The effect of some instruments for prothrombin time testing on the International Sensitivity Index (ISI) of two rabbit tissue thromboplastin reagents.

M Poggio1, A M van den Besselaar, E A van der Velde, R M Bertina.   

Abstract

Two commercial rabbit tissue thromboplastins were calibrated against the International Reference Preparation for rabbit thromboplastin (coded RBT/79) by one laboratory using the manual technique, a semi-automatic electro-mechanical coagulometer and three different automatic photo-optical instruments. The calibration of the two reagents was performed in three and two different exercises, respectively, and showed good reproducibility of the procedure. The purpose of calibration is providing a formula for the assessment of the International Normalized Ratio (INR) for patients receiving oral anticoagulants. The World Health Organization (WHO) model for thromboplastin calibration leads to the equation INR = RISI, in which R is the prothrombin time ratio and ISI the International Sensitivity Index of the calibrated thromboplastin/instrument system. This equation was adequate for one reagent, but not for the other when it was used in combination with the four instruments. At therapeutic intensities of anticoagulation, the deviation from the WHO-model observed for the second reagent was clinically insignificant. The WHO model was fully adequate when the second reagent was used with the manual technique. For both thromboplastins, there were statistically significant differences in ISI between the four instruments. The largest difference--amounting to approximately 10%--was observed between two photo-optical instruments. The order of instruments with increasing ISI was the same for the two reagents. It is recommended that thromboplastin manufacturers specify the instruments used for calibration of their reagent.

Entities:  

Mesh:

Substances:

Year:  1989        PMID: 2595661

Source DB:  PubMed          Journal:  Thromb Haemost        ISSN: 0340-6245            Impact factor:   5.249


  6 in total

Review 1.  Oral anticoagulant therapy: Antithrombotic Therapy and Prevention of Thrombosis, 9th ed: American College of Chest Physicians Evidence-Based Clinical Practice Guidelines.

Authors:  Walter Ageno; Alexander S Gallus; Ann Wittkowsky; Mark Crowther; Elaine M Hylek; Gualtiero Palareti
Journal:  Chest       Date:  2012-02       Impact factor: 9.410

2.  Assessment of value of calibrated lyophilised plasmas to determine International Sensitivity Index for coagulometers.

Authors:  K Clarke; D A Taberner; J M Thomson; J A Morris; L Poller
Journal:  J Clin Pathol       Date:  1992-01       Impact factor: 3.411

3.  Multicentre evaluation of the Thrombotest International Sensitivity Index used with a steel ball coagulometer.

Authors:  A M van den Besselaar; R H Peters
Journal:  J Clin Pathol       Date:  1996-05       Impact factor: 3.411

4.  Local INR correction: justification for a simplified approach.

Authors:  S Craig; K J Stevenson; J M Dufty; D A Taberner
Journal:  J Clin Pathol       Date:  1997-09       Impact factor: 3.411

5.  Calculation of System International Sensitivity Index: how many calibrant plasmas are required?

Authors:  J M Dufty; S Craig; K J Stevenson; D A Taberner
Journal:  J Clin Pathol       Date:  1997-01       Impact factor: 3.411

6.  Monitoring of anticoagulant therapy in heart disease: considerations for the current assays.

Authors:  Mohammadali Boroumand; Hamidreza Goodarzynejad
Journal:  J Tehran Heart Cent       Date:  2010-05-31
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.