Field Evaluation of Capillary Blood Samples as a Collection Specimen for the Rapid Diagnosis of Ebola Virus Infection During an Outbreak Emergency.

Thomas Strecker1, Bernadett Palyi2, Heinz Ellerbrok3, Sylvie Jonckheere4, Hilde de Clerck4, Joseph Akoi Bore5, Martin Gabriel6, Kilian Stoecker7, Markus Eickmann1, Michel van Herp4, Pierre Formenty8, Antonino Di Caro9, Stephan Becker10.   

Abstract

BACKGROUND: Reliable reverse transcription polymerase chain reaction (RT-PCR)-based diagnosis of Ebola virus infection currently requires a blood sample obtained by intravenous puncture. During the current Ebola outbreak in Guinea, we evaluated the usability of capillary blood samples collected from fingersticks of patients suspected of having Ebola virus disease (EVD) for field diagnostics during an outbreak emergency.
METHODS: A total of 120 venous and capillary blood samples were collected from 53 patients admitted to the Ebola Treatment Centre in Guéckédou, Guinea, between July and August 2014. All sample specimens were analyzed by RT-PCR using the RealStar Filovirus Screen RT-PCR Kit 1.0 from altona Diagnostics (Germany). We compared samples obtained by venipuncture and those obtained by capillary blood sampling absorbed onto swab devices.
RESULTS: The resulting sensitivity and specificity of tests performed with capillary blood samples were 86.8% (95% confidence interval [CI], 71.9%-95.6%; 33/38 patients) and 100% (95% CI, 84.6%-100%; 22/22 patients), respectively.
CONCLUSIONS: Our data suggest that capillary blood samples could serve as an alternative to venous blood samples for the diagnosis of EVD in resource-limited settings during a crisis. This can be of particular advantage in cases when venipuncture is difficult to perform-for example, with newborns and infants or when adult patients reject venipuncture for cultural or religious reasons.
© The Author 2015. Published by Oxford University Press on behalf of the Infectious Diseases Society of America.

Entities:  

Keywords:  Ebola virus; Ebola virus disease; capillary blood; field diagnostic

Mesh:

Substances:

Year:  2015        PMID: 25991465      PMCID: PMC4530726          DOI: 10.1093/cid/civ397

Source DB:  PubMed          Journal:  Clin Infect Dis        ISSN: 1058-4838            Impact factor:   9.079


The current Ebola virus outbreak in West Africa was first described in March 2014 in Guinea [1]. Since then, Ebola has widely spread to the neighboring countries of Liberia and Sierra Leone. Although intense and widespread transmission of the current Ebola outbreak has been limited to these 3 countries, cases were also reported in Mali, Nigeria, and Senegal as well as in Europe and the United States [2, 3]. As of 11 March 2015, the World Health Organization (WHO) reported a total of 24 282 cases of Ebola virus disease (EVD) with 9976 deaths [4], making the ongoing outbreak in West Africa by far the geographically most widespread and most complex Ebola outbreak since Ebola virus was first discovered in 1976 [5]. Reported EVD case fatality rates vary from 31% to 74% in the main affected West African countries [6-11]. Effective outbreak containment measures include the strict isolation of EVD patients, which requires reliable diagnosis. Current methods to diagnose suspected Ebola virus infection include reverse transcription polymerase chain reaction (RT-PCR), antigen-capture enzyme-linked immunosorbent assay (ELISA), and immunoglobulin M (IgM) and immunoglobulin G (IgG) ELISA [12-19]. While detection of viral RNA or viral antigen has been proven effective for diagnosis of Ebola virus infections from the early until late stage of illness, serological detection of specific IgM and IgG antibodies can only be used for diagnosis of EVD at the late stage in the acute phase of the disease and during convalescence. All current techniques for the reliable laboratory diagnosis of Ebola virus infection require whole blood, plasma, or serum collected by venipuncture. In addition, the use of oral fluid specimens has been suggested as a noninvasive, sensitive method for the detection of Ebola virus infections during outbreak investigations [20]. Venous blood sampling requires trained and skilled medical personnel, bears high risk for needlestick injuries, and is especially difficult to perform with newborns and infants. Moreover, patients in African settings often refuse venipuncture due to cultural and religious beliefs. To overcome these problems, we evaluated the suitability of capillary blood sampling from fingersticks in the diagnosis of EVD. The main objective of this study was to compare the sensitivity and specificity of tests that used capillary blood samples with samples obtained by venous puncture during an outbreak emergency. Furthermore, we assessed in a limited study the stability of capillary blood swab samples when stored at room temperature vs 4°C.

MATERIALS AND METHODS

Ethical Approval

The National Committee of Ethics in Medical Research of Guinea approved the use of diagnostic leftover samples for this study (permit number 11/CNERS/14). As the samples had been collected as part of the public health response to contain the outbreak in Guinea, informed consent was not obtained from patients.

Sample Collection

The study was conducted during the current outbreak of EVD in Guinea between July and August 2014. The study included 53 patients presenting with symptoms compatible with the WHO case definition for EVD [21]. Clinical samples were collected by Médecins sans Frontières (MSF) medical health personnel at the Ebola Treatment Centre (ETC) in Guéckédou, Guinea, wearing personal protective equipment following MSF safety guidelines. For each patient, whole blood obtained by venipuncture and capillary blood samples taken from fingersticks were collected at the same time. Venipuncture whole blood samples were collected using 6.0 mL BD Vacutainer collection tubes containing ethylenediaminetetraacetic acid (EDTA) (Becton Dickinson, East Rutherford, New Jersey). Capillary whole blood samples were collected from fingerpricks using BD Microtainer contact activated lancets, type 21G (Becton Dickinson, East Rutherford, New Jersey). Collected capillary blood was adsorbed on viscose swab collection applicators (Sarstedt, Germany) and immediately placed into a labeled sterile collection tube without viral transport medium. Samples were stored at room temperature (the mean temperature in Guéckédou is 23.8°C in July and August) until they were transferred to the laboratory for analysis. All samples were packaged with a triple packaging system following WHO safety recommendations for sample shipment. All sample tubes, plastic containers, and sealed plastic bags used for transport were surface disinfected with a 0.5% hypochlorite solution, and transported to the European Mobile Laboratory in Guéckédou, where they arrived on average within 1 hour after collection. Matching venipuncture blood sample and fingerstick sample specimen from the same patient were labeled with the same unique laboratory identification number and an additional mark for capillary blood samples. All sample specimens were subjected to the same handling and storage conditions. One hundred twenty venous and capillary blood samples from 53 patients with suspected EVD were obtained.

Laboratory Tests

Infectious sample specimens were handled in the European Mobile Laboratory in Guéckédou, Guinea, by qualified and trained laboratory personnel using a mobile safety glovebox cabinet, equipped with P3 and ABC filters and operated with negative pressure. All venous blood samples and capillary blood samples were examined for detection of Ebola virus RNA by RT-PCR using RealStar Filovirus Screen RT-PCR Kit 1.0 (altona Diagnostics, Germany) [22], specifically targeting the L gene as described previously [15]. Negative and positive controls were included in every batch of samples analyzed. To ensure the integrity of Ebola virus–specific real-time RT-PCR results by indicating potential RT-PCR inhibition, an internal control was analyzed in parallel for each patient sample. RT-PCR was performed using the SmartCycler technology (Cepheid). To this end, aliquots (140 µL) from each sample were inactivated by adding 560 µL of the guanidine thiocyanate lysis buffer AVL. Following 10 minutes of incubation, 560 µL of absolute ethanol was added to the aliquots. All sample tubes were surface disinfected using 0.5% hypochlorite solution for 5 minutes prior to release from the glovebox. For RT-PCR analysis of venipuncture blood samples, 50 µL of EDTA whole blood (plus 90 µL of nuclease-free water; Qiagen, Germany) or 140 µL plasma was used for RNA extraction using the QIAmp Viral RNA kit (Qiagen, Germany). Plasma was obtained by centrifugation of EDTA whole blood at 3500 rpm for 10 minutes. Comparison of Ebola viral loads using RT-PCR in 50 µL of EDTA whole blood or 140 µL plasma revealed similar results (unpublished field data). Swabs were swirled in 300 µL of nuclease-free water (Qiagen, Germany) in a 1.5-mL Eppendorf tube to redissolve biological material, and 140 µL of supernatant was used for RNA extraction with the QIAmp Viral RNA kit. Extraction and purification of RNA from venous and capillary blood samples was done according to the manufacturer's instructions with an additional second wash step using AW2 buffer. For differential diagnosis, admitted patients in the ETC were tested for malaria infections using venous EDTA whole blood for the qualitative detection of Plasmodium species antigens in a commercially available in vitro immunochromatographic assay using BinaxNOW Malaria (Alere).

Statistical Analysis

To determine sensitivity and specificity of the capillary blood sample tests, RT-PCR results were compared to those obtained with venous blood samples. Sensitivity and specificity of the tests were calculated using a 2-by-2 contingency table.

RESULTS

During the Ebola outbreak in Guinea, 53 Guinean patients presenting with clinical symptoms of EVD were included in the study. Among the 53 patients, 18 (33.9%) were men and 35 (66.1%) were women. The mean age was 27.4 years (range, 1–80 years). Venous and capillary blood samples were collected simultaneously from each patient and analyzed for the presence of Ebola virus RNA by real-time RT-PCR. Of the 60 capillary blood samples analyzed, 33 tested positive and 27 negative for EVD. For the 60 corresponding venous blood samples, 38 tested positive and 22 negative for EVD. Identical results between capillary and venous samples were observed for 55 samples, while 5 samples revealed false-negative results using capillary blood samples. For the 55 capillary blood samples with identical results, 33 tested EVD positive and 22 tested EVD negative. In our study, no false-positive results were observed using capillary blood samples as a diagnostic specimen. The sensitivity and the specificity of tests performed with capillary blood samples absorbed onto swab devices compared to tests performed with venous blood samples were 86.8% (95% confidence interval [CI], 71.9%–95.6%; 33/38 patients) and 100% (95% CI, 84.6%–100%; 22/22 patients), respectively (Table 1). Table 2 summarizes the comparative analysis of RT-PCR–based EVD detection in venous blood and capillary blood samples according to the time of sampling after onset of clinical symptoms. With the exception of 1 sample pair, pairwise comparison of venous and corresponding capillary blood samples revealed lower levels of Ebola RNA in capillary blood samples, regardless of the time to sampling after symptom onset. False-negative RT-PCR results caused by RT-PCR inhibition have been reported for patients with viral hemorrhagic fevers [23]. However, analysis of cycle threshold (Ct) values for the corresponding internal control for each individual patient sample indicated that a higher Ct value or even a negative RT-PCR result for Ebola virus–specific RNA in capillary blood samples was not due to severe RT-PCR inhibition (Supplementary Tables 1 and 2, Supplementary Figure 1).
Table 1.

Sensitivity and Specificity of Tests Performed on Capillary Blood Specimens for the Diagnosis of Ebola Virus Infection

Ebola Virus Detection With Capillary Blood SamplesEbola Virus Detection With Standard Venous Blood Samples
PositiveNegativeTotal
Positive33033
Negative52227
Total382260

Data are No. of samples. Results of Ebola virus detection in capillary fingerstick samples were compared with samples obtained by venous blood puncture using real-time reverse transcription polymerase chain reaction. The sensitivity for detection of Ebola virus in capillary blood samples was 86.8% (95% confidence interval [CI], 71.9%–95.6%), the specificity was 100% (95% CI, 84.6%–100%), and the negative likelihood ratio was 0.13 (95% CI, .06–.30).

Table 2.

Comparison of Ebola Virus Load Between Venous Blood and Capillary Blood Samples Using Real-time Reverse Transcription Polymerase Chain Reaction

No. of Tested SamplePatient No.Day of Sampling After Symptom OnsetEVD Detection in Venous Blood (Ct Value)Semiquantitative Viral Load ResultaEVD Detection in Capillary Blood (Ct Value)Semiquantitative Viral Load ResultbCt Difference Between Venous and Capillary Blood Samplesc
1Patient 1Day 1Positive (22.47)+++Positive (23.44)++++0.97
2Patient 2Day 1Positive (15.49)+++Positive (18.42)++++2.93
3Patient 3Day 1Positive (25.59)++Positive (29.26)+++3.67
4Patient 4Day 2Positive (20.31)+++Positive (28.28)+++7.97
5Patient 5Day 2Positive (21.38)+++Positive (26.39)+++5.01
6Patient 6Day 2Positive (26.03)++Positive (30.47)+++4.44
7Patient 7Day 2Positive (30.48)++Negative (−)NA
8Patient 8Day 3Positive (16.42)+++Positive (18.27)++++1.85
9Patient 9Day 3Positive (18.10)+++Positive (19.65)++++1.55
10Patient 10Day 3Positive (28.08)++Positive (21.01)+++−7.07
11Patient 11Day 3Positive (23.39)+++Positive (26.46)+++3.07
12Patient 12Day 3Positive (18.77)+++Positive (21.45)++++2.68
13Patient 13Day 3Positive (23.94)+++Positive (28.65)+++4.71
14Patient 14Day 3Positive (24.32)+++Positive (27.26)+++2.94
15Patient 15Day 3Positive (22.64)+++Positive (26.42)+++3.78
16Patient 16Day 3Positive (20.74)+++Positive (25.32)+++4.58
17Patient 17Day 3Positive (19.48)+++Positive (20.60)++++1.12
18Patient 18Day 3Positive (26.17)++Negative (−)NA
19Patient 19Day 3Positive (27.03)++Negative (−)NA
20Patient 20Day 4Positive (15.31)+++Positive (20.47)++++5.16
21Patient 18Day 4Positive (22.23)+++Positive (28.41)+++6.18
22Patient 21Day 4Positive (20.19)+++Positive (25.87)+++5.68
23Patient 22Day 4Positive (21.46)+++Positive (24.36)++++2.90
24Patient 23Day 4Positive (17.71)+++Positive (20.29)++++2.58
25Patient 24Day 5Positive (19.01)+++Positive (26.12)+++7.11
26Patient 25Day 5Positive (21.12)+++Positive (23.74)++++2.62
27Patient 4Day 5Positive (21.02)+++Positive (28.10)+++7.08
28Patient 26Day 5Positive (21.99)+++Positive (23.72)++++1.73
29Patient 27Day 6Positive (17.06)+++Positive (21.33)++++4.27
30Patient 28Day 6Positive (21.84)+++Positive (27.45)+++5.61
31Patient 29Day 6Positive (22.04)+++Positive (28.05)+++6.01
32Patient 30Day 6Positive (21.74)+++Positive (26.00)+++4.26
33Patient 18Day 7Positive (16.00)+++Positive (17.00)++++1.00
34Patient 31Day 9Positive (25.04)++Positive (30.01)+++4.97
35Patient 32Day 15Positive (31.77)++Negative (−)NA
36Patient 33Day 17Positive (17.04)+++Positive (19.43)++++2.39
37Patient 34Day 18Positive (24.35)+++Positive (25.60)+++1.25
38Patient 35Not knownPositive (30.61)++Negative (−)NA

Abbreviations: Ct, cycle threshold; EVD, Ebola virus disease; NA, not applicable.

a Ct values are classified in subsequent categories of 0–24 (highly positive), 25–34 (positive), and 35–40 (weak positive) and correspond with +++, ++, and + results, respectively.

b Negative real-time reverse transcription polymerase chain reaction result is indicated by “–.”

c For Ct differences, “+” indicates higher viral load in venous blood sample, whereas “–” indicates higher viral load in capillary blood sample.

Sensitivity and Specificity of Tests Performed on Capillary Blood Specimens for the Diagnosis of Ebola Virus Infection Data are No. of samples. Results of Ebola virus detection in capillary fingerstick samples were compared with samples obtained by venous blood puncture using real-time reverse transcription polymerase chain reaction. The sensitivity for detection of Ebola virus in capillary blood samples was 86.8% (95% confidence interval [CI], 71.9%–95.6%), the specificity was 100% (95% CI, 84.6%–100%), and the negative likelihood ratio was 0.13 (95% CI, .06–.30). Comparison of Ebola Virus Load Between Venous Blood and Capillary Blood Samples Using Real-time Reverse Transcription Polymerase Chain Reaction Abbreviations: Ct, cycle threshold; EVD, Ebola virus disease; NA, not applicable. a Ct values are classified in subsequent categories of 0–24 (highly positive), 25–34 (positive), and 35–40 (weak positive) and correspond with +++, ++, and + results, respectively. b Negative real-time reverse transcription polymerase chain reaction result is indicated by “–.” c For Ct differences, “+” indicates higher viral load in venous blood sample, whereas “–” indicates higher viral load in capillary blood sample. The discrepancy in Ct values may stem from differences in recovery efficiency of viral RNA between both sample specimen types. Alternatively, differences in viral load levels in capillary and venous blood may account for the observed Ct variations, particularly in the early phase of symptomatic patients. Sequential sampling of an individual patient allowed us to monitor Ebola viral load levels in venous and capillary blood samples during the course of clinical illness (Table 3). This patient, suspected of having EVD based on clinical symptoms, was admitted 1 day after the onset of symptoms to the ETC. At the day of admission, laboratory results were negative for EVD in both sample types. RT-PCR testing was repeated 48 hours later and the patient tested positive for Ebola RNA in the venous blood sample but remained Ebola RNA negative in the capillary blood sample. On day 4, both sample types tested EVD positive, with the venous blood sample showing a higher viral load compared with the capillary blood sample. On day 7, EVD-positive test results revealed similar Ct values for capillary blood and venous blood, indicating comparable viral load. These data indicate potential differences in viremia between venous and capillary blood according to time of disease onset. This might also explain the 2 additional discordant results between venous blood and capillary blood from day 2 and day 3 after onset of symptoms (Table 2, samples 7 and 19). Yet with low numbers of patients, further studies are needed to compare the kinetics of viral load in venous and capillary blood during the course of infection.
Table 3.

Real-time Reverse Transcription Polymerase Chain Reaction Results From an Individual Patient in the Course of a Follow-up Study During His Stay at the Ebola Treatment Center, Guéckédou, Guinea, August 2014

Day of Sampling After Symptom OnsetEVD Detection in Venous Blood (Ct Value)Semiquantitative Viral Load ResultaEVD Detection in Capillary Blood (Ct Value)Semiquantitative Viral Load Result
Day 1Negative (−)Negative (−)
Day 3Positive (26.17)++Negative (−)
Day 4Positive (22.23)+++Positive (28.41)++
Day 7Positive (16.00)+++Positive (17.00)+++

Abbreviations: Ct, cycle threshold; EVD, Ebola virus disease.

a Ct values are classified in subsequent categories of 0–24 (highly positive), 25–34 (positive), and 35–40 (weak positive) and correspond with +++, ++, and + results, respectively. Negative real-time reverse transcription polymerase chain reaction result is indicated by “−.”

Real-time Reverse Transcription Polymerase Chain Reaction Results From an Individual Patient in the Course of a Follow-up Study During His Stay at the Ebola Treatment Center, Guéckédou, Guinea, August 2014 Abbreviations: Ct, cycle threshold; EVD, Ebola virus disease. a Ct values are classified in subsequent categories of 0–24 (highly positive), 25–34 (positive), and 35–40 (weak positive) and correspond with +++, ++, and + results, respectively. Negative real-time reverse transcription polymerase chain reaction result is indicated by “−.” For differential diagnostic, 52 out of 53 newly admitted patients were tested for malaria infections (Plasmodium species) using an antigen-based rapid diagnostic test. Among the patients who tested positive for EVD, 13 patients scored positive and 21 patients scored negative for malaria. For the patients who tested EVD negative, 8 patients scored positive and 10 patients scored negative for malaria. Comparison of results between matched capillary blood and venous blood samples and malaria coinfections showed that malaria coinfections had no influence on the sensitive and specific detection of EVD when using capillary blood samples (data not shown). Due to the favorable location in Guéckédou with the European Mobile Laboratory adjacent to the MSF ETC, blood samples collected from patients, including the capillary blood specimens, could be treated for laboratory diagnosis within 1 hour to a few hours. However, in other settings, timely delay between specimen collection and laboratory analysis might be much greater. In cases where EVD diagnostics can not be performed within 24 hours after sample collection, the WHO recommends a storage temperature between 0°C and 5°C for sample specimens to maintain sample preservation [24]. In a limited test, we addressed sample stability under field conditions and observed rapid sample degradation when capillary blood swab samples were stored for >24 hours at room temperature compared with corresponding samples stored at 4°C, indicating an impact of storage conditions on sensitive Ebola nucleic acid detection by RT-PCR using capillary blood swab samples (data not shown).

DISCUSSION

Efficient interruption of Ebola virus transmission chains critically depends on reliable and rapid laboratory diagnosis of patients suspected of having EVD. The laboratory results are needed to confirm suspected cases and to execute subsequent isolation measures. The high numbers of patients with suspected EVD admitted to ETCs in the affected countries Liberia, Sierra Leone, and Guinea often requires large-scale sampling of patients. A major challenge is sampling of venous blood in patients with suspected EVD, posing a serious risk to healthcare workers due to accidental needlestick injuries [25]. The less invasive collection of capillary blood samples could be an alternative approach, strongly reducing the risk for medical personnel. In addition, capillary blood samples might be useful for monitoring viral load during the course of disease where repeated sampling by venous puncture would be too much of a strain for severely suffering EVD patients. The aim of this study was therefore to evaluate the feasibility of using capillary blood samples from fingertips for detection of Ebola virus and to compare the results with blood samples obtained by venipuncture. Analyzing capillary blood specimens collected from 53 patients with suspected EVD, we demonstrate for the first time the applicability of blood samples derived from fingersticks for the detection of EVD by RT-PCR. As shown in Table 2, pairwise comparison of venous blood samples and capillary blood samples revealed in general a good correlation of test results for patients who were found to be positive for EVD. Compared to venous blood samples, capillary blood samples from fingerpricks revealed a sensitivity of 86.8% (95% confidence interval [CI], 71.9%–95.6%; 33/38 patients) and a specificity of 100% (95% CI, 84.6%–100%; 22/22 patients), respectively. Our findings thus support the use of capillary blood samples as an alternative diagnostic specimen to venous blood samples in an outbreak situation. Due to the observed 100% specificity of test results in our field study, we consider a positive EVD laboratory result from capillary blood samples valid for reliable diagnostic evaluation. Capillary blood sampling may therefore represent a suitable diagnostic approach for initial rapid screening of a large number of patients in the midst of an outbreak, allowing accelerated EVD patient management and isolation measures. Suspected cases testing negative for EVD using capillary blood samples should be additionally tested using venous blood samples, the current gold standard specimen for RT-PCR–based EVD diagnosis. Similarly, at present, venous-derived blood is recommended for testing patients during the convalescent phase to enhance decision making on their discharge and reintroduction in the community. Those cases when capillary blood samples were less sensitive or tested negative compared with venous blood samples may relate to the stage of disease. At present, no information is available regarding Ebola viral load kinetics in capillary blood during the course of infection. Future studies should address the viremia in the capillary blood according to the stage of clinical illness to ensure valid interpretation of diagnostic results. Alternatively, the differences in viral load between the 2 specimen types might be due to inherent sampling and collection procedures (Supplementary Figure 2). Although introducing identical blood volumes (eg, from whole blood vacutainers into sample preparation) is easy to achieve, adsorption and desorption of capillary blood from swab fingerpricks might vary, thus explaining the observed differences in Ct values. Further studies should address standardization of collection procedures and sampling volumes for the swab approach. Finally, in this study we compared EDTA-anticoagulated blood collected by venipuncture and capillary blood samples that were deposited onto viscose swab devices, exhibiting no anticoagulation activity. Protective effects of EDTA on stability of pathogen-derived RNA in blood samples have been reported [26]. Despite this fact, we haven chosen these swab devices for our comparison study, as the European Mobile Laboratory units in Guéckédou, Guinea, and Foya, Liberia, received a considerable number of capillary blood samples absorbed onto viscose swabs as a diagnostic sample specimen from both infants and adults for EVD testing (unpublished observation). Although these types of swabs are routinely used for sensitive detection of Ebola virus in oral fluids in the current EVD outbreak in West Africa [27, 28], at present very little is known about the stability of Ebola virus RNA or potential degradation effects in capillary blood samples absorbed onto swabs. Rapid RNA degradation might result in lower detection sensitivity or even render a test falsely negative. Thus, the value of swabs as a collection and storage device for capillary blood samples for diagnostic purposes needs to be further evaluated and compared with capillary blood samples collected into a microcollection container that has EDTA as an anticoagulant. Capillary blood sampling offers some advantages over blood sampling by venipuncture. Sampling can be done by medical health personnel other than experienced and trained phlebotomists. Furthermore, for patient convenience and in some circumstances for cultural reasons and religious beliefs, this procedure may be more acceptable than venous blood draw. Thus, the less invasive collection of samples eliminates the anxiety associated with venipuncture and may lead to higher acceptance of blood testing in settings where cultural traditions otherwise might interfere with effective outbreak management. Moreover, capillary blood collection is the preferred method of blood specimen collection for newborns and infants. These practical reasons might outbalance the higher sensitivity observed in some cases for venous blood analysis, particularly in the critical situation of an outbreak. Reliable detection and laboratory confirmation of Ebola virus is key to successfully controlling the current EVD outbreak in West Africa. The WHO has identified an ongoing need to develop sensitive and specific tests for diagnosis of EVD using accessible point-of-care (POC) technologies, including POC fingerstick tests that are capable of providing early diagnosis of EVD and Ebola viral load measurement. Our report is the first study that compares the sensitivity of RT-PCR–based EVD tests on capillary vs venous blood samples. Although a more extensive evaluation of capillary blood specimens under field conditions is still needed, our results point toward the use of capillary blood samples as an appropriate collection specimen for sensitive and specific diagnosis of EVD. In conclusion, this study demonstrates the value of capillary fingerstick blood as a clinical sample specimen for virologic diagnosis of Ebola virus infection during an outbreak emergency.

Supplementary Data

Supplementary materials are available at Clinical Infectious Diseases online (http://cid.oxfordjournals.org). Supplementary materials consist of data provided by the author that are published to benefit the reader. The posted materials are not copyedited. The contents of all supplementary data are the sole responsibility of the authors. Questions or messages regarding errors should be addressed to the author.
  24 in total

Review 1.  Laboratory diagnostic systems for Ebola and Marburg hemorrhagic fevers developed with recombinant proteins.

Authors:  Masayuki Saijo; Masahiro Niikura; Tetsuro Ikegami; Ichiro Kurane; Takeshi Kurata; Shigeru Morikawa
Journal:  Clin Vaccine Immunol       Date:  2006-04

2.  The first case of Ebola virus disease acquired outside Africa.

Authors:  J Manuel Parra; Octavio J Salmerón; María Velasco
Journal:  N Engl J Med       Date:  2014-11-19       Impact factor: 91.245

3.  Short-term stability of pathogen-specific nucleic acid targets in clinical samples.

Authors:  Mohammad R Hasan; Rusung Tan; Ghada N Al-Rawahi; Eva Thomas; Peter Tilley
Journal:  J Clin Microbiol       Date:  2012-10-10       Impact factor: 5.948

4.  Case fatality rate for Ebola virus disease in west Africa.

Authors:  Adam J Kucharski; W John Edmunds
Journal:  Lancet       Date:  2014-09-23       Impact factor: 79.321

5.  Emergence of Zaire Ebola virus disease in Guinea.

Authors:  Sylvain Baize; Delphine Pannetier; Lisa Oestereich; Toni Rieger; Lamine Koivogui; N'Faly Magassouba; Barrè Soropogui; Mamadou Saliou Sow; Sakoba Keïta; Hilde De Clerck; Amanda Tiffany; Gemma Dominguez; Mathieu Loua; Alexis Traoré; Moussa Kolié; Emmanuel Roland Malano; Emmanuel Heleze; Anne Bocquin; Stephane Mély; Hervé Raoul; Valérie Caro; Dániel Cadar; Martin Gabriel; Meike Pahlmann; Dennis Tappe; Jonas Schmidt-Chanasit; Benido Impouma; Abdoul Karim Diallo; Pierre Formenty; Michel Van Herp; Stephan Günther
Journal:  N Engl J Med       Date:  2014-04-16       Impact factor: 91.245

6.  Management of pregnant women infected with Ebola virus in a treatment centre in Guinea, June 2014.

Authors:  F M Baggi; A Taybi; A Kurth; M Van Herp; A Di Caro; R Wölfel; S Günther; T Decroo; H Declerck; S Jonckheere
Journal:  Euro Surveill       Date:  2014-12-11

7.  Clinical presentation of patients with Ebola virus disease in Conakry, Guinea.

Authors:  Elhadj Ibrahima Bah; Marie-Claire Lamah; Tom Fletcher; Shevin T Jacob; David M Brett-Major; Amadou Alpha Sall; Nahoko Shindo; William A Fischer; Francois Lamontagne; Sow Mamadou Saliou; Daniel G Bausch; Barry Moumié; Tim Jagatic; Armand Sprecher; James V Lawler; Thierry Mayet; Frederique A Jacquerioz; María F Méndez Baggi; Constanza Vallenas; Christophe Clement; Simon Mardel; Ousmane Faye; Oumar Faye; Baré Soropogui; Nfaly Magassouba; Lamine Koivogui; Ruxandra Pinto; Robert A Fowler
Journal:  N Engl J Med       Date:  2014-11-05       Impact factor: 91.245

8.  Detection of Ebola virus in oral fluid specimens during outbreaks of Ebola virus hemorrhagic fever in the Republic of Congo.

Authors:  Pierre Formenty; Eric Maurice Leroy; Alain Epelboin; Francois Libama; Marco Lenzi; Hinrich Sudeck; Philippe Yaba; Yokouidé Allarangar; Paul Boumandouki; Virginot Blad Nkounkou; Christian Drosten; Allen Grolla; Heinz Feldmann; Cathy Roth
Journal:  Clin Infect Dis       Date:  2006-04-26       Impact factor: 9.079

9.  Evidence for a decrease in transmission of Ebola virus--Lofa County, Liberia, June 8-November 1, 2014.

Authors:  Aditya Sharma; Nico Heijenberg; Clement Peter; Josephus Bolongei; Bruce Reeder; Tamba Alpha; Esther Sterk; Hugues Robert; Andreas Kurth; Angela Cannas; Anne Bocquin; Thomas Strecker; Christopher Logue; Antonino Di Caro; Thomas Pottage; Constanze Yue; Kilian Stoecker; Roman Wölfel; Martin Gabriel; Stephan Günther; Inger Damon
Journal:  MMWR Morb Mortal Wkly Rep       Date:  2014-11-21       Impact factor: 17.586

10.  Ebola virus disease in West Africa--the first 9 months of the epidemic and forward projections.

Authors:  Bruce Aylward; Philippe Barboza; Luke Bawo; Eric Bertherat; Pepe Bilivogui; Isobel Blake; Rick Brennan; Sylvie Briand; Jethro Magwati Chakauya; Kennedy Chitala; Roland M Conteh; Anne Cori; Alice Croisier; Jean-Marie Dangou; Boubacar Diallo; Christl A Donnelly; Christopher Dye; Tim Eckmanns; Neil M Ferguson; Pierre Formenty; Caroline Fuhrer; Keiji Fukuda; Tini Garske; Alex Gasasira; Stephen Gbanyan; Peter Graaff; Emmanuel Heleze; Amara Jambai; Thibaut Jombart; Francis Kasolo; Albert Mbule Kadiobo; Sakoba Keita; Daniel Kertesz; Moussa Koné; Chris Lane; Jered Markoff; Moses Massaquoi; Harriet Mills; John Mike Mulba; Emmanuel Musa; Joel Myhre; Abdusalam Nasidi; Eric Nilles; Pierre Nouvellet; Deo Nshimirimana; Isabelle Nuttall; Tolbert Nyenswah; Olushayo Olu; Scott Pendergast; William Perea; Jonathan Polonsky; Steven Riley; Olivier Ronveaux; Keita Sakoba; Ravi Santhana Gopala Krishnan; Mikiko Senga; Faisal Shuaib; Maria D Van Kerkhove; Rui Vaz; Niluka Wijekoon Kannangarage; Zabulon Yoti
Journal:  N Engl J Med       Date:  2014-09-22       Impact factor: 91.245

View more
  13 in total

1.  Molecular Diagnostic Field Test for Point-of-Care Detection of Ebola Virus Directly From Blood.

Authors:  Jason W Benzine; Kerry M Brown; Krystle N Agans; Ronald Godiska; Chad E Mire; Krishne Gowda; Brandon Converse; Thomas W Geisbert; David A Mead; Yogesh Chander
Journal:  J Infect Dis       Date:  2016-09-16       Impact factor: 5.226

Review 2.  The Point-of-Care Laboratory in Clinical Microbiology.

Authors:  Michel Drancourt; Audrey Michel-Lepage; Sylvie Boyer; Didier Raoult
Journal:  Clin Microbiol Rev       Date:  2016-07       Impact factor: 26.132

Review 3.  Ebola virus disease and pregnancy: A review of the current knowledge of Ebola virus pathogenesis, maternal, and neonatal outcomes.

Authors:  Lisa M Bebell; Titilope Oduyebo; Laura E Riley
Journal:  Birth Defects Res       Date:  2017-03-15       Impact factor: 2.344

4.  Rapid Bedside Inactivation of Ebola Virus for Safe Nucleic Acid Tests.

Authors:  Maiken Worsøe Rosenstierne; Helen Karlberg; Karoline Bragstad; Gunnel Lindegren; Malin Lundahl Stoltz; Cristiano Salata; Anne-Marte Bakken Kran; Susanne Gjeruldsen Dudman; Ali Mirazimi; Anders Fomsgaard
Journal:  J Clin Microbiol       Date:  2016-07-27       Impact factor: 5.948

Review 5.  Presence and Persistence of Ebola or Marburg Virus in Patients and Survivors: A Rapid Systematic Review.

Authors:  Julii Brainard; Katherine Pond; Lee Hooper; Kelly Edmunds; Paul Hunter
Journal:  PLoS Negl Trop Dis       Date:  2016-02-29

6.  Operational Research during the Ebola Emergency.

Authors:  Gabriel Fitzpatrick; Tom Decroo; Bertrand Draguez; Rosa Crestani; Axelle Ronsse; Rafael Van den Bergh; Michel Van Herp
Journal:  Emerg Infect Dis       Date:  2017-07       Impact factor: 6.883

7.  External quality assessment study for ebolavirus PCR-diagnostic promotes international preparedness during the 2014 - 2016 Ebola outbreak in West Africa.

Authors:  Heinz Ellerbrok; Sonja Jacobsen; Pranav Patel; Toni Rieger; Markus Eickmann; Stephan Becker; Stephan Günther; Dhamari Naidoo; Livia Schrick; Kathrin Keeren; Angelina Targosz; Anette Teichmann; Pierre Formenty; Matthias Niedrig
Journal:  PLoS Negl Trop Dis       Date:  2017-05-01

8.  Determining the effect of different environmental conditions on Ebola virus viability in clinically relevant specimens.

Authors:  Bernadett Palyi; Nora Magyar; Judit Henczko; Balint Szalai; Agnes Farkas; Thomas Strecker; Maria Takacs; Zoltan Kis
Journal:  Emerg Microbes Infect       Date:  2018-03-29       Impact factor: 7.163

9.  Performance of the GeneXpert Ebola Assay for Diagnosis of Ebola Virus Disease in Sierra Leone: A Field Evaluation Study.

Authors:  Amanda E Semper; M Jana Broadhurst; Jade Richards; Geraldine M Foster; Andrew J H Simpson; Christopher H Logue; J Daniel Kelly; Ann Miller; Tim J G Brooks; Megan Murray; Nira R Pollock
Journal:  PLoS Med       Date:  2016-03-29       Impact factor: 11.069

10.  The role of rapid diagnostics in managing Ebola epidemics.

Authors:  Pierre Nouvellet; Tini Garske; Harriet L Mills; Gemma Nedjati-Gilani; Wes Hinsley; Isobel M Blake; Maria D Van Kerkhove; Anne Cori; Ilaria Dorigatti; Thibaut Jombart; Steven Riley; Christophe Fraser; Christl A Donnelly; Neil M Ferguson
Journal:  Nature       Date:  2015-12-03       Impact factor: 49.962

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.