Literature DB >> 25811461

Instability in the COPD diagnosis upon repeat testing vary with the definition of COPD.

Rogelio Perez-Padilla1, Fernando C Wehrmeister2, Maria Montes de Oca3, Maria Victorina Lopez4, Jose R Jardim5, Adriana Muino4, Gonzalo Valdivia6, Julio Pertuze6, Ana Maria B Menezes2.   

Abstract

BACKGROUND: A low FEV1/FVC from post-bronchodilator spirometry is required to diagnose COPD. Both the FEV1 and the FVC can vary over time; therefore, individuals can be given a diagnosis of mild COPD at one visit, but have normal spirometry during the next appointment, even without an intervention.
METHODS: We analyzed two population-based surveys of adults with spirometry carried out for the same individuals 5-9 years after their baseline examination. We determined the factors associated with a change in the spirometry interpretation from one exam to the next utilizing different criteria commonly used to diagnose COPD.
RESULTS: The rate of an inconsistent diagnosis of mild COPD was 11.7% using FEV1/FVC <0.70, 5.9% using FEV1/FEV6 <the lower limit of the normal range, LLN and 4.1% using the GOLD stage 2-4 criterion. The most important factor associated with diagnostic inconsistency was the closeness of the ratio to the LLN during the first examination. Inconsistency decreased with a lower FEV1.
CONCLUSIONS: Using FEV1/FEV6 <LLN or GOLD stage 2-4 as the criterion for airflow obstruction reduces inconsistencies in the diagnosis of mild COPD. Further improvement could be obtained by defining a borderline zone around the LLN (e.g. plus or minus 0.6 SD), or repeating the test in patients with borderline results.

Entities:  

Mesh:

Substances:

Year:  2015        PMID: 25811461      PMCID: PMC4374954          DOI: 10.1371/journal.pone.0121832

Source DB:  PubMed          Journal:  PLoS One        ISSN: 1932-6203            Impact factor:   3.240


Introduction

Population-based prevalence of poorly reversible airflow obstruction (COPD) has been recently estimated for a variety of countries in Latin America [1] and other continents [2] with standardized methods including post-bronchodilator spirometry. In those and other surveys, it has been clear that the several criteria to define airflow obstruction would produce important differences in COPD prevalence, which would be lower with criteria based on FEV1/FVC or FEV1/FEV6 below the 5th percentile (lower limit of normal, LLN), than with the traditional Global Initiative for Obstructive Lung Disease (GOLD) definition (FEV1/FVC<0.7). The majority of the emphasis has been placed on cross-sectional exams and on the impact of COPD criteria on population prevalence, but consistency of diagnosis is also very important for these persons because false positives label them as having a potentially severe disease and false negatives could lead to denial of useful treatments. Additional complications would derive from inconsistency of diagnosis after serial spirometric tests due to known and expected variations in spirometric results after repeated testing that may change the diagnosis along time[3] adding to systematic changes due to aging, and to worsening or improvements in airflow obstruction. In a population-based survey, COPD prevalence based on FEV1/FEV6 airflow obstruction prevalence [4]. We hypothesized that there would be instability in the designation of abnormal spirometric results depending on the COPD definition used. Therefore, we conducted this study to determine the frequency of change in the diagnosis in the same subjects studied in two population-based surveys, as well as the correlates for these inconsistencies.

Methods

The detailed methods of the PLATINO baseline[5] and follow-up studies [6] are available elsewhere. Between the years 2003 and 2005, population-based surveys were conducted employing standardized methodology in five large Latin-American metropolitan areas: Sao Paulo (Brazil), Mexico City (Mexico), Montevideo (Uruguay), Santiago (Chile), and Caracas (Venezuela). We successfully interviewed 1,000 subjects aged 40 years or older in Sao Paulo, 1,063 in Mexico City, 943 in Montevideo, 1,208 in Santiago, and 1,357 in Caracas. Spirometry testing was performed for 963 (97.9%) subjects in Sao Paulo, 1,000 (98.3%) in Mexico City, 885 (97.1%) in Montevideo, 1,173 (99.8%) in Santiago, and 1,294 (98.4%) in Caracas[1]. The questionnaire is available at the PLATINO website: http://www.platino-alat.org. Ethical approval was obtained from the Institutional Review Boards at all five sites (Instituto Nacional de Enfermedades Respiratorias, Universidad Central de Venezuela, Pontificia Universidad Católica de Chile, Federal University of Pelotas and Federal University of Sao Paulo) for the baseline exam and at the three sites participating in the follow-up exams. All participants signed a written informed consent approved by the Boards. Spirometry was undertaken in individuals who did not present any exclusion criteria (99% of the sample) using an ultrasonic spirometer (EasyOne; ndd Medical Technologies, Zurich, Switzerland). Spirometry was performed before (pre-BD) and 15 minutes after the administration of 200 μg of Salbutamol (post-BD) according to the American Thoracic Society (ATS) criteria of acceptability and reproducibility [7]. The quality control exercises showed that >90% of the tests fulfilled ATS quality criteria [5]. Follow-up studies were conducted in Montevideo, Santiago, and Sao Paulo 5, 6, and 9 years after the baseline surveys, respectively[6]. Only individuals with valid spirometric data at baseline were eligible for a follow-up exam. Individuals were visited at their homes based on the contact information provided by these during the baseline exam. For the purposes of the present study, only individuals with valid postbronchodilator spirometric tests during both surveys were analyzed. Data analyses contained the description of the sample characteristics and the calculation of the intraclass correlation coefficient (ICC Rho) between the spirometric parameters (as a continuous variable) in the two tests and the concordance in the diagnosis of COPD (as dichotomic data using the ICC Rho and the Kappa coefficient) with the following criteria: a) LLN [8-10]—defined as the lower 5th percentile for predicted post-BD FEV1/FVC based on equations derived from the baseline study in a sub-set of healthy and never smoking subjects [11]; b) a ratio of the post-BD FEV1 over FVC < 0.70 according to the Global Initiative for Obstructive Lung Disease (GOLD)[12,13]; c) FEV1/FVC6 11]; d) GOLD stages 2–4 defined as FEV1/FVC<0.7 & FEV1<80% predicted, that has been used increasingly to add specificity, and similar indices based on LLN, e.g. FEV1/FVC & FEV1 The probability of an inconsistent spirometric diagnosis during the second examination (one with COPD and the other with no COPD) versus consistent results (both with the same diagnosis) was estimated from logistic regression models, including as independent variables the deviation of the measured FEV1/FVC and FEV1/FEV6 (in standard deviations [SD] from predicted) from the LLN and, in addition, adjustment by several confounders (forced expiratory time, age, gender, current smoking, cumulative smoking in pack-years, the presence of respiratory symptoms, and previous physician diagnosis of respiratory diseases obtained from a questionnaire). Weight and height were measured, and Body mass index (BMI) was categorized into two groups: normal and overweight/obesity. Although in the present work we do not analyze adverse outcomes in detail[14], we investigated if indices giving less inconsistencies on repeated testing also better predicted risk of death and the risk of two or more exacerbations in the previous year. In the presence of airflow obstruction (for the different definitions) we estimated risk of death during follow-up from proportional hazard models (Cox regression done in all individuals including those who died and did not performed the second spirometry) and risk of more than 2 exacerbations in the previous year with a logistic regression model taking into account age and gender. All of the analyses were adjusted for study site (country) and were stratified by gender.

Results

Table 1 describes the main baseline characteristics of the study participants. Follow-up exams were conducted for 885 adults in Montevideo, 1,173 in Santiago, and 963 in Sao Paulo. Information was obtained for 758 (85.6%), 993 (84.7%) and 748 (77.7%) subjects, respectively, of whom 2,026 had a good quality post-BD spirometry test during both examinations (68.8% of those with post bronchodilator testing in the first evaluation and 75.6% of those with post bronchodilator testing in the first evaluation surviving or lost at the time of the second evaluation) (see Table 2). Follow-up rates for each independent variable category were around 80% [5,6]
Table 1

Characteristics of the 2,026 adults who participated in both examinations and had a valid spirometry test.

First examSecond exam
Mean (%)SD (or 95%CI)Mean (%)SD (or 95%CI)
Men (95%CI)40.639.0–42.2--
Age (years)55.610.962.210.8
Height (cm)160.49.7159.89.8
BMI (Kg/m2)28.35.728.85.3
FEV1 pre-BD (L)2.650.782.420.77
FVC pre-BD (L)3.560.993.240.97
FEV1 post-BD (L)2.740.772.500.77
FVC post-BD (L)3.540.963.260.96
FEV1/FVC pre-BD74.38.274.78.5
FEV1/FVC post-BD77.78.176.88.5
FEV1/FEV6 pre-BD78.16.277.76.9
FEV1/FEV6 post-BD80.46.279.26.8
COV FEV1/FVC pre-BD0.800.37–1.470.780.34–1.44
COV FEV1/FVC post-BD0.670.30–1.250.680.31–1.34
COV FEV1/FEV6 pre-BD0.470.23–0.910.450.20–0.86
COV FEV1/FEV6 post-BD0.420.19–0.760.450.20–0.86
Good quality pre-BD* (95%CI)9089–928381–85
Good quality post-BD* (95%CI)9190–937977–81
FET pre-BD (s)11.73.99.93.9
FET post-BD (s)10.33.49.43.7
History of Asthma (95%CI)16.414.6–18.116.014.1–17.1
History of COPD (95%CI)4.23.3–5.16.25.1–7.3
Current smoker (95%CI)31.128.8–33.525.323.2–27.4

BMI = weight/ height2; COV = Coefficient of variability, expressed as medians, 25th and 75th percentiles. SD = Standard deviation; FET = forced expiratory time; pre-BD = before bronchodilator; post-BD = after Bronchodilator. Good quality = three acceptable tests with two best FEV1 and FVC within <150mL. LLN = lower limit of normal; 95%CI = 95% confidence interval.

Table 2

Population eligible, spirometries done, individuals switching COPD categories and percentage with 2 or more exacerbations according to several definitions of airflow obstruction (COPD).

CriteriaNumber with post BD spirometry at baselineDead at follow upAlive or lost at follow upNumber with post BD spirometry at follow upNumber switching COPD categories (%)Percentage with 2+ exacerbations in previous year at 2nd evaluation (95%CI)
FEV1/FVC<0.7 (GOLD)50296 (19.1)406 (80.9)300 (73.9)91 (30.3)2.4 (1.3–4.2)
 Remaining2440169 (6.9)2271 (93.1)1726 (76.0)146 (8.5)2.1 (1.6–2.8)
FEV1/FVC <LLN27450 (18.2)224 (81.8)167 (74.6)68 (40.7)2.9 (1.5–5.7)
 Remaining2668215 (8.1)2453 (91.9)1859 (75.8)63 (3.4)2.1 (1.6–2.7)
FEV1/FEV6 <LLN25148 (19.1)203 (80.9)149 (73.4)45 (30.2)3.6 (1.9–6.8)
 Remaining2691217 (8.1)2474(91.9)1877 (75.9)86 (4.6)2.0 (1.6–2.7)
GOLD 2–419147 (24.6)144 (75.4)110 (76.4)29 (26.4)4.2 (2.1–8.2)
 Remaining2751218 (7.9)2533 (92.1)1916 (75.6)54 (2.8)2.0 (1.6–2.6)
FEV1/FVC & FEV1<LLN10125 (24.8)76 (75.2)58 (76.3)21 (36)7.9 (4.0–15.1)
 Remaining2841240 (8.4)2601 (91.6)1968 (69.3)23 (1.2)2.0 (1.5–2.6)
FEV1/FEV6 & FEV1<LLN10625 (23.6)81 (76.4)61 (75.3)20 (32.8)7.5 (3.8–14.4)
 Remaining2836240 (8.5)2596 (91.5)1965 (69.3)34 (1.7)2.0 (1.5–2.6)

PostBD = post bronchodilator; Switched categories includes the number and the percentage with reference to previous column spirometry at follow up. 2026 subjects with two spirometries; GOLD 2–4 = FEV1/FVC<0.7 & FEV1<80% predicted; LLN = lower limit of normal, the 5th percentile adjusted for gender, age and height. 95%CI = 95% confidence intervaL

BMI = weight/ height2; COV = Coefficient of variability, expressed as medians, 25th and 75th percentiles. SD = Standard deviation; FET = forced expiratory time; pre-BD = before bronchodilator; post-BD = after Bronchodilator. Good quality = three acceptable tests with two best FEV1 and FVC within <150mL. LLN = lower limit of normal; 95%CI = 95% confidence interval. PostBD = post bronchodilator; Switched categories includes the number and the percentage with reference to previous column spirometry at follow up. 2026 subjects with two spirometries; GOLD 2–4 = FEV1/FVC<0.7 & FEV1<80% predicted; LLN = lower limit of normal, the 5th percentile adjusted for gender, age and height. 95%CI = 95% confidence intervaL Compared with the first examination, individuals with a follow-up exam had less current smoking, and with slightly lower lung function. Mean Forced expiratory time (FET) decreased by about 1 second but was >6 seconds on average, and the rate of valid tests decreased from 90% to 83%. The coefficient of variability of FEV1/FEV6 was lower than that of FEV1/FVC, both before and after BD in both examinations. Fig. 1 depicts a scatter plot of the FEV1/FVC and Fig. 2 the FEV1/FEV6 both expressed as Z-scores of the PLATINO predicted values.
Fig 1

Post-Bronchodilator (BD) FEV1/FVC in the first vs.

the second examination, both expressed as Z- score of the PLATINO predicted values. Lines at -1.645 represent the 5th percentile of the measurement (LLN). Compare with Fig. 2, and observe that dispersion is worse for FEV1/FVC (Rho concordance 0.74) than for FEV1/FEV6 (Rho concordance 0.82).

Fig 2

Post-bronchodilator (BD) FEV1/FEV6 in the first versus the second examination, both expressed as Z- score of the PLATINO predicted values.

Lines at -1.645 represent the 5th percentile of the measurement (LLN). Compare with Fig. 1, and observe that dispersion is worse for FEV1/FVC (Rho concordance 0.74) than for FEV1/FEV6 (Rho concordance 0.82).

Post-Bronchodilator (BD) FEV1/FVC in the first vs.

the second examination, both expressed as Z- score of the PLATINO predicted values. Lines at -1.645 represent the 5th percentile of the measurement (LLN). Compare with Fig. 2, and observe that dispersion is worse for FEV1/FVC (Rho concordance 0.74) than for FEV1/FEV6 (Rho concordance 0.82).

Post-bronchodilator (BD) FEV1/FEV6 in the first versus the second examination, both expressed as Z- score of the PLATINO predicted values.

Lines at -1.645 represent the 5th percentile of the measurement (LLN). Compare with Fig. 1, and observe that dispersion is worse for FEV1/FVC (Rho concordance 0.74) than for FEV1/FEV6 (Rho concordance 0.82). Dispersion is narrower with FEV1/FEV6 (Intraclass correlation coefficient Rho 0.81, 95%CI 0.78–0.84) when compared to the FEV1/FVC (ICC Rho 0.77, 95%CI 0.74–0.79). Reliability for for diagnosis (FEV1/FVC or FEV1/FEV6 as a dichotomic variable) measured similarly by the ICC Rho were much lower numerically than the ICC Rho coefficients for original variables: for GOLD criteria was 0.57 (95%CI 0.54–0.60), for GOLD stages 2–4 was 0.64 (95%CI 0.61.0.66), for FEV1/FEV6 Inconsistent diagnoses occurred with all three criteria, but were more common when utilizing the GOLD criterion. Tables 2 and 3 describes the individuals switching spirometric diagnoses in the two examinations according to the airflow obstruction criteria as a proportion those with and without airflow obstruction (Table 2) or as a proportion of the total population (Table 3). Population inconsistencies were highest for GOLD criterion (11.7%), intermediate for FEV1/FVCairflow obstruction and inconsistency in diagnosis was further reduced using indices requiring a low FEV1 in addition to a low ratio: GOLD-2–4 (4.1%), FEV1/FVC &FEV1airflow obstruction in both evaluations had increased risk of exacerbations compared to those with inconsistent airflow obstruction regardless of the criterion of airflow obstruction. For example those with FEV1/FVCCOPD diagnosis.
Table 3

Prevalences of airflow obstruction, frequency of inconsistent longitudinal diagnosis and hazard ratio depending on the definition.

CriteriaCOPD 195%CICOPD 295%CIInconsistency (%)95%CIHazard ratio*95%CIEXACERBATIONS (OR)95%CI
FEV1/FVC<0.7 (GOLD)14.813.2–16.417.515.9–19.211.710.3–13.11.771.3–2.41.90.9–3.9
FEV1/FVC <LLN8.27.0–9.48.06.8–9.26.55.4–7.51.531.1–2.21.90.8–3.9
FEV1/FEV6 <LLN7.46.2–8.510.28.9–11.55.94.8–6.91.701.2–2.42.51.1–5.0
GOLD 2–45.44.4–6.46.65.5–7.74.13.2–5.02.491.8–3.53.11.4–7.2
FEV1/FVC & FEV1<LLN2.82.0–3.53.52.7–4.32.21.5–2.83.061.9–4.95.42.4–13.1
FEV1/FEV6 & FEV1<LLN3.02.2–3.74.13.2–4.92.72.0–3.42.921.8–4.65.22.2–12.1

COPD 1 and COPD 2 = sample prevalence (as percentage) of COPD in the first and second evaluation in the 2026 subjects with two spirometries; GOLD 2–4 = FEV1/FVC<0.7 & FEV1<80% predicted; LLN = lower limit of normal, the 5th percentile adjusted for gender, age and height. Hazard ratio = the risk of death (n = 2348) from a Cox model adjusting by age and gender based on the presence of airflow obstruction depending on the definition used. Inconsistency is the percentage of participants with a change in diagnosis of airflow obstruction (either a new diagnosis in the second evaluation or the presence of airflow obstruction in the first evaluation but not in the second). Exacerbations = odds ratio of having 2 or more exacerbations in the previous year assessed at the second evaluation in a model adjusted by age and gender. 95%CI = 95% confidence interval.

If inconsistency were expressed as percentage of airflow obstruction prevalence at first evaluation (Table 2 and column 6/column 2 from Table 3), differences among indices would be minor. In addition although prevalence of airflow obstruction may be similar in both evaluations, "abnormal" individuals are not necessarily the same: some of them had a "new diagnosis" at the second evaluation and some "normalized" (see Table 2). For example, from the total of individuals with COPD diagnosis at first examination (see Table 2), 30% would be reverted at second evaluation if using GOLD criteria, 26% if using GOLD stages 2–4, and 33% if using FEV1/FEV6 & FEV1 COPD 1 and COPD 2 = sample prevalence (as percentage) of COPD in the first and second evaluation in the 2026 subjects with two spirometries; GOLD 2–4 = FEV1/FVC<0.7 & FEV1<80% predicted; LLN = lower limit of normal, the 5th percentile adjusted for gender, age and height. Hazard ratio = the risk of death (n = 2348) from a Cox model adjusting by age and gender based on the presence of airflow obstruction depending on the definition used. Inconsistency is the percentage of participants with a change in diagnosis of airflow obstruction (either a new diagnosis in the second evaluation or the presence of airflow obstruction in the first evaluation but not in the second). Exacerbations = odds ratio of having 2 or more exacerbations in the previous year assessed at the second evaluation in a model adjusted by age and gender. 95%CI = 95% confidence interval. A new diagnosis of COPD was observed in 7.3% (GOLD), 3.2% (FEV1/FVCCOPD was observed in 4.4% (GOLD), 3.3% (FEV1/FVCairflow obstruction. The majority of individuals with new airflow diagnosis had a drop of FEV1 in the second evaluation (from 68–82% depending on the criteria) and a minority an increase in FVC or FEV6 (18 to 34%). In the individuals whose airflow obstruction normalized in the second evaluation this was due (70–94%) to a decrease in FVC or FEV6 (the denominator) whereas an increase in FEV1 was observed less frequently (17–48%). A decrease in FEV1 was associated with an increase in BMI, continuous smoking, and use of bronchodilators or corticosteroids whereas an increase in FVC or FEV6 was associated with a decrease in BMI or lack of use or bronchodilators or corticosteroids but predicting variables explained less than 3% of the variation of the change in FVC, FEV6 or FEV1. Therefore >97% of changes in FEV1, FVC or FEV6 were unaccounted for by the collected variables. Incident airflow obstruction was associated consistently with age and continuous smoking regardless of the definition but elimination of airflow obstruction was also associated with increased age. Again the variability of the change in spirometric measurements explained by the models was <3%. If a borderline “buffer zone” is constructed 0.6 SD above and below the LLN of FEV1/FEV6, this would include 68% of measurements during the first examination and 63% of the measurements during the second examination. None of the individuals would have changed from no COPD in the first examination to COPD in the second examination (performed 5–9 years later), and only four individuals would have changed from COPD in the first exam to no COPD during the second exam. Therefore, nearly all of the changes in diagnosis would fell within the borderline zone. In addition, in this large borderline zone, only 7/1383 individuals (0.5%) had an FEV1<80% predicted (GOLD stage 2) and none had an FEV1 <50% predicted (GOLD stages 3 or 4). The obstructed category was 16% of the total, and included all GOLD stage 3–4 individuals whereas the normal category was also 16% of the total. The borderline zone could be constructed of different sizes, balancing the proportion of false COPD positives and the proportion of individuals with borderline results. In the logistic regression model (table 4), inconsistency was associated very powerfully with absolute distance of FEV1/FVC or FEV1/FEV6 (expressed as Z- scores) from the LLN. That is, the closer the measured FEV1/FVC or FEV1/FEV6 to the LLN, the higher the probability of an inconsistent diagnosis during the second examination. For example, the Odds ratio (OR) for an inconsistent diagnosis was 12 (95%CI 8–18) for FEV1/FVCparticipants. Also, the lower the FEV1, the lower the inconsistency (OR 0.65–0.68 per SD from normality, see Table 5 and Fig. 3).
Table 4

Predicting inconsistency in airflow obstruction diagnosis by a logistic regression model.

Inconsistency in FEV1/FVC<LLNInconsistency in FEV1/FEV6<LLN
Odds Ratio95% CIPOdds Ratio95% CIP
Distance of FEV1/FVC or FEV1/FEV6 from the LLN (standard deviations)
 2–37.80.9–640.060.90.2–4.80.9
 1 <29.61.2–760.032.80.6–130.17
 0.6 <139.15–312<0.00008.51.9–380.005
 <0.610814–842<0.0000175.818–319<0.00001
FEV1 post BD as score Z0.680.55–0.85<0.000010.750.60–0.960.02
FET (s) first evaluation1.21.1–1.30.0011.061.0–1.130.04
FET (s) second evaluation1.11.03–1.1<0.00011.020.97–1.070.45
Pseudo R2 0.330.38

LLN = lower limit of normal, 5th percentile of the ratio. FET = forced expiratory time; Model was also adjusted by age, gender, asthma, current smoking at second evaluation, a physician diagnosis of asthma. PostBD = post bronchodilator. The closer the ratio and the FEV1 to the diagnostic threshold, the higher the risk of inconsistency on repeated testing.

Table 5

Inconsistencies in the COPD diagnosis (post bronchodilator spirometry) and some of the predisposing factors.

Same participants in two surveys 5–9 years apart.

CategoryPercentage95%CI
FEV 1 /FVC<0.70 (GOLD) INCONSISTENT 11.710.3–13.1
 New GOLD7.36.0–8.5
  Drop in FEV1  67.8 60–75
  Continue smoking 27.4 20–35
  Increase in FVC 53.4 45–62
  Use of inhaled medications 9.6 5–14
   Physician diagnosed asthma 8.0 5–11
 Reversed GOLD 4.4 3.5–5.3
  Increase in FEV1  25.3 16–34
  Use of inhaled medications 7.8 2–13
  Quit smoking 29.1 10–49
  Drop in FVC 94.5 90–99
  Increase in weight 74.7 66–84
   Physician diagnosed asthma 5.6 4–9
FEV 1 /FEV 6 <LLN INCONSISTENT 5.94.8–6.9
 New FEV1/FEV6 <LLN4.43.4–5.2
  Drop in FEV1  81.9 72–92
  Continue smoking 40.0 27–53
  Increase in FVC 34.4 22–47
  Use of inhaled medications 13.1 4–21
   Physician diagnosed asthma 7.1 5–10
 Reversed FEV1/FEV6 <LLN1.81.2–2.4
  Increase in FEV1  19.4 6–33
  Use of inhaled medications 16.6 4–29
   Quit smoking 25.0 1–63
  Drop in FVC 91.7 82–100
  Increase in weight 66.6 50–82
   Physician diagnosed asthma 3.5 2–6
FEV 1 /FVC<LLN INCONSISTENT 6.55.4–7.5
 New FEV1/FVC<LLN3.22.2–4.0
  Drop in FEV1  69.8 58–81
  Continue smoking 32.3 20–44
  Increase in FVC 63.4 51–76
  Use of inhaled medications 11.1 3–19
   Physician diagnosed asthma 4.4 2.7–7.2
 Reversed FEV1/FVC<LLN3.32.5–4.1
  Increase in FEV1  17.6 8–27
  Use of inhaled medications 13.2 5–21
   Quit smoking 31 11–53
  Drop in FVC 94.1 88–99
  Increase in weight 72.1 61–83
  Physician diagnosed asthma 5.6 4–9

Percentages on the left part of columns are from the total studied population, whereas those on the right are strata (overlapping) from the above categories. 95%CI = 95% confidence interval; Inconsistent = different result between the two examinations. New GOLD = fulfilling FEV1/FVC<0.70 only during the second exam (similar for the other airflow obstruction criteria). Reversed GOLD = met GOLD criterion during the first examination, but not during the second examination; LLN = lower limit of normal

Fig 3

Cancelled diagnosis of airflow obstruction as percentage (vertical axis) decrease with lower FEV1 (horizontal axis) regardless of the criteria used.

The graph depicts the results for three common criteria: GOLD (dashed line), FEV1/FVC

LLN = lower limit of normal, 5th percentile of the ratio. FET = forced expiratory time; Model was also adjusted by age, gender, asthma, current smoking at second evaluation, a physician diagnosis of asthma. PostBD = post bronchodilator. The closer the ratio and the FEV1 to the diagnostic threshold, the higher the risk of inconsistency on repeated testing.

Inconsistencies in the COPD diagnosis (post bronchodilator spirometry) and some of the predisposing factors.

Same participants in two surveys 5–9 years apart. Percentages on the left part of columns are from the total studied population, whereas those on the right are strata (overlapping) from the above categories. 95%CI = 95% confidence interval; Inconsistent = different result between the two examinations. New GOLD = fulfilling FEV1/FVC<0.70 only during the second exam (similar for the other airflow obstruction criteria). Reversed GOLD = met GOLD criterion during the first examination, but not during the second examination; LLN = lower limit of normal

Cancelled diagnosis of airflow obstruction as percentage (vertical axis) decrease with lower FEV1 (horizontal axis) regardless of the criteria used.

The graph depicts the results for three common criteria: GOLD (dashed line), FEV1/FVC A longer forced expiratory time was also associated with inconsistent diagnosis (OR 1.06–1.22), and a history of wheezing or asthma in some models, but usually with a borderline statistical significance. Other variables analyzed were age, continuous smoking, BMI, gender, clinical diagnosis of COPD, use of respiratory medications and the duration of the expiratory maneuvers, but all together they explained about 7–17% of the variability of inconsistent diagnosis, and the statistical significance disappeared (except for forced expiratory time) once we included in the models the proximity of the spirometry measurements to the threshold for airflow obstruction and FEV1. The closeness of FEV1/FVC or FEV1/FEV6 to the LLN was the main predictor of switching diagnosis in the two evaluations whether it was a new airflow obstruction in the second evaluation or an airflow obstruction found in the first evaluation disappearing in the second (see S2 Table with information for FEV1/FEV6).

Discussion

We have described the rates and correlates of inconsistent interpretations of airway obstruction in adults during population-based follow-up examinations according to several criteria of airflow obstruction. There were three important findings. Firstly, the inconsistency of COPD diagnosis was observed with all criteria because these are based on sharp cut-off points (the LLN or a FEV1/FVC<0.7 for the GOLD criteria plus FEV1<80% of predicted for GOLD stage 2–4 or FEV1COPD or FEV1 for GOLD stages). Utilizing the GOLD criterion, an individual with an FEV1/FVC of 0.69 would be diagnosed with COPD, and during a follow-up exam, if the ratio were 0.7, the subject would no longer be diagnosed as having COPD. Intraclass correlation coefficient, a measure of concordance was lower numerically for COPD diagnosis than for FEV1/FVC or FEV1/FEV6 as continuous variables. The same problem exists with strict cut-off values based on LLN. Although certain inconsistency is expected, the recommended diagnosis of COPD assumes a single spirometry and therefore does not consider longitudinal inconsistency. Secondly, airflow obstruction criteria leading to lower population prevalence also produce less inconsistency on repeated testing: the FEV1/FVC had less concordance in the two tests than FEV1/FEV6 whereas the consistency of COPD diagnosis was highest for FEV1/FEV6 airflow obstruction requiring a low FEV1 (GOLD stage 2–4). Thirdly, the closeness of the FEV1/FVC, FEV1/FEV6 and FEV1 to the thresholds or cut points, the higher the possibility of a change in diagnosis on repeated testing (Table 5) independently if the switch represents a new airflow obstruction or a "reversed" airflow obstruction (S2 Table). Previous studies have shown that the within-test reproducibility of FEV1/FEV6 based on 6-second maneuvers was smaller when compared with the traditional FEV1/FVC [9]. We previously reported that utilizing FEV1/FEV6 COPD prevalence in the same populations as criteria based on FEV1/FVC (GOLD or COPD in other words, by creating a buffer zone immediately around the LLN, as suggested previously by the National Lung Health Education Program (NLHEP) in the United States [15]. This zone would reduce or avoid the changes in diagnosis from extreme groups i.e. from COPD to normal, or from normal to COPD, instead forcing the majority of changes to occur from and to the borderline zone. This strategy would increase confidence in spirometric diagnosis and would be relatively easy to implement compared to a more sound approach of keeping as continuous FEV1/FVC, FEV1/FEV6 and FEV1 and estimating risks of disease or adverse outcomes or likelihood of response to treatments. Clinicians usually possess information on individual patients that is not available to Epidemiologists and therefore can use this information to estimate the pre-test disease probability. They can then consider the magnitude of the abnormality from the spirometry test results in order to estimate the post-test probability of COPD. They can also order follow-up tests that will further increase or decrease the probability of COPD, such as a Diffusing capacity (DLCO) test, an exhaled nitric oxide test, induced sputum for eosinophil counts, or a lung Computed tomography scan (looking for emphysema). Clinicians can also perform an asthma therapy trial to determine whether the patient responds, rendering asthma much more likely than COPD. Clinical practice guidelines for hypertension recommend that prior to prescribing blood pressure medication, 3–6 repeated measurements of blood pressure over weeks to months be performed [16]. Perhaps a similar conservative approach should be taken with spirometry for COPD, especially if the results fall within the borderline zone. For these borderline cases, a conservative approach seems adequate because a false COPD label entails costs and potential adverse effects for the patient, including unwarranted bronchodilator prescription or further expensive tests. On the other hand, missing a case of mild COPD has little consequence, as all smokers should be advised to quit smoking regardless of spirometric results, and the majority of drug treatments are meant for moderate or severe COPD. The main strengths of this study include the population-based sampling, high quality of the spirometry tests (15), the measurement of post-BD spirometry during both examinations and the relatively high rates of follow-up after 6–9 years. Limitations of our study include the following: A lack of annual spirometry testing, which could have determined whether or not study participants were “rapid decliners” consistent with COPD progression [15]. Shifts from airway obstruction to normal spirometry during the 5–9 year period could have occurred due to factors that we did not measure, such as respiratory infections or exposure to irritants during the first examinations (but not during the second examination), smoking cessation, or treatment for asthma, COPD, or heart failure; or reductions in exposures causing bronchospasm at the time of the first examination. However, an inconsistent airflow obstruction diagnosis, with any definition was associated significantly with the proximity to the limit of normality and the expiratory time of the tests and therefore likely mostly determined by intrinsic properties of the tests (reliability) rather than true pathobiological processes. Finally, in the current work we did not analyze in detail adverse outcomes, but thresholds giving less prevalence of airflow obstruction, also give less inconsistencies and are associated to a greater risk of death and of exacerbations. In conclusion, current criteria for airflow obstruction, even those based on age- and gender- adjusted FEV1/FVC or FEV1/FEV6 COPD diagnosis more likely if tests are around the threshold defining airflow obstruction. Diagnosing airflow obstruction using the 6-second spirometry (FEV1/FEV6 patient to weigh the consequences of a false-positive versus a false-negative disease label before making clinical decisions.

Dataset in Excel (XLS) format.

(XLS) Click here for additional data file.

Reliability by the intraclass correlation coefficient (ICC) and 95%CI between two spirometry tests (at baseline and follow-up) in the same individuals.

(DOCX) Click here for additional data file.

Determinants of new and reversed airflow obstruction diagnosis by a logistic regression model using as criteria FEV1/FEV6 (DOCX) Click here for additional data file.

  15 in total

Review 1.  Office spirometry for lung health assessment in adults: A consensus statement from the National Lung Health Education Program.

Authors:  G T Ferguson; P L Enright; A S Buist; M W Higgins
Journal:  Chest       Date:  2000-04       Impact factor: 9.410

2.  [Spirometric reference values in 5 large Latin American cities for subjects aged 40 years or over].

Authors:  Rogelio Pérez-Padilla; Gonzalo Valdivia; Adriana Muiño; María Victorina López; María Nelly Márquez; María Montes de Oca; Carlos Tálamo; Carmen Lisboa; Julio Pertuzé; José Roberto B Jardim; Ana María B Menezes
Journal:  Arch Bronconeumol       Date:  2006-07       Impact factor: 4.872

3.  A population-based cohort study on chronic obstructive pulmonary disease in Latin America: methods and preliminary results. The PLATINO Study Phase II.

Authors:  Ana Maria Baptista Menezes; Adriana Muiño; Maria Victorina López-Varela; Gonzalo Valdivia; Carmen Lisboa; José Roberto Jardim; Maria Montes de Oca; Carlos Tálamo; Fernando César Wehrmeister; Rogelio Perez-Padilla
Journal:  Arch Bronconeumol       Date:  2013-12-12       Impact factor: 4.872

Review 4.  Global strategy for the diagnosis, management, and prevention of chronic obstructive pulmonary disease. NHLBI/WHO Global Initiative for Chronic Obstructive Lung Disease (GOLD) Workshop summary.

Authors:  R A Pauwels; A S Buist; P M Calverley; C R Jenkins; S S Hurd
Journal:  Am J Respir Crit Care Med       Date:  2001-04       Impact factor: 21.405

5.  Standardization of Spirometry, 1994 Update. American Thoracic Society.

Authors: 
Journal:  Am J Respir Crit Care Med       Date:  1995-09       Impact factor: 21.405

6.  Chronic obstructive pulmonary disease in five Latin American cities (the PLATINO study): a prevalence study.

Authors:  Ana Maria B Menezes; Rogelio Perez-Padilla; José Roberto B Jardim; Adriana Muiño; Maria Victorina Lopez; Gonzalo Valdivia; Maria Montes de Oca; Carlos Talamo; Pedro C Hallal; Cesar G Victora
Journal:  Lancet       Date:  2005-11-26       Impact factor: 79.321

7.  International variation in the prevalence of COPD (the BOLD Study): a population-based prevalence study.

Authors:  A Sonia Buist; Mary Ann McBurnie; William M Vollmer; Suzanne Gillespie; Peter Burney; David M Mannino; Ana M B Menezes; Sean D Sullivan; Todd A Lee; Kevin B Weiss; Robert L Jensen; Guy B Marks; Amund Gulsvik; Ewa Nizankowska-Mogilnicka
Journal:  Lancet       Date:  2007-09-01       Impact factor: 79.321

8.  Impact of bronchodilator use on the prevalence of COPD in population-based samples.

Authors:  Rogelio Pérez-Padilla; Pedro Curi Hallal; Juan Carlos Vázquez-García; Adriana Muiño; María Máquez; María Victorina López; María Montes de Oca; Carlos Tálamo; Gonzalo Valdivia; Julio Pertuzé; Jose Jardim; Ana Maria B Menezes
Journal:  COPD       Date:  2007-06       Impact factor: 2.409

9.  Comparison of spirometry criteria for the diagnosis of COPD: results from the BOLD study.

Authors:  W M Vollmer; Thorn Gíslason; P Burney; P L Enright; A Gulsvik; A Kocabas; A S Buist
Journal:  Eur Respir J       Date:  2009-05-21       Impact factor: 16.671

10.  Reliability of FEV1/FEV6 to diagnose airflow obstruction compared with FEV1/FVC: the PLATINO longitudinal study.

Authors:  Rogelio Perez-Padilla; Fernando C Wehrmeister; Bartolome R Celli; Maria Victorina Lopez-Varela; Maria Montes de Oca; Adriana Muiño; Carlos Talamo; Jose R Jardim; Gonzalo Valdivia; Carmen Lisboa; Ana Maria B Menezes
Journal:  PLoS One       Date:  2013-08-01       Impact factor: 3.240

View more
  6 in total

1.  Increased Airway Wall Thickness is Associated with Adverse Longitudinal First-Second Forced Expiratory Volume Trajectories of Former World Trade Center workers.

Authors:  Rafael E de la Hoz; Xiaoyu Liu; John T Doucette; Anthony P Reeves; Laura A Bienenfeld; Juan P Wisnivesky; Juan C Celedón; David A Lynch; Raúl San José Estépar
Journal:  Lung       Date:  2018-05-24       Impact factor: 2.584

2.  One-off Spirometry Is Insufficient to Rule In or Rule Out Mild to Moderate Chronic Obstructive Pulmonary Disease.

Authors:  Shumonta Quaderi; John R Hurst
Journal:  Am J Respir Crit Care Med       Date:  2017-08-01       Impact factor: 21.405

3.  Spirometry is not enough to diagnose COPD in epidemiological studies: a follow-up study.

Authors:  Elena Andreeva; Marina Pokhaznikova; Anatoly Lebedev; Irina Moiseeva; Olga Kuznetsova; Jean-Marie Degryse
Journal:  NPJ Prim Care Respir Med       Date:  2017-11-14       Impact factor: 2.871

4.  The PLATINO study: description of the distribution, stability, and mortality according to the Global Initiative for Chronic Obstructive Lung Disease classification from 2007 to 2017.

Authors:  Ana M Menezes; Fernando C Wehrmeister; Rogelio Perez-Padilla; Karynna P Viana; Claudia Soares; Hana Müllerova; Gonzalo Valdivia; José R Jardim; Maria Montes de Oca
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2017-05-18

5.  Reliability and usefulness of spirometry performed during admission for COPD exacerbation.

Authors:  Alberto Fernández-Villar; Cristina Represas-Represas; Cecilia Mouronte-Roibás; Cristina Ramos-Hernández; Ana Priegue-Carrera; Sara Fernández-García; José Luis López-Campos
Journal:  PLoS One       Date:  2018-03-26       Impact factor: 3.240

Review 6.  Brazilian studies on pulmonary function in COPD patients: what are the gaps?

Authors:  Agnaldo José Lopes; Pedro Lopes de Melo
Journal:  Int J Chron Obstruct Pulmon Dis       Date:  2016-07-11
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.