Observer performance for adaptive, image-based denoising and filtered back projection compared to scanner-based iterative reconstruction for lower dose CT enterography.

Joel G Fletcher1, Amy K Hara, Jeff L Fidler, Alvin C Silva, John M Barlow, Rickey E Carter, Adam Bartley, Maria Shiung, David R Holmes, Nicolas K Weber, David H Bruining, Lifeng Yu, Cynthia H McCollough.   

Abstract

PURPOSE: The purpose of this study was to compare observer performance for detection of intestinal inflammation for low-dose CT enterography (LD-CTE) using scanner-based iterative reconstruction (IR) vs. vendor-independent, adaptive image-based noise reduction (ANLM) or filtered back projection (FBP).
METHODS: Sixty-two LD-CTE exams were performed. LD-CTE images were reconstructed using IR, ANLM, and FBP. Three readers, blinded to image type, marked intestinal inflammation directly on patient images using a specialized workstation over three sessions, interpreting one image type/patient/session. Reference standard was created by a gastroenterologist and radiologist, who reviewed all available data including dismissal Gastroenterology records, and who marked all inflamed bowel segments on the same workstation. Reader and reference localizations were then compared. Non-inferiority was tested using Jackknife free-response ROC (JAFROC) figures of merit (FOM) for ANLM and FBP compared to IR. Patient-level analyses for the presence or absence of inflammation were also conducted.
RESULTS: There were 46 inflamed bowel segments in 24/62 patients (CTDIvol interquartile range 6.9-10.1 mGy). JAFROC FOM for ANLM and FBP were 0.84 (95% CI 0.75-0.92) and 0.84 (95% CI 0.75-0.92), and were statistically non-inferior to IR (FOM 0.84; 95% CI 0.76-0.93). Patient-level pooled confidence intervals for sensitivity widely overlapped, as did specificities. Image quality was rated as better with IR and AMLM compared to FBP (p < 0.0001), with no difference in reading times (p = 0.89).
CONCLUSIONS: Vendor-independent adaptive image-based noise reduction and FBP provided observer performance that was non-inferior to scanner-based IR methods. Adaptive image-based noise reduction maintained or improved upon image quality ratings compared to FBP when performing CTE at lower dose levels.

Entities:  

Mesh:

Year:  2015        PMID: 25725794      PMCID: PMC4450115          DOI: 10.1007/s00261-015-0384-1

Source DB:  PubMed          Journal:  Abdom Imaging        ISSN: 0942-8925


  29 in total

1.  Endoscopic skipping of the distal terminal ileum in Crohn's disease can lead to negative results from ileocolonoscopy.

Authors:  Sunil Samuel; David H Bruining; Edward V Loftus; Brenda Becker; Joel G Fletcher; Jayawant N Mandrekar; Alan R Zinsmeister; William J Sandborn
Journal:  Clin Gastroenterol Hepatol       Date:  2012-04-10       Impact factor: 11.382

2.  Adaptive nonlocal means filtering based on local noise level for CT denoising.

Authors:  Zhoubo Li; Lifeng Yu; Joshua D Trzasko; David S Lake; Daniel J Blezek; Joel G Fletcher; Cynthia H McCollough; Armando Manduca
Journal:  Med Phys       Date:  2014-01       Impact factor: 4.071

3.  The impact of abdominal computed tomography in a tertiary referral centre emergency department on the management of patients with inflammatory bowel disease.

Authors:  E Israeli; S Ying; B Henderson; J Mottola; T Strome; C N Bernstein
Journal:  Aliment Pharmacol Ther       Date:  2013-07-09       Impact factor: 8.171

4.  EMerging BiomARKers in Inflammatory Bowel Disease (EMBARK) study identifies fecal calprotectin, serum MMP9, and serum IL-22 as a novel combination of biomarkers for Crohn's disease activity: role of cross-sectional imaging.

Authors:  William A Faubion; Joel G Fletcher; Sharon O'Byrne; Brian G Feagan; Willem Js de Villiers; Bruce Salzberg; Scott Plevy; Deborah D Proctor; John F Valentine; Peter D Higgins; Jeffrey M Harris; Lauri Diehl; Lilyan Wright; Gaik Wei Tew; Diana Luca; Karen Basu; Mary E Keir
Journal:  Am J Gastroenterol       Date:  2013-10-15       Impact factor: 10.864

Review 5.  CT enterography: concepts and advances in Crohn's disease imaging.

Authors:  Mahmoud M Al-Hawary; Ravi K Kaza; Joel F Platt
Journal:  Radiol Clin North Am       Date:  2013-01       Impact factor: 2.303

6.  Methods for clinical evaluation of noise reduction techniques in abdominopelvic CT.

Authors:  Eric C Ehman; Lifeng Yu; Armando Manduca; Amy K Hara; Maria M Shiung; Dayna Jondal; David S Lake; Robert G Paden; Daniel J Blezek; Michael R Bruesewitz; Cynthia H McCollough; David M Hough; Joel G Fletcher
Journal:  Radiographics       Date:  2014 Jul-Aug       Impact factor: 5.333

7.  Effect of reduced radiation exposure and iterative reconstruction on detection of low-contrast low-attenuation lesions in an anthropomorphic liver phantom: an 18-reader study.

Authors:  Ajit H Goenka; Brian R Herts; Nancy A Obuchowski; Andrew N Primak; Frank Dong; Wadih Karim; Mark E Baker
Journal:  Radiology       Date:  2014-03-10       Impact factor: 11.105

8.  Automatic selection of tube potential for radiation dose reduction in vascular and contrast-enhanced abdominopelvic CT.

Authors:  Lifeng Yu; Joel G Fletcher; Katharine L Grant; Rickey E Carter; David M Hough; John M Barlow; Terri J Vrtiska; Eric E Williamson; Phillip M Young; Brian C Goss; Maria Shiung; Shuai Leng; Ranier Raupach; Bernhard Schmidt; Thomas Flohr; Cynthia H McCollough
Journal:  AJR Am J Roentgenol       Date:  2013-08       Impact factor: 3.959

9.  Prospective evaluation of prior image constrained compressed sensing (PICCS) algorithm in abdominal CT: a comparison of reduced dose with standard dose imaging.

Authors:  Meghan G Lubner; Perry J Pickhardt; David H Kim; Jie Tang; Alejandro Munoz del Rio; Guang-Hong Chen
Journal:  Abdom Imaging       Date:  2015-01

10.  Reducing radiation dose in CT enterography.

Authors:  Andrew J Del Gaizo; Joel G Fletcher; Lifeng Yu; Robert G Paden; Garrett Clay Spencer; Shuai Leng; Annelise M Silva; Jeff L Fidler; Alvin C Silva; Amy K Hara
Journal:  Radiographics       Date:  2013 Jul-Aug       Impact factor: 5.333

View more
  7 in total

1.  Impact of number of repeated scans on model observer performance for a low-contrast detection task in computed tomography.

Authors:  Chi Ma; Lifeng Yu; Baiyu Chen; Christopher Favazza; Shuai Leng; Cynthia McCollough
Journal:  J Med Imaging (Bellingham)       Date:  2016-05-26

Review 2.  Small Bowel Imaging: an Update.

Authors:  Jordi Rimola; Julián Panés
Journal:  Curr Gastroenterol Rep       Date:  2016-07

3.  Effect of Radiation Dose Reduction and Reconstruction Algorithm on Image Noise, Contrast, Resolution, and Detectability of Subtle Hypoattenuating Liver Lesions at Multidetector CT: Filtered Back Projection versus a Commercial Model-based Iterative Reconstruction Algorithm.

Authors:  Justin Solomon; Daniele Marin; Kingshuk Roy Choudhury; Bhavik Patel; Ehsan Samei
Journal:  Radiology       Date:  2017-02-07       Impact factor: 11.105

Review 4.  Advanced imaging techniques for small bowel Crohn's disease: what does the future hold?

Authors:  Inês Pita; Fernando Magro
Journal:  Therap Adv Gastroenterol       Date:  2018-02-12       Impact factor: 4.409

5.  Application of Vendor-Neutral Iterative Reconstruction Technique to Pediatric Abdominal Computed Tomography.

Authors:  Woo Hyeon Lim; Young Hun Choi; Ji Eun Park; Yeon Jin Cho; Seunghyun Lee; Jung Eun Cheon; Woo Sun Kim; In One Kim; Jong Hyo Kim
Journal:  Korean J Radiol       Date:  2019-09       Impact factor: 3.500

6.  Low-dose CT image and projection dataset.

Authors:  Taylor R Moen; Baiyu Chen; David R Holmes; Xinhui Duan; Zhicong Yu; Lifeng Yu; Shuai Leng; Joel G Fletcher; Cynthia H McCollough
Journal:  Med Phys       Date:  2020-12-16       Impact factor: 4.071

7.  Is the Mixed Use of Magnetic Resonance Enterography and Computed Tomography Enterography Adequate for Routine Periodic Follow-Up of Bowel Inflammation in Patients with Crohn's Disease?

Authors:  Jiyeon Ha; Seong Ho Park; Jung Hee Son; Ji Hun Kang; Byong Duk Ye; So Hyun Park; Bohyun Kim; Sang Hyun Choi; Sang Hyoung Park; Suk-Kyun Yang
Journal:  Korean J Radiol       Date:  2021-09-13       Impact factor: 3.500

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.