Alessandro Corzani1, Matteo Ziacchi2, Mauro Biffi2, Igor Diemberger2, Cristian Martignani2, Giuseppe Boriani2.
Abstract
Inappropriate ICD shocks are common adverse events; they are mainly due to supraventricular arrhythmias and secondly are related to noise, undersensing, oversensing, device malfunctions. We present a case of inappropriate device therapy due to myopotential oversensing in a patient with a subcutaneous ICD (s-ICD). A 58 years old male with an s-ICD during the device interrogation showed a previous episode of suspected sustained ventricular tachycardia at 210 bpm, which was effectively treated with ICD shock. The patient experienced the electrical shock while holding a big gas-cylinder in his arms. The EGM analysis revealed many irregular ventricular signals of low amplitude lasting for 24 s and interrupted by the shock. The device showed no malfunctions. This is the first case report of inappropriate S-ICD shock related to myopotential over-sensing. By recording intracardiac EGM, we demonstrated that the noise was created by the activity of the pectorals muscles.Entities:
Keywords: Device inappropriate shocks; Implantable cardioverter defibrillators; Myopotential over-sensing; Subcutaneous ICD; Ventricular tachyarrhythmia