Literature DB >> 25817187

Corrosion and Fretting of a Modular Hip System: A Retrieval Analysis of 60 Rejuvenate Stems.

Ivan De Martino1, Joseph B Assini2, Marcella E Elpers3, Timothy M Wright3, Geoffrey H Westrich2.   

Abstract

Femoral stems with dual-taper modularity were introduced to allow independent control of length, offset, and version. Corrosion and fretting related to micromotion at the neck-stem junction are thought to stimulate an adverse local tissue reaction (ALTR). Analysis of 60 consecutively retrieved modular-neck stem implants (Rejuvenate, Stryker) revised primarily for ALTR was done to determine the variables influencing corrosion and fretting patterns at the neck-stem interface. Taper damage evaluation was performed with stereomicrocopic analysis with two observers. Evidence of fretting and corrosion was seen at the neck-stem taper in all implants, including three implants revised for periprosthetic fractures within four weeks of the index surgery indicating that this process starts early. Femoral stems paired with the long overall neck lengths had significantly higher corrosion scores. Correlation of the corrosion severity at particular locations with the length of implantation suggests that the neck-stem junction experiences cyclic cantilever bending in vivo. The positive correlation between the length of implantation and fretting/corrosion scores bodes poorly for patients who still have this implant. Scanning electron microscopy on a subset of specimens was also performed to evaluate the black corrosion material. We strongly urge frequent follow-up exams for every patient with this particular modular hip stem.
Copyright © 2015 Elsevier Inc. All rights reserved.

Entities:  

Keywords:  complications; dual taper; metal corrosion; modular neck; modular total hip; taper corrosion

Mesh:

Year:  2015        PMID: 25817187     DOI: 10.1016/j.arth.2015.03.010

Source DB:  PubMed          Journal:  J Arthroplasty        ISSN: 0883-5403            Impact factor:   4.757


  22 in total

1.  Revision for taper corrosion at the neck-body junction following total hip arthroplasty: pearls and pitfalls.

Authors:  Mitchell C Weiser; Darwin D Chen
Journal:  Curr Rev Musculoskelet Med       Date:  2016-03

2.  Cementless modular neck stems: are they a safe option in primary total hip arthroplasty?

Authors:  José M Pelayo-de-Tomás; José L Rodrigo-Pérez; Carlos D Novoa-Parra; Alejandro Lizaur-Utrilla; María Morales-Suárez-Varela; José Antonio Blas-Dobón
Journal:  Eur J Orthop Surg Traumatol       Date:  2017-10-29

Review 3.  Corrosion of Metallic Biomaterials: A Review.

Authors:  Noam Eliaz
Journal:  Materials (Basel)       Date:  2019-01-28       Impact factor: 3.623

4.  CORR Insights®: Do Well-functioning THAs Retrieved at Autopsy Exhibit Evidence of Fretting and Corrosion?

Authors:  Ronald Emilio Delanois
Journal:  Clin Orthop Relat Res       Date:  2018-10       Impact factor: 4.176

5.  Spontaneous Fractures of a Modern Modular Uncemented Femoral Stem.

Authors:  Chelsea N Koch; Laura Serrano Mateo; Stephen Kayiaros; Kevin A Cassidy; Seth A Jerabek; Alejandro Gonzalez Della Valle
Journal:  HSS J       Date:  2016-06-21

6.  Comparison of metal ion levels in patients with hip resurfacing versus total hip arthroplasty.

Authors:  Craig W Forsthoefel; Nicholas M Brown; Mark L Barba
Journal:  J Orthop       Date:  2017-08-31

7.  Result of Modular Necks in Primary Total Hip Arthroplasty with a Average Follow-up of Four Years.

Authors:  Jemin Yi; Kye Young Han; Young Jun Nam; Keun Woo Kim
Journal:  Hip Pelvis       Date:  2016-09-30

Review 8.  Metal-on-metal total hip arthroplasty: risk factors for pseudotumours and clinical systematic evaluation.

Authors:  Ming Han Lincoln Liow; Young-Min Kwon
Journal:  Int Orthop       Date:  2016-10-20       Impact factor: 3.075

9.  Fretting-corrosion of CoCr-alloys against TiAl6V4: The importance of molybdenum in oxidative biological environments.

Authors:  M A Wimmer; S Radice; D Janssen; A Fischer
Journal:  Wear       Date:  2021-03-10       Impact factor: 4.695

10.  Pseudotumor recurrence in a post-revision total hip arthroplasty with stem neck modularity: A case report.

Authors:  Bhumit R Desai; Gonzalo E Sumarriva; George F Chimento
Journal:  World J Orthop       Date:  2020-02-18
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.