OBJECTIVE: Accuracy of in vivo activity quantification improves after the correction of penetrated and scattered photons. However, accurate assessment is not possible with physical experiment. We have used Monte Carlo Simulation to accurately assess the contribution of penetrated and scattered photons in the photopeak window. MATERIALS AND METHODS: Simulations were performed with Simulation of Imaging Nuclear Detectors Monte Carlo Code. The simulations were set up in such a way that it provides geometric, penetration, and scatter components after each simulation and writes binary images to a data file. These components were analyzed graphically using Microsoft Excel (Microsoft Corporation, USA). Each binary image was imported in software (ImageJ) and logarithmic transformation was applied for visual assessment of image quality, plotting profile across the center of the images and calculating full width at half maximum (FWHM) in horizontal and vertical directions. RESULTS: The geometric, penetration, and scatter at 140 keV for low-energy general-purpose were 93.20%, 4.13%, 2.67% respectively. Similarly, geometric, penetration, and scatter at 140 keV for low-energy high-resolution (LEHR), medium-energy general-purpose (MEGP), and high-energy general-purpose (HEGP) collimator were (94.06%, 3.39%, 2.55%), (96.42%, 1.52%, 2.06%), and (96.70%, 1.45%, 1.85%), respectively. For MEGP collimator at 245 keV photon and for HEGP collimator at 364 keV were 89.10%, 7.08%, 3.82% and 67.78%, 18.63%, 13.59%, respectively. CONCLUSION: Low-energy general-purpose and LEHR collimator is best to image 140 keV photon. HEGP can be used for 245 keV and 364 keV; however, correction for penetration and scatter must be applied if one is interested to quantify the in vivo activity of energy 364 keV. Due to heavy penetration and scattering, 511 keV photons should not be imaged with HEGP collimator.
OBJECTIVE: Accuracy of in vivo activity quantification improves after the correction of penetrated and scattered photons. However, accurate assessment is not possible with physical experiment. We have used Monte Carlo Simulation to accurately assess the contribution of penetrated and scattered photons in the photopeak window. MATERIALS AND METHODS: Simulations were performed with Simulation of Imaging Nuclear Detectors Monte Carlo Code. The simulations were set up in such a way that it provides geometric, penetration, and scatter components after each simulation and writes binary images to a data file. These components were analyzed graphically using Microsoft Excel (Microsoft Corporation, USA). Each binary image was imported in software (ImageJ) and logarithmic transformation was applied for visual assessment of image quality, plotting profile across the center of the images and calculating full width at half maximum (FWHM) in horizontal and vertical directions. RESULTS: The geometric, penetration, and scatter at 140 keV for low-energy general-purpose were 93.20%, 4.13%, 2.67% respectively. Similarly, geometric, penetration, and scatter at 140 keV for low-energy high-resolution (LEHR), medium-energy general-purpose (MEGP), and high-energy general-purpose (HEGP) collimator were (94.06%, 3.39%, 2.55%), (96.42%, 1.52%, 2.06%), and (96.70%, 1.45%, 1.85%), respectively. For MEGP collimator at 245 keV photon and for HEGP collimator at 364 keV were 89.10%, 7.08%, 3.82% and 67.78%, 18.63%, 13.59%, respectively. CONCLUSION: Low-energy general-purpose and LEHR collimator is best to image 140 keV photon. HEGP can be used for 245 keV and 364 keV; however, correction for penetration and scatter must be applied if one is interested to quantify the in vivo activity of energy 364 keV. Due to heavy penetration and scattering, 511 keV photons should not be imaged with HEGP collimator.
Entities:
Keywords:
Collimator; Monte Carlo Simulation; parallel-hole
In order to improve photon statistics in the image, photopeak window is widened, thus the contribution of penetrated and scattered photons are included in the image. The geometric components of the photons are the photons which were detected without interaction inside the collimator. Gamma camera cannot classify the image forming photons into geometric, penetrated, or scattered photons. Therefore, with measuring the radio-activity, we cannot accurately characterize the collimator which includes the accurate assessment of penetrated and scattered photons. Single-photon emission computed tomography images are reconstructed on the assumption that detected photons are those photons which are emitted from a source located in a pixel area (rectangular area of constant width) and entered the collimator hole parallel to the hole axis. These criteria are only achieved if the collimator hole have infinite length and infinitely small hole size. Parallel-hole collimator has finite length and finite hole size and, therefore, the detected photons are from the area enclosed by a cone whose vertex is at the point of interaction site of the photon on the crystal. In this case, the number of detected photons increases with the distance of the source from the collimator encompassing more area and therefore spatial resolution decreases. Moreover, photons entering the collimator hole at small angle can be detected after penetrating the septa and/or scattering inside the collimator and introduces error in the quantification of in vivo activity and results in reduced image contrast.It is possible to track and record the life history of the individual photon originating from the source that ultimately deposits its complete energy inside the crystal using Monte Carlo Simulation. Therefore, with the help of Monte Carlo Simulation technique, accurate assessment of the geometric, penetration and scatter contribution inside the photopeak window can be made. The aim of the study was to estimate the geometric, penetration, and scatter components for parallel-hole collimators using Simulation of Imaging Nuclear Detectors (SIMIND) Monte Carlo Simulation code.
MATERIALS AND METHODS
In order to characterize the parallel-hole collimators we simulated the Millennium VG high-performance dual head gamma camera equipped with advanced XP digital detector GE Medical Systems, Milwaukee, WI, USA. having 1 inch NaI (Tl) crystal thickness, intrinsic spatial resolution of 0.450 cm and energy resolution of 8.80% at 140 keV. The dimension of the crystal was 500 × 400 mm. The parallel-hole collimators used was VG low-energy general-purpose (LEGP), low-energy high-resolution (LEHR), medium-energy general-purpose (MEGP) and high-energy general-purpose (HEGP) collimators. The collimator data used during the simulation are given in Table 1.
(a) Image of the point source with square ROI, (b) horizontal profile, (c) vertical profile and (d) result
(a) Image of the point source with square ROI, (b) horizontal profile, (c) vertical profile and (d) result
Algorithm to assess geometric, penetrated and scattered photons
The authors of the SIMIND Monte Carlo code have used the delta scattering methods to sample the photons interaction through the collimators[4]. The path length of the photons has been sampled according to the equation 1.Where ln is the natural logarithm, R is the sampled uniform random number, and μmax is the largest attenuation coefficient within the heterogeneous media.The collimator can be considered as heterogeneous medium of air (0 g/cm3) and lead. Therefore, the ratio of μ/μmax for the collimator will be either 0 or 1. In this case, it is only necessary to test whether the end point is within the collimator hole or not. It is not necessary to find the geometric coordinates of end point of the photon.The algorithm they have used to assess geometric, penetrated and scattered photons is briefly summarized as follows:Step 1: Record the entrance point of the gamma photon on the collimator face.Step 2: Determine whether the entrance point is inside the collimator hole or in the lead septa based on a mathematical formula mentioned in the reference[4] and keep the recordStep 3: Sample the path length of the photon using equation (1) and determine what will happen to the photon at the end of the sampled path. If R < μ/μmax, then location is the final end point of the photon following the real interaction. Otherwise, continue tracking of photon by a new distance sampled according to equation (1). Record the number of interactions between the entrance point and final end point of the photon. The history of the gamma photon is terminated when all of their energy was absorbed within the collimator material or escaped from collimator.Step 4: Record the exit point on the collimator face towards NaI (Tl) crystal.Step 5: Compare the entrance and the exit points. If the entrance and exit point is inside the hole of the collimator, then flag the photon as the geometric photon. If some (one or more) interaction has occurred inside the lead material then flag the photon as scattered photons, otherwise flag the photon as the penetrated photon.Steps 6: Calculate the percentage of geometric, scattered and penetrated photons by dividing these numbers with the total number of photons that makes hits with the detector.Step 7: Geometric, scattered and penetrated photon when makes hit with detector, track them individually and sampled for their contribution in the photopeak window. Calculate the percentage of geometric, scattered, penetrated in the photopeak window by dividing with the total number of photopeak counts.
RESULTS
We evaluated the geometric, septal penetration, and scattering component in parallel-hole collimators (LEGP, LEHR, MEGP, and HEGP) using radio-active point source having energy ranging from 75 keV to 511 keV using Monte Carlo Simulation. The result of the simulation is given in Table 2. Figure 2 shows the variation of geometric, penetration, and scatter component with energy in LEGP, LEHR, MEGP, and HEGP collimators, respectively. Figure 3 shows images of point source obtained as a result of the simulation. It is clear that the geometric component has decreased with increase in energy, very sharp transition in case of LEGP and LEHR while comparatively smooth transition in MEGP and HEGP collimators [Figure 2]. Penetration and scatter component has increased with increase in energy showing sharp increase in case of LEGP and LEHR, with a smooth increase in MEGP and HEGP collimator.
Table 2
The result of the simulation
Figure 2
The variation of geometric, penetration, and scatter component with energy in low-energy general-purpose (a), low-energy high-resolution (b), medium-energy general-purpose (c) and high-energy general-purpose (d) collimators, respectively
Figure 3
Image of the point source created at the end of each twenty simulation. Low-energy general-purpose (a-e), low-energy high-resolution (f-j), medium-energy general-purpose (k-o), high-energy general-purpose (p-t) images were simulated at 75 keV, 140 keV, 245 keV, 364 keV, and 511 keV, respectively
The result of the simulationThe variation of geometric, penetration, and scatter component with energy in low-energy general-purpose (a), low-energy high-resolution (b), medium-energy general-purpose (c) and high-energy general-purpose (d) collimators, respectivelyImage of the point source created at the end of each twenty simulation. Low-energy general-purpose (a-e), low-energy high-resolution (f-j), medium-energy general-purpose (k-o), high-energy general-purpose (p-t) images were simulated at 75 keV, 140 keV, 245 keV, 364 keV, and 511 keV, respectively
Visual and quantitative assessment of the point source images
Image of the point source created at the end of each twenty simulation is shown in Figure 3. As shown in Figure 3, we have found the point source image superimposed on intense foggy background, and star artefacts (the camera-wide tails showing six-fold symmetry) resulting in the loss of image contrast. For a particular selection of collimator, the fogginess in the images has increased with increase in energy. This may be because of the selected higher energy photons to be imaged with respective collimators; at these energies respective collimators are becoming virtually transparent. This is evident from the calculated value of high septa penetration and scattering obtained as a result of the simulation in these selected combination of collimator and energy of the gamma photon shown in Table 2. It is important to note that Figure 3e, the foggiest image has the highest value of penetration. The 6-fold symmetry of tails is associated with the hexagonal-hole shape of the collimator used in the simulation. The calculated vertical and horizontal FWHM on the images shown in Figure 3 is given in Table 3.
Table 3
FWHM of the point source images of Figure 3
FWHM of the point source images of Figure 3For a particular value of collimator, the FWHM of the point source (point spread function) have found to increase with increase in the energy, that is, with increase in energy, the system shows poorer spatial resolution. This may be due to increase in penetration and scatter inside the collimator with increase in photon energy. The spatial resolution (V = 9.962 mm and H = 9.52 mm at 140 keV) observed with LEHR collimator is best spatial resolution observed in this study. In this study we had selected two low-energy collimators (LEGP and LEHR), the better resolution of LEHR than that of LEGP may be due to smaller diameter of the holes (diameter = 0.150 cm, LEHR, and diameter = 0.190 cm, LEGP) [Table 2]. The spatial resolution observed at 140 keV for MEGP and HEGP in comparison to LEGP collimator may be due to combined effect of larger diameter of the holes (diameter = 0.300 cm for MEGP and diameter = 0.400 cm for HEGP, increases sensitivity) and comparatively low percentage of scattered and penetrated components in case of MEGP and HEGP collimators (penetration and scatter = 6.80% for LEGP, 3.58% for MEGP, and 3.30% for HEGP) [Table 2].
DISCUSSION
The effect of penetration and scatter have been studied both qualitatively and quantitatively in clinical and phantom study and found that the accuracy of quantification improves by incorporating the compensation of collimator detecting response function.[56789101112131415] The collimator detector response function has four component namely intrinsic response, geometric, penetration and scatter component of the collimator parameters. Accurate quantification of in vivo activity requires the assessment of the contribution of penetrated and scattered photons so that the correction can be made. We assessed these contributions using Monte Carlo Simulation.Although the geometric component does not depend on the energy, for all the four investigated parallel-hole collimator we found the value of geometric component at 75 keV was less than that at 140 keV. This may be because the one inch crystal thickness (used in the simulation) is optimized for imaging radioisotope having energy more than 100–511 keV and gamma camera is optimized for 140 keV most widely used radioisotope in nuclear medicine. 75 keV photon will penetrate approximately half the distance than that of 140 keV and light will travel from that point to the end of the crystal to PMT. The amount of light reaching the PMT will be less due to absorption and scattering of light inside the NaI (Tl) crystal. Therefore, the photopeak efficiency will be less in case of 75 keV photons. This may be the reason for less geometric component for 75 keV in comparison to 140 keV. Low-energy photon (75 keV) can also undergo multiple Compton scattering and final photopeak absorption, and this will increase the photopeak counts. Both the phenomena, that is, multiple scattering leading to final photopeak absorption and diffusion of light over a longer path before reaching PMTs, also degrade the spatial resolution.[16] For all the four collimators, we have found more FWHM (less spatial resolution) in comparison to the 140 keV photon [Table 3].The value of geometric component has decreased with increase in energy of the gamma photons with all the collimator. One inch crystal is sufficient to absorb all 140 keV gamma photon completely; however, it is not sufficient to stop all high energy photon above 245 keV. And, therefore, efficiency of one inch NaI (Tl) crystal is low. The low geometric component may be due to low efficiency of one inch crystal thickness used in the simulation. This result is in good agreement with the previous publications by Shafaei et al. they have estimated geometric, penetration, and scatter component for the energy range of 250–450 keV for DST-XLi dual-head gamma camera with HEGP collimator[17]. The poor spatial resolution was observed in the case of MEGP in comparison to LEHR and LEGP for 140 keV photon. It may be attributed to the larger hole diameter and increased septa thickness.The septa thickness is designed to control the number of photons penetrating the septa and generally it is such that less than 5% penetrated photons are allowed. We have found less than 5% septa penetration in both LEHR and LEGP collimators which is in agreement with the general principle of designing collimators septa thickness for imaging low-energy isotope up to 140 keV energy. We have found very high value of septa penetration and scatter, at energy higher than 140 keV for LEGP and LEHR collimator so that it is advised not to use these collimators for imaging high energy radioisotopes. Septa penetration was found to be more than 5% for both the collimators designed to image medium-energy or high energy radioisotopes. For MEGP collimator designed to image 245 keV, the septa penetration was 5.43% and HEGP designed to image 364 keV photons, we have found the value of septa penetration as 18.63% which is quite high, the contribution due to penetration and scattering is equal to 32.22%. Therefore, correction for penetration and scattering becomes very important in case of I-131 imaging when we are interested in the quantification of in vivo I-131 activity.The length of the collimator holes affects resolution and sensitivity. As holes are made longer, gamma rays that are not quite perpendicular to the collimator are more likely to be absorbed in the septa before they reach the scintillation crystal. For MEGP and HEGP collimator, the hole length is 5.800 cm and 6.000 cm respectively. Hole length are increasing to maintain the resolution and hole size is increasing to maintain the sensitivity. Hole size for MEGP and HEGP collimator are 0.300 cm and 0.400 cm, respectively. Septa thickness is also increased to reduce the septal penetration in these collimators at high energy.The inaccuracy in the calculation of FWHM [S. No. 3, 4, 5, 8, 9, 10, 14, 15, 19, 20 in Table 3] was observed in case of all the combination of collimator and energy of photons listed in Table 3. The calculated FWHM is nearly same for MEGP collimator (approximately 11.55 mm, S. No. 13, 14, and 15). However, by looking at the Figure 2m-o, it is obvious that the calculated FWHM cannot be accepted. The second example is in case of HEGP collimator, the calculate FWHM value shows that HEGP collimator has better resolution for 511 keV in comparison to the 364 keV photons, that is not acceptable which is clear from the Figure 2s and t. This inaccuracy in calculation of FWHM may be attributed to heavy septa penetration and scattering inside the collimator that has introduced star artefacts. The plugin used to calculate FWHM fits a Gaussian function to a horizontal and vertical cut that is centered on the brightest point in the image and returns the FWHM of the function both in numerical and graphical formats.There is an obvious star artefact in Figure 3n is due to septa penetration, it is clear from the figure that a bright line is passing vertically through the brightest point source, therefore, vertical profile have high value of FWHM, while no such line exists in the horizontal direction through the brightest points, therefore, horizontal FWHM will be much less in comparison to vertical FWHM.When the contribution of penetration and scattering is very high, the image of a point source becomes very blurred and becomes very difficult to put ROI in order to quantify the activity. The gamma camera modeled in our study is not the same as other investigators; therefore, a quantitative comparison is not possible. However, there are few studies related with assessment of geometric, penetration, and scatter components. Dewaraja et al.[12] have also calculated geometric, penetration and scatter components of events within 20% window at 364 keV for point source in air and found 73% of the events in the photopeak window had either scattered or penetrated the collimator. They simulated for NaI (Tl) crystal of size 24 × 40 cm and 0.95 cm crystal thickness, HEGP collimator with square size hole = 3.54 mm, septa = 1 mm, thickness = 5.84 cm.Shafaei et al.[17] simulated for 3/8 inch NaI (Tl) crystal thickness, energy resolution 9.8% at 140 keV, HEGP collimator hexagonal-hole, collimator thickness 5.4 cm, septa thickness 0.16 cm, and found 64% either penetration or scatter in the photopeak window.We have found 32.22% of the event, the 20% photopeak window center at 364 keV either penetrated or scattered in the HEGP collimator (hexagonal-hole, hole diameter = 0.447 cm, collimator thickness 6.60 cm, septa thickness 0.535 cm). And 65.11% of the event the 20% photopeak window center at 364 keV either penetrated or scattered in the MEGP collimator (hexagonal-hole, hole diameter = 0.335 cm, collimator thickness 5.80 cm, septa thickness 0.350 cm).It is important to take into account the effect of backscatter from the PMTs and shielding material surrounding the crystal because these photons might re-enter into the crystal. De Vries et al.[18] have shown that a single slab of 66% Pyrex is sufficient to model backscatter. We have used a single slab of 5 cm thick Pyrex just below the crystal to include the effect of backscatter in the simulation.
CONCLUSION
Low-energy general-purpose and LEHR collimator is best to image 140 keV photon. HEGP can be used for 245 keV and 364 keV; however, correction for penetration and scatter must be applied if one is interested to quantify the in vivo activity of energy 364 keV. Due to heavy penetration and scattering, 511 keV photons should not be imaged with HEGP collimator. The result of this study can be used for optimal collimator design and development of new correction algorithm, because for evaluation and design of scatter correction algorithm, proper understanding of how the scattered photons are distributed and about its properties is required.
Authors: H C Gifford; M A King; R G Wells; W G Hawkins; M V Narayanan; P H Pretorius Journal: IEEE Trans Med Imaging Date: 2000-05 Impact factor: 10.048
Authors: Manoj V Narayanan; Michael A King; P Hendrik Pretorius; Seth T Dahlberg; Frederick Spencer; Ellen Simon; Eric Ewald; Edward Healy; Kirk MacNaught; Jeffrey A Leppo Journal: J Nucl Med Date: 2003-11 Impact factor: 10.057