Literature DB >> 25921329

National dosimetric audit network finds discrepancies in AAA lung inhomogeneity corrections.

Leon Dunn1, Joerg Lehmann2, Jessica Lye3, John Kenny4, Tomas Kron5, Andrew Alves3, Andrew Cole6, Jackson Zifodya7, Ivan Williams8.   

Abstract

This work presents the Australian Clinical Dosimetry Service's (ACDS) findings of an investigation of systematic discrepancies between treatment planning system (TPS) calculated and measured audit doses. Specifically, a comparison between the Anisotropic Analytic Algorithm (AAA) and other common dose-calculation algorithms in regions downstream (≥2cm) from low-density material in anthropomorphic and slab phantom geometries is presented. Two measurement setups involving rectilinear slab-phantoms (ACDS Level II audit) and anthropomorphic geometries (ACDS Level III audit) were used in conjunction with ion chamber (planar 2D array and Farmer-type) measurements. Measured doses were compared to calculated doses for a variety of cases, with and without the presence of inhomogeneities and beam-modifiers in 71 audits. Results demonstrate a systematic AAA underdose with an average discrepancy of 2.9 ± 1.2% when the AAA algorithm is implemented in regions distal from lung-tissue interfaces, when lateral beams are used with anthropomorphic phantoms. This systemic discrepancy was found for all Level III audits of facilities using the AAA algorithm. This discrepancy is not seen when identical measurements are compared for other common dose-calculation algorithms (average discrepancy -0.4 ± 1.7%), including the Acuros XB algorithm also available with the Eclipse TPS. For slab phantom geometries (Level II audits), with similar measurement points downstream from inhomogeneities this discrepancy is also not seen. Crown
Copyright © 2015. Published by Elsevier Ltd. All rights reserved.

Keywords:  Algorithm; Audit; Dose; Dosimetry; Radiotherapy

Mesh:

Year:  2015        PMID: 25921329     DOI: 10.1016/j.ejmp.2015.04.002

Source DB:  PubMed          Journal:  Phys Med        ISSN: 1120-1797            Impact factor:   2.685


  13 in total

1.  Credentialing of radiotherapy centres in Australasia for TROG 09.02 (Chisel), a Phase III clinical trial on stereotactic ablative body radiotherapy of early stage lung cancer.

Authors:  Tomas Kron; Brent Chesson; Nicholas Hardcastle; Melissa Crain; Natalie Clements; Mark Burns; David Ball
Journal:  Br J Radiol       Date:  2018-03-06       Impact factor: 3.039

2.  Virtual bronchoscopy-guided lung SAbR: dosimetric implications of using AAA versus Acuros XB to calculate dose in airways.

Authors:  P Kinkopf; A Modiri; Kun-Chang Yu; Y Yan; P Mohindra; R Timmerman; A Sawant; E Vicente
Journal:  Biomed Phys Eng Express       Date:  2021-09-15

3.  Radiotherapy of lung cancers: FFF beams improve dose coverage at tumor periphery compromised by electronic disequilibrium.

Authors:  Oleg N Vassiliev; Stephen F Kry; He C Wang; Christine B Peterson; Joe Y Chang; Radhe Mohan
Journal:  Phys Med Biol       Date:  2018-09-28       Impact factor: 3.609

4.  On the use of AAA and AcurosXB algorithms for three different stereotactic ablative body radiotherapy (SABR) techniques: Volumetric modulated arc therapy (VMAT), intensity modulated radiation therapy (IMRT) and 3D conformal radiotherapy (3D-CRT).

Authors:  Abdulrahman Tajaldeen; Prabhakar Ramachandran; Salem Alghamdi; Moshi Geso
Journal:  Rep Pract Oncol Radiother       Date:  2019-03-23

5.  A depth dose study between AAA and AXB algorithm against Monte Carlo simulation using AIP CT of a 4D dataset from a moving phantom.

Authors:  Roger Cai Xiang Soh; Guan Heng Tay; Wen Siang Lew; James Cheow Lei Lee
Journal:  Rep Pract Oncol Radiother       Date:  2018-09-03

6.  A clinical database to assess action levels and tolerances for the ongoing use of Mobius3D.

Authors:  David Jolly; Leon Dunn; John Kenny
Journal:  J Appl Clin Med Phys       Date:  2016-11-30       Impact factor: 2.102

7.  Automated data mining of a plan-check database and example application.

Authors:  Leon Dunn; David Jolly
Journal:  J Appl Clin Med Phys       Date:  2018-06-29       Impact factor: 2.102

8.  Dose deviations induced by respiratory motion for radiotherapy of lung tumors: Impact of CT reconstruction, plan complexity, and fraction size.

Authors:  Erlend P S Sande; Ana M Acosta Roa; Taran P Hellebust
Journal:  J Appl Clin Med Phys       Date:  2020-03-12       Impact factor: 2.102

9.  Clinical Implications of TiGRT Algorithm for External Audit in Radiation Oncology.

Authors:  Daryoush Shahbazi-Gahrouei; Mohsen Saeb; Shahram Monadi; Iraj Jabbari
Journal:  Adv Biomed Res       Date:  2017-09-21

10.  A Homogeneous Water-Equivalent Anthropomorphic Phantom for Dosimetric Verification of Radiotherapy Plans.

Authors:  Manikandan Arjunan; Sureka Chandra Sekaran; Biplab Sarkar; Sujatha Manikandan
Journal:  J Med Phys       Date:  2018 Apr-Jun
View more