| Literature DB >> 25931745 |
Abstract
[Purpose] The purpose of the present study was to compare the reliability of 2 hand-held dynamometers (HHD-1, 2) with different designs, by performing isometric knee muscle extension measurements two times each. [Subjects] The subjects were 40 young healthy adults. [Methods] The reliability of the measurements was examined using Bland-Altman analysis.Entities:
Keywords: Hand-held dynamometer; Muscle strength; Reliability
Year: 2015 PMID: 25931745 PMCID: PMC4395729 DOI: 10.1589/jpts.27.851
Source DB: PubMed Journal: J Phys Ther Sci ISSN: 0915-5287
Values of isometric knee extension muscle strength
| Value | Body weight ratio | ||||
|---|---|---|---|---|---|
| kgf | kgf/kg | ||||
| HHD-1 | 1st | 44.1 | (17.7) | 0.737 | (0.205) |
| 2nd | 46.1 | (17.3) | 0.774 | (0.195) | |
| large | 46.9 | (17.4) | 0.712 | (0.192) | |
| HHD-2 | 1st | 42.5 | (16.5) | 0.712 | (0.192) |
| 2nd | 44.4 | (17.7) | 0.744 | (0.207) | |
| large | 45.4 | (18.0) | 0.787 | (0.194) | |
Mean (SD), HHD-1: μTas F-1; HHD-2: Mobie; 1st: The 1st value; 2nd: The 2nd value; large: The value of the larger of two measurements
Isometric knee extension strength measurement reliabilities of two hand-held dynamometers
| Group | n | ICC (1,1) | Bland-Altman analysis | ||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Point estimation | (95% CI) | LOA | SEM | Fixed bias | Proportional bias | Random error | |||||
| 95% CI | bias* | slope** | bias* | MDC | |||||||
| HHD-1a) | 40 | 0.958 | 0.923 to 0.978 | −8.80–4.60 | 0.70 | −3.70 to −0.60 | exist | 0.016 | p=0.711 | n-exi | |
| 0.909 | 0.835 to 0.953 | −0.15–0.07 | 0.01 | −0.06 to −0.01 | exist | −0.076 | p=0.138 | n-exi | |||
| HHD-2a) | 40 | 0.953 | 0.913 to 0.975 | −9.40–7.00 | 0.90 | −3.00 to 0.70 | n−exi | −0.028 | p=0.630 | n-exi | 11.4 |
| 0.913 | 0.843 to 0.953 | −0.14–0.08 | 0.01 | −0.06 to −0.01 | exist | 0.026 | p=0.586 | n-exi | |||
| HHD1-2b) | 40 | 0.928 | 0.868 to 0.961 | −5.60–10.00 | 0.90 | 0.40 to 3.90 | exist | 0.013 | p=0.808 | n-exi | |
| 0.846 | 0.729 to 0.915 | −0.18–0.13 | 0.02 | −0.06 to 0.01 | n−exi | −0.095 | p=0.198 | n-exi | 0.22 | ||
Top: Measured value; Lower; Body weight ratio. HHD-1: μ tas; HHD-2: Mobie. a) Reliability of the 1st value and the 2nd value. b) Reliability of the value of the larger of two measurements by both devices. ICC: intraclass correlation coefficient; 95% CI: 95% confidence interval; LOA: limits of agreement; SEM: standard error of measurements; *: presence of bias, exist: present; n-exi: not-present; **: Slope of regression line; MDC: minimal detectable change