Limitations and late complications of third-generation automatic cardioverter-defibrillators.

S O Nunain1, M Roelke, T Trouton, S Osswald, Y H Kim, G Sosa-Suarez, D R Brooks, B McGovern, M Guy, D F Torchiana.   

Abstract

BACKGROUND: This study examines the limitations and complex management problems associated with the use of tiered-therapy, implantable cardioverter-defibrillators (ICDs). METHODS AND
RESULTS: The study group comprises the first 154 patients undergoing implantation of tiered-therapy ICDs at our institution. Pulse generators from three different manufacturers were used. In 39 patients, a complete nonthoracotomy lead system was used. The perioperative mortality was 1.3%. Of these 154 patients, 37% experienced late postoperative problems. Twenty-one patients required system revision within 36.5 months (mean, 8.57 +/- 11.3) of surgery. Reasons for revision were spurious shocks due to electrode fractures (3) or electrode adapter malfunction (2), inadequate signal from endocardial rate-sensing electrodes (3), superior vena cava or right ventricular coil migration (5), failure to correct tachyarrhythmias due to a postimplant rise in defibrillation threshold (5), or pulse generator failure (3). One of these patients required system removal for infection after revision of an endocardial lead. A further 32 patients received inappropriate shocks for atrial fibrillation with a rapid ventricular response or sinus tachycardia. Two of these patients also received shocks for ventricular tachycardia initiated by antitachycardia pacing triggered by atrial fibrillation. Ventricular pacing for bradycardia was associated with inappropriate shocks due to excessive autogain in 2 patients.
CONCLUSIONS: Despite the major diagnostic and therapeutic advantages of tiered-therapy ICDs, a significant proportion of patients continue to experience hardware-related complications or receive inappropriate shocks.

Entities:  

Mesh:

Substances:

Year:  1995        PMID: 7697850     DOI: 10.1161/01.cir.91.8.2204

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  18 in total

1.  Inappropriate implantable defibrillator discharge caused by a retained pacemaker lead fragment.

Authors:  L Lickfett; C Wolpert; W Jung; S Spehl; L Pizzulli; B Esmailzadeh; B Lüderitz
Journal:  J Interv Card Electrophysiol       Date:  1999-07       Impact factor: 1.900

2.  A 17 year experience of inappropriate shock therapy in patients with implantable cardioverter-defibrillators: are we getting any better?

Authors:  C A Rinaldi; R D Simon; A Baszko; J Bostock; D Elliot; C A Bucknall; J S S Gill
Journal:  Heart       Date:  2004-03       Impact factor: 5.994

Review 3.  Atrial fibrillation in carcinoid heart disease: The role of serotonin. A review of the literature.

Authors:  C Langer; C Piper; J Vogt; J Heintze; T Butz; O Lindner; W Burchert; C Kersting; D Horstkotte
Journal:  Clin Res Cardiol       Date:  2006-11-24       Impact factor: 5.460

4.  Management of patients with implantable cardioverter defibrillators at emergency departments.

Authors:  Chung-Pin Liu; Yi-Lwun Ho; Yen-Hung Lin; Yen-Bin Liu; Wei-Tien Chang; Chien-Hua Huang; Wen-Jone Chen
Journal:  Emerg Med J       Date:  2007-02       Impact factor: 2.740

Review 5.  Endovascular extraction techniques: Part 2: Complications and indications.

Authors:  F A Bracke; A Meijer; B van Gelder
Journal:  Neth Heart J       Date:  2001-05       Impact factor: 2.380

6.  Inappropriate shock for myopotential over-sensing in a patient with subcutaneous ICD.

Authors:  Alessandro Corzani; Matteo Ziacchi; Mauro Biffi; Igor Diemberger; Cristian Martignani; Giuseppe Boriani
Journal:  Indian Heart J       Date:  2015-02-12

7.  Prevalence of sensing abnormalities in dual chamber implantable cardioverter defibrillators.

Authors:  Mohammad Saeed; Anna Jin; Gregory Pontone; Steve Higgins; Michael Gold; David Harari; Steven Nunley; Mark S Link; Munther K Homoud; N A Mark Estes; Paul J Wang
Journal:  Ann Noninvasive Electrocardiol       Date:  2003-07       Impact factor: 1.468

Review 8.  Should all implantable cardioverter defibrillators for ventricular arrhythmias be dual-chamber devices?

Authors:  K L Lee; C P Lau
Journal:  Curr Cardiol Rep       Date:  2001-11       Impact factor: 2.931

9.  Right ventricular rapid pacing in catheter ablation of atrial fibrillation: a novel application for cryoballoon pulmonary vein isolation.

Authors:  K R Julian Chun; Alexander Fürnkranz; Boris Schmidt; Andreas Metzner; Roland Tilz; Thomas Zerm; Ilka Köster; Buelent Koektuerk; Melanie Konstantinidou; Feifan Ouyang; Karl Heinz Kuck
Journal:  Clin Res Cardiol       Date:  2009-06-10       Impact factor: 5.460

10.  [New algorithms for discrimination between supraventricular and ventricular tachyarrhythmias in patients with implantable cardioverter/defibrillator].

Authors:  J Neuzner; M Schlepper
Journal:  Herzschrittmacherther Elektrophysiol       Date:  1997-03
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.