: Braz J Phys Ther ISSN: 1413-3555 Impact factor: 3.377
Figure 1.(A) sphygmomanometer in the original format; (B) inner sphygmomanometer cuff; (C) modified device (10X5X2 cm); (D) modified sphygmomanometer.
Demographic and clinical characteristics of the study subjects.
| Variable | CG elderly (n=26) | Parkinsonian (n=24) |
|---|---|---|
| Men | 11 (42%) | 10 (42%) |
| Women | 15 (58%) | 14 (58%) |
| Age (years) | 63.4 (7.2) | 65.5 (6.2) |
| BMI (kg/m2) | 22.5 (3.9) | 24.9 (2.0) |
| Right UL dominant | 19 (73%) | 18 (75%) |
| Left UL dominant | 7 (27%) | 6 (25%) |
| Hoehn and Yahr Classification | - - | 2 (1/3) |
| Time since PD diagnosed (years) | - - | 7.2 (4.1) |
| MST Right (mm Hg) | ||
| First evaluation | 55.95 (21.28) | 59.35 (20.34) |
| Second evaluation | 56.25 (21.02) | 67.43 (21.25) |
| Third evaluation | 60.22 (20.43) | 65.71 (14.64) |
| MST Left (mm Hg) | ||
| First evaluation | 57.15 (16.73) | 60.66 (18.35) |
| Second evaluation | 55.38 (15.97) | 70.91 (23.21) |
| Third evaluation | 66.15 (25.24) | 72.96 (17.66) |
CG: control group; PD: Parkinson's disease; BMI: body mass index; UL: upper limb; MST: modified sphygmomanometer test. The data are expressed as the frequency (percentage), median (interquartile range), or mean (standard deviation).
Pearson correlation coefficient and simple linear regression analysis between the modified sphygmomanometer test and the portable Jamar dynamometer.
| Correlation coefficient
( | Simple linear regression ( | Regression Equation | |
|---|---|---|---|
| HGS R (CG) | 0.67* | 0.41 | y=1.4562 + 2.1532x |
| HGS L (CG) | 0.65* | 0.43 | y=8.6356 + 1.9897x |
| HGS R (PD) | 0.68* | 0.46 | y=0.9996 + 1.7848x |
| HGS L (PD) | 0.45* | 0.20 | y=20.7755 + 1.3719x |
CG: control group; PD: Parkinson's disease; HGS: handgrip strength; R: right; L: left.
(P<0.05).
Reproducibility (reliability and agreement) of the modified sphygmomanometer test (MST).
| Reliability | Agreement | Agreement | |
|---|---|---|---|
| MST R (CG) | 0.79 (0.55-0.95) | 2.56 | 7.09 |
| MST L (CG) | 0.88 (0.75-0.95) | 2.29 | 6.34 |
| MST R (PD) | 0.89 (0.62-0.96) | 2.55 | 7.06 |
| MST L (PD) | 0.83 (0.50-0.95) | 2.67 | 7.40 |
MST: modified sphygmomanometer test; R: right; L: left. CG: control group; PD: Parkinson's disease; ICC: intraclass correlation coefficient; SEM: standard error of measurement; MDC: minimum detectable change.
Figure 2.Interrater agreement according to the Bland-Altman method. CG: control group; PD: Parkinson's disease; MST: Modified sphygmomanometer test; SD: standard deviation.