| Literature DB >> 25931075 |
Stefan Moeller1, Garth Brown1, Georgi Dakovski1, Bruce Hill1, Michael Holmes1, Jennifer Loos1, Ricardo Maida1, Ernesto Paiser1, William Schlotter1, Joshua J Turner1, Alex Wallace1, Ulf Jastrow2, Svea Kreis2, Andrey A Sorokin2, Kai Tiedtke2.
Abstract
A gas monitor detector was implemented and characterized at the Soft X-ray Research (SXR) instrument to measure the average, absolute and pulse-resolved photon flux of the LCLS beam in the energy range between 280 and 2000 eV. The detector is placed after the monochromator and addresses the need to provide reliable absolute pulse energy as well as pulse-resolved measurements for the various experiments at this instrument. This detector provides a reliable non-invasive measurement for determining flux levels on the samples in the downstream experimental chamber and for optimizing signal levels of secondary detectors and for the essential need of data normalization. The design, integration into the instrument and operation are described, and examples of its performance are given.Entities:
Keywords: FEL; diagnostics; pulse energy detector
Year: 2015 PMID: 25931075 PMCID: PMC4416676 DOI: 10.1107/S1600577515006098
Source DB: PubMed Journal: J Synchrotron Radiat ISSN: 0909-0495 Impact factor: 2.616
Figure 1Optical lay-out of the SXR instrument. The front-end enclosure (FEE) includes the two LCLS gas detectors with attenuator system and the offset and steering mirrors. The SXR monochromator is located in hutch 1 (AMO hutch) with its exit slit in hutch 2 (SXR hutch) which is followed by the GMD, the KB mirrors and experimental chamber.
Figure 2Schematic diagram of the gas monitor detection. The blue arrow represents the FEL beam. The red arrows represent the ion and the green arrows the electron trajectories. The applied high-voltage scheme and the signal readout channels are shown. Figure reprinted with permission from Tiedtke et al. (2014 ▶).
Figure 3Single-pulse ion time-of-flight spectrum (uncorrected) for krypton at 1300 eV.
Figure 4Schematic of the mechanical assembly. The detector is located in the center chamber with differential pumping stages on the upstream and downstream side. The upstream aperture assembly and downstream gate valve with stand and strut system are also shown.
Figure 5Measured transmission of the SXR instrument in zeroth order of the monochromator (transmission of the KB optics is not included). The error bars combine the variability of the limited measurement points and the estimated 5% uncertainty of the pulse energy measurement.
Measured transmission and average number of photons per pulse per mJ of the SXR instrument in zero-order mode of the monochromator (KB optics not included)
| Photon energy (eV) | Transmission (%) | Average number of photons per pulse per mJ |
|---|---|---|
| 495 | 23 3 | 2.8 0.4 1012 |
| 773 | 22 3 | 1.7 0.3 1012 |
| 1300 | 22 3 | 1.1 0.2 1012 |
| 1600 | 24 4 | 1 0.1 1012 |
| 1840 | 30 5 | 1 0.2 1012 |
| 2000 | 30 5 | 0.9 0.1 1012 |
Figure 6Ion pulse signal versus FEE gas detector correlation for varying attenuation levels. Target gas argon, photon energy 1300 eV, zero-order monochromator mode. A linear fit (black straight line) shows the deviation at higher pulse energies.