Variability in CT lung-nodule volumetry: Effects of dose reduction and reconstruction methods.

Stefano Young1, Hyun J Grace Kim1, Moe Moe Ko1, War War Ko1, Carlos Flores1, Michael F McNitt-Gray1.   

Abstract

PURPOSE: Measuring the size of nodules on chest CT is important for lung cancer staging and measuring therapy response. 3D volumetry has been proposed as a more robust alternative to 1D and 2D sizing methods. There have also been substantial advances in methods to reduce radiation dose in CT. The purpose of this work was to investigate the effect of dose reduction and reconstruction methods on variability in 3D lung-nodule volumetry.
METHODS: Reduced-dose CT scans were simulated by applying a noise-addition tool to the raw (sinogram) data from clinically indicated patient scans acquired on a multidetector-row CT scanner (Definition Flash, Siemens Healthcare). Scans were simulated at 25%, 10%, and 3% of the dose of their clinical protocol (CTDIvol of 20.9 mGy), corresponding to CTDIvol values of 5.2, 2.1, and 0.6 mGy. Simulated reduced-dose data were reconstructed with both conventional filtered backprojection (B45 kernel) and iterative reconstruction methods (SAFIRE: I44 strength 3 and I50 strength 3). Three lab technologist readers contoured "measurable" nodules in 33 patients under each of the different acquisition/reconstruction conditions in a blinded study design. Of the 33 measurable nodules, 17 were used to estimate repeatability with their clinical reference protocol, as well as interdose and inter-reconstruction-method reproducibilities. The authors compared the resulting distributions of proportional differences across dose and reconstruction methods by analyzing their means, standard deviations (SDs), and t-test and F-test results.
RESULTS: The clinical-dose repeatability experiment yielded a mean proportional difference of 1.1% and SD of 5.5%. The interdose reproducibility experiments gave mean differences ranging from -5.6% to -1.7% and SDs ranging from 6.3% to 9.9%. The inter-reconstruction-method reproducibility experiments gave mean differences of 2.0% (I44 strength 3) and -0.3% (I50 strength 3), and SDs were identical at 7.3%. For the subset of repeatability cases, inter-reconstruction-method mean/SD pairs were (1.4%, 6.3%) and (-0.7%, 7.2%) for I44 strength 3 and I50 strength 3, respectively. Analysis of representative nodules confirmed that reader variability appeared unaffected by dose or reconstruction method.
CONCLUSIONS: Lung-nodule volumetry was extremely robust to the radiation-dose level, down to the minimum scanner-supported dose settings. In addition, volumetry was robust to the reconstruction methods used in this study, which included both conventional filtered backprojection and iterative methods.

Entities:  

Mesh:

Year:  2015        PMID: 25979066      PMCID: PMC5148179          DOI: 10.1118/1.4918919

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  23 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  Evaluation of lung MDCT nodule annotation across radiologists and methods.

Authors:  Charles R Meyer; Timothy D Johnson; Geoffrey McLennan; Denise R Aberle; Ella A Kazerooni; Heber Macmahon; Brian F Mullan; David F Yankelevitz; Edwin J R van Beek; Samuel G Armato; Michael F McNitt-Gray; Anthony P Reeves; David Gur; Claudia I Henschke; Eric A Hoffman; Peyton H Bland; Gary Laderach; Richie Pais; David Qing; Chris Piker; Junfeng Guo; Adam Starkey; Daniel Max; Barbara Y Croft; Laurence P Clarke
Journal:  Acad Radiol       Date:  2006-10       Impact factor: 3.173

3.  Validation of CT dose-reduction simulation.

Authors:  Parinaz Massoumzadeh; Steven Don; Charles F Hildebolt; Kyongtae T Bae; Bruce R Whiting
Journal:  Med Phys       Date:  2009-01       Impact factor: 4.071

Review 4.  Noncalcified lung nodules: volumetric assessment with thoracic CT.

Authors:  Marios A Gavrielides; Lisa M Kinnard; Kyle J Myers; Nicholas Petrick
Journal:  Radiology       Date:  2009-04       Impact factor: 11.105

5.  Volumetric quantification of lung nodules in CT with iterative reconstruction (ASiR and MBIR).

Authors:  Baiyu Chen; Huiman Barnhart; Samuel Richard; Marthony Robins; James Colsher; Ehsan Samei
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

6.  CT volumetry of artificial pulmonary nodules using an ex vivo lung phantom: influence of exposure parameters and iterative reconstruction on reproducibility.

Authors:  Mark O Wielpütz; Mathieu Lederlin; Jacek Wroblewski; Julien Dinkel; Monika Eichinger; Jürgen Biederer; Hans-Ulrich Kauczor; Michael Puderbach
Journal:  Eur J Radiol       Date:  2013-05-31       Impact factor: 3.528

7.  Variability of semiautomated lung nodule volumetry on ultralow-dose CT: comparison with nodule volumetry on standard-dose CT.

Authors:  Patrick A Hein; Valentina C Romano; Patrik Rogalla; Christian Klessen; Alexander Lembcke; Lars Bornemann; Volker Dicken; Bernd Hamm; Hans-Christian Bauknecht
Journal:  J Digit Imaging       Date:  2008-09-05       Impact factor: 4.056

8.  Evaluating variability in tumor measurements from same-day repeat CT scans of patients with non-small cell lung cancer.

Authors:  Binsheng Zhao; Leonard P James; Chaya S Moskowitz; Pingzhen Guo; Michelle S Ginsberg; Robert A Lefkowitz; Yilin Qin; Gregory J Riely; Mark G Kris; Lawrence H Schwartz
Journal:  Radiology       Date:  2009-07       Impact factor: 11.105

9.  Accuracy of lung nodule volumetry in low-dose CT with iterative reconstruction: an anthropomorphic thoracic phantom study.

Authors:  K W Doo; E-Y Kang; H S Yong; O H Woo; K Y Lee; Y-W Oh
Journal:  Br J Radiol       Date:  2014-07-16       Impact factor: 3.039

10.  An architecture for computer-aided detection and radiologic measurement of lung nodules in clinical trials.

Authors:  Matthew S Brown; Richard Pais; Peiyuan Qing; Sumit Shah; Michael F McNitt-Gray; Jonathan G Goldin; Iva Petkovska; Lien Tran; Denise R Aberle
Journal:  Cancer Inform       Date:  2007-05-12
View more
  16 in total

1.  Technical Note: FreeCT_ICD: An open-source implementation of a model-based iterative reconstruction method using coordinate descent optimization for CT imaging investigations.

Authors:  John M Hoffman; Frédéric Noo; Stefano Young; Scott S Hsieh; Michael McNitt-Gray
Journal:  Med Phys       Date:  2018-06-01       Impact factor: 4.071

2.  Technical Note: Design and implementation of a high-throughput pipeline for reconstruction and quantitative analysis of CT image data.

Authors:  John Hoffman; Nastaran Emaminejad; Muhammad Wahi-Anwar; Grace H Kim; Matthew Brown; Stefano Young; Michael McNitt-Gray
Journal:  Med Phys       Date:  2019-04-03       Impact factor: 4.071

3.  Technical Note: FreeCT_wFBP: A robust, efficient, open-source implementation of weighted filtered backprojection for helical, fan-beam CT.

Authors:  John Hoffman; Stefano Young; Frédéric Noo; Michael McNitt-Gray
Journal:  Med Phys       Date:  2016-03       Impact factor: 4.071

4.  Noninvasive radiomics signature based on quantitative analysis of computed tomography images as a surrogate for microvascular invasion in hepatocellular carcinoma: a pilot study.

Authors:  Shaimaa Bakr; Sebastian Echegaray; Rajesh Shah; Aya Kamaya; John Louie; Sandy Napel; Nishita Kothary; Olivier Gevaert
Journal:  J Med Imaging (Bellingham)       Date:  2017-08-21

5.  Interchangeability between real and three-dimensional simulated lung tumors in computed tomography: an interalgorithm volumetry study.

Authors:  Marthony Robins; Justin Solomon; Jocelyn Hoye; Taylor Smith; Yuese Zheng; Lukas Ebner; Kingshuk Roy Choudhury; Ehsan Samei
Journal:  J Med Imaging (Bellingham)       Date:  2018-09-24

6.  Techniques for virtual lung nodule insertion: volumetric and morphometric comparison of projection-based and image-based methods for quantitative CT.

Authors:  Marthony Robins; Justin Solomon; Pooyan Sahbaee; Martin Sedlmair; Kingshuk Roy Choudhury; Aria Pezeshk; Berkman Sahiner; Ehsan Samei
Journal:  Phys Med Biol       Date:  2017-08-22       Impact factor: 3.609

7.  The effect of radiation dose reduction on computer-aided detection (CAD) performance in a low-dose lung cancer screening population.

Authors:  Stefano Young; Pechin Lo; Grace Kim; Matthew Brown; John Hoffman; William Hsu; Wasil Wahi-Anwar; Carlos Flores; Grace Lee; Frederic Noo; Jonathan Goldin; Michael McNitt-Gray
Journal:  Med Phys       Date:  2017-03-14       Impact factor: 4.071

8.  Celebrating Contributions of Imaging Technology to Fight against Cancer at the 50th NCA Anniversary.

Authors:  Yantian Zhang; Robert Nordstrom
Journal:  Radiol Imaging Cancer       Date:  2022-08

9.  Quantitative assessment of nonsolid pulmonary nodule volume with computed tomography in a phantom study.

Authors:  Marios A Gavrielides; Benjamin P Berman; Mark Supanich; Kurt Schultz; Qin Li; Nicholas Petrick; Rongping Zeng; Jenifer Siegelman
Journal:  Quant Imaging Med Surg       Date:  2017-12

10.  Reproducibility of lung nodule radiomic features: Multivariable and univariable investigations that account for interactions between CT acquisition and reconstruction parameters.

Authors:  Nastaran Emaminejad; Muhammad Wasil Wahi-Anwar; Grace Hyun J Kim; William Hsu; Matthew Brown; Michael McNitt-Gray
Journal:  Med Phys       Date:  2021-04-13       Impact factor: 4.506

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.