Addition of magnetic resonance imaging to computed tomography-based three-dimensional conformal radiotherapy planning for postoperative treatment of astrocytomas: Changes in tumor volume and isocenter shift.

Puneet Kumar Bagri1, Akhil Kapoor1, Daleep Singh1, Mukesh Kumar Singhal1, Satya Narayan1, Harvindra Singh Kumar1.   

Abstract

INTRODUCTION: Postoperative radiotherapy is the current gold standard treatment in astrocytomas. Computed tomography (CT)-based radiotherapy planning leads to either missing of the tumor volume or underdosing. The aim of this prospective study was to study the changes in tumor volume on addition of magnetic resonance imaging (MRI) to CT-based three-dimensional radiotherapy treatment planning of astrocytomas.
MATERIALS AND METHODS: Twenty-five consecutive patients of astrocytoma (WHO grades I-IV) for postoperative three-dimensional conformal radiotherapy were included in this prospective study. Postoperative tumor volumes were contoured on CT-based images and recontoured on CT-MRI images after automated MRI co-registration on treatment planning system Eclipse 8.9.15 as per ICRU-50 report. Tumor volumes were compared with each other. RESULT: The MRI-based mean and median tumor volume was 24.24 cc ± 13.489 and 18.72 cc (range 5.6-46.48 cc), respectively, while for CT it was 19.4 cc ± 11.218 and 16.24 cc (range: 5.1-38.72 cc), respectively. The mean and median isocenter shift between CT and MRI was 4.05 mm and 4.39 mm (range 0.92-6.32 mm), respectively. There is a linear relationship between MRI and CT volume with a good correlation coefficient of R (2) = 0.989, and MRI-based tumor volume was 1.208 times as compared to CT volume. Statistical analysis using paired sample t-test for the difference in CT and MRI tumor volume was highly significant (P < 0.001).
CONCLUSION: Addition of MRI to the CT-based three-dimensional radiation treatment planning reduces the chances of geographical miss or tumor under dosing. Thus, MRI should be an integral part of three-dimensional planning of astrocytomas.

Entities:  

Keywords:  Astrocytoma; computed tomography-magnetic resonance imaging fusion; isocenter shift; postoperative radiotherapy; tumor volume

Year:  2015        PMID: 25839014      PMCID: PMC4382776          DOI: 10.4103/2278-330X.149939

Source DB:  PubMed          Journal:  South Asian J Cancer        ISSN: 2278-330X


Introduction

Modern anatomic imaging technologies, such as computed tomography (CT) and magnetic resonance imaging (MRI), provide a fully three-dimensional model of the cancer patient's anatomy. Such advanced imaging allows the radiation oncologists to accurately delineate tumor volumes and their relationship with other critical normal organs. CT provides geometrically precise scans, but it gives less information about the soft tissues as compared to MRI. The resolution of the brain parenchyma is far superior in an MRI as it provides greater detail and better topographic distribution of the abnormal lesions.[1234] The possibility of obtaining high-quality images without the bone artifact makes MRI particularly attractive for evaluating lesions in the posterior fossa, at the vertex, and the base of the skull. Although MRI is adequately sensitive for the detection of intracranial lesions, it lacks specificity in pronouncing tumor cell type or grade. Further, MRI abnormalities seen in astrocytomas overlap with those seen in stroke and demyelinating disease. CT-MRI fusion is the standard imaging technique for three-dimensional radiotherapy treatment planning for brain tumors. The aim of this prospective study was to study the advantages of addition of MRI- to CT-based three-dimensional radiotherapy treatment planning of primary brain tumors.

Materials and Methods

Twenty-five consecutive patients with histopathologically proven astrocytoma (WHO grades I-IV) referred to our regional cancer institute for postoperative three-dimensional conformal radiotherapy were included in this prospective study. Immobilization was done using ORFIT thermoplastic cast (ORFIT Industries, Belgium), and contrast-enhanced CT and MRI was done with the cast in situ. Computed tomography simulation was done with the patients immobilized on the flat couch of the diagnostic CT unit (CT Hispeed Nxi, Wipro GE, Bangalore, India). Following early intravenous contrast, axial scans of 3 mm slice thickness were taken from the vertex to the base of the skull. These images were transferred to the radiation therapy treatment planning system (TPS) Eclipse 8.9.15 (Varian, Palo Alto, USA). Magnetic resonance imaging scans were taken with a standard head coil (1.5 T, Philips, USA). A custom-made Perspex base plate was used for fixing the cast during MRI, which was similar to the one used during CT acquisition. Spin-echo (SE) sequence, having the least image distortion, was used to obtain images from the vertex to the base of the skull.[5] The conventional SE T2-weighted plain and postgadolinium contrast T1-weighted axial MRI images of 3 mm slice thickness were obtained. The MRI images were transferred to the radiation therapy TPS Eclipse 8.9.15. Postoperative tumor volumes were contoured on CT-based images as per the recommendations of ICRU-50 report.[6] Tumor volumes were recontoured on composite CT-MRI images after automated MRI co-registration. All the contourings were done by same person to eliminate interobserver variability, and the contouring process was double-blinded. Tumor volumes were measured and compared with each other. Geometrical center of the target volume was taken as isocenter of tumor volume contoured on CT and MRI. Isocenter shift of tumor was determined with the distance formula. Total tumor dose delivered was 50.4–64.8 Gy depending on astrocytoma grade. Statistical analysis was done using SPSS statistical software for Windows, version 20.0 (IBM Corp., Armonk, NY, USA).

Results

For postoperative three-dimensional conformal radiotherapy planning, CT simulation was done, and MRI was carried out within 3 days of CT simulation. The mean and median interval between surgery and postoperative CT was 23.32 and 24 days. The mean and median time interval between CT and MRI was 1.48 days and 1 day, respectively. High-grade astrocytoma patients were more than with low grade (17:8). The main operative procedures were either subtotal resection or tumor decompression. Total resection was done in only two patients. Table 1 shows the baseline characteristics of the patients.
Table 1

Patient characteristics (n=25)

Patient characteristics (n=25) The MRI-based mean and median tumor volume was 24.24 cc ± 13.489 and 18.72 cc (range 5.6–46.48 cc), respectively, while for CT it was 19.4 cc ± 11.218 and 16.24 cc (range: 5.1–38.72 cc) respectively. The mean and median isocenter shift between CT and MRI was 4.05 mm and 4.39 mm (range 0.92–6.32 mm), respectively. There is a linear relationship between MRI and CT volume with a good correlation coefficient of R2 = 0.989 and MRI-based tumor volume was 1.208 times as compared to CT volume [Figure 1a]. Bland-Altman test shows that MRI volume is larger than CT volume [Figure 1b]. Statistical analysis using paired sample t-test for the difference in CT and MRI tumor volume was highly significant (P < 0.001). Compared to MRI, CT significantly underestimated the volumes, resulting in a mean difference between the imaging modalities (CT volume − MRI volume) of − 4.684 ± 2.754 cc (P < 0.001) [Table 2]. The percentage difference calculated taking the CT volume as baseline [(CT-based volume − MRI-based volume)/CT-based volume × 100] shows that for tumor volume, the mean percentage difference was − 25.9% (SD: ±11.6). The isocenter shift decreases with an increase in tumor volume [Figure 1c].
Figure 1

(a) Scatter diagram for computed tomography (CT) and magnetic resonance imaging (MRI) based gross tumor volume. The curve represents linear relationship (R2 = 0.989). MRI shows 1.208 times more volume as compared to CT volume. (b) Bland-Altman test for computed tomography and MRI-based tumor volume. (c) Isocenter shift decreases with an increase in tumor volume. (d) Histogram with normal distribution curve depicting percentage difference in tumor volume for low grade versus high-grade astrocytoma. (e) Histogram with normal distribution curve for percentage difference in tumor volume for overall patients included in the study

Table 2

Mean differences in tumor volumes and isocenter shift as evident on CT and MRI

(a) Scatter diagram for computed tomography (CT) and magnetic resonance imaging (MRI) based gross tumor volume. The curve represents linear relationship (R2 = 0.989). MRI shows 1.208 times more volume as compared to CT volume. (b) Bland-Altman test for computed tomography and MRI-based tumor volume. (c) Isocenter shift decreases with an increase in tumor volume. (d) Histogram with normal distribution curve depicting percentage difference in tumor volume for low grade versus high-grade astrocytoma. (e) Histogram with normal distribution curve for percentage difference in tumor volume for overall patients included in the study Mean differences in tumor volumes and isocenter shift as evident on CT and MRI The comparison of CT and MRI-based tumor volumes was also done separately for low grade (grades I and II) and high grade (grades III and IV) astrocytoma [Table 2]. The MRI-based volumes were significantly higher (P < 0.001) in the low grade as well as high-grade astrocytoma. Figure 1d shows the histogram with normal distribution curve depicting percentage difference in tumor volume for low grade versus high-grade astrocytoma. Figure 1e shows the histogram for percentage difference in tumor volume for overall patients included in the study.

Discussion

Whole brain radiotherapy was the standard of care in the era of nonavailability of sophisticated imaging modalities. However, with poor outcomes after the treatment of malignant astrocytoma, partial brain irradiation after surgery is the current gold standard. It has been realized that the region of recurrence in malignant astrocytoma is usually the tumor bed itself, stressing on the fact that accurate target volume delineation is mandatory especially while giving the boost dose to the surgical bed. Delineation of the target volume is highly dependent on the observer (both inter and intra observer variations).[78] With the help of CT-MRI fusion based planning, the chances of geographical miss are minimal. Treatment planning based on CT only will lead to either miss the tumor volume or under dosing of the tumor. In this study, use of MRI caused 25.9% increase in gross tumor volume. This showed the importance of MRI in tumor delineation for tumors in the brain. Shift of isocenter also showed the possibilities of under dosing of the tumor. The fact that MRI shows larger tumor volume compared to CT is well in accord with the published results. As the tumor volume increases, the isocenter shift decreases. This clearly indicates that the chances of tumor miss are more when tumor is small and planning is done based only on CT scan. CT-MRI fusion is necessary for planning of brain tumors to rule out geographical misses. Initial study for radiation treatment planning using MRI demonstrates a 30% reduction of treatment volume.[9] But the other studies show that the use of MRI in tumor volume delineation increases the tumor volume.[10111213] Hence, MRI is superior to provide the detail of tumor volume and volume of organs at risk.[14]

Conclusion

We conclude that addition of MRI- to the CT-based three-dimensional radiation treatment planning reduces the chances of geographical miss or tumor under dosing. Thus, MRI should be an integral part of three-dimensional planning of astrocytomas.
  12 in total

1.  Reproducibility of geometric distortion in magnetic resonance imaging based on phantom studies.

Authors:  T Mizowaki; Y Nagata; K Okajima; M Kokubo; Y Negoro; N Araki; M Hiraoka
Journal:  Radiother Oncol       Date:  2000-11       Impact factor: 6.280

Review 2.  The impact of gross tumor volume (GTV) and clinical target volume (CTV) definition on the total accuracy in radiotherapy theoretical aspects and practical experiences.

Authors:  Elisabeth Weiss; Clemens F Hess
Journal:  Strahlenther Onkol       Date:  2003-01       Impact factor: 3.621

3.  Interobserver variations in gross tumor volume delineation of brain tumors on computed tomography and impact of magnetic resonance imaging.

Authors:  C Weltens; J Menten; M Feron; E Bellon; P Demaerel; F Maes; W Van den Bogaert; E van der Schueren
Journal:  Radiother Oncol       Date:  2001-07       Impact factor: 6.280

Review 4.  Imaging of gliomas.

Authors:  T N Byrne
Journal:  Semin Oncol       Date:  1994-04       Impact factor: 4.929

Review 5.  Target volumes in radiotherapy for high-grade malignant glioma of the brain.

Authors:  E P Jansen; L G Dewit; M van Herk; H Bartelink
Journal:  Radiother Oncol       Date:  2000-08       Impact factor: 6.280

6.  Comparison of computed tomography and magnetic resonance based target volume in brain tumors.

Authors:  R Prabhakar; K P Haresh; T Ganesh; R C Joshi; P K Julka; G K Rath
Journal:  J Cancer Res Ther       Date:  2007 Apr-Jun       Impact factor: 1.805

Review 7.  A systematic overview of radiation therapy effects in brain tumours.

Authors:  Gertrud Berg; Erik Blomquist; Eva Cavallin-Ståhl
Journal:  Acta Oncol       Date:  2003       Impact factor: 4.089

8.  The clinical utility of magnetic resonance imaging in 3-dimensional treatment planning of brain neoplasms.

Authors:  A F Thornton; H M Sandler; R K Ten Haken; D L McShan; B A Fraass; M L La Vigne; B R Yanke
Journal:  Int J Radiat Oncol Biol Phys       Date:  1992       Impact factor: 7.038

9.  Implications of contrast-enhanced CT-based and MRI-based target volume delineations in radiotherapy treatment planning for brain tumors.

Authors:  Niloy R Datta; Rajasekar David; Rakesh K Gupta; Punita Lal
Journal:  J Cancer Res Ther       Date:  2008 Jan-Mar       Impact factor: 1.805

10.  Radiologic diagnosis, staging, and follow-up of adult central nervous system primary malignant glioma.

Authors:  M T Madison; W A Hall; R E Latchaw; D J Loes
Journal:  Radiol Clin North Am       Date:  1994-01       Impact factor: 2.303

View more
  1 in total

1.  Impact on Radiation Dose and Volume V57 Gy of the Brain on Recurrence and Survival of Patients with Glioblastoma Multiformae.

Authors:  Igor Stojkovski; Valentina Krstevska; Snezhana Smichkoska
Journal:  Radiol Oncol       Date:  2017-11-01       Impact factor: 2.991

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.