Localization of pacing and defibrillator leads using standard x-ray views is frequently inaccurate and is not reproducible.

Larry R Jackson1, Jonathan P Piccini, James P Daubert, Lynne M Hurwitz Koweek, Brett D Atwater.   

Abstract

PURPOSE: While the accuracy of roentography for evaluation of lead tip position compared with three-dimensional imaging techniques has been well described, there remains considerable variability in the interpretation of the reproducibility of standard x-ray for right ventricular (RV) and left ventricular (LV) lead position. The aim of this study was to evaluate the accuracy and reliability of right ventricular (RV) and left ventricular (LV) lead tip position as determined by board-certified cardiac electrophysiologists (EP) using standard x-ray views.
METHODS: EP interpretations of RV and LV lead tip position using standard x-ray views (posterior-anterior, lateral, and left anterior oblique) were compared to thoracic computed tomography (CT). The accuracy of x-ray interpretation was compared to the reference CT location, and the reproducibility of x-ray interpretation was tested using the free-marginal Kappa statistic.
RESULTS: A total of 58 EPs were invited to participate in the survey with a response rate of 43 % (25/58). The agreement between x-ray and CT lead tip position (accuracy) was 37 % for RV lead, 33 % for longitudinal LV lead, and 41 % for short axis LV lead. Reproducibility was 64 % for RV lead tip (k = 0.46), 58 % for longitudinal LV lead tip (k = 0.37), and 39 % for short axis LV lead tip (k = 0.24).
CONCLUSIONS: Conventional roentography is limited in its ability to accurately and reliably determine pacing lead tip position.

Entities:  

Mesh:

Year:  2015        PMID: 25715917     DOI: 10.1007/s10840-015-9984-5

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  18 in total

1.  The impact of the right ventricular lead position on response to cardiac resynchronization therapy.

Authors:  Fakhar Z Khan; Pegah Salahshouri; Rudy Duehmke; Philip A Read; Peter J Pugh; Maros Elsik; David Begley; Simon P Fynn; David P Dutka; Munmohan S Virdee
Journal:  Pacing Clin Electrophysiol       Date:  2011-01-05       Impact factor: 1.976

2.  Effect of right ventricular pacing lead on left ventricular dyssynchrony in patients receiving cardiac resynchronization therapy.

Authors:  Leyla Elif Sade; Ozlem Demir; Ilyas Atar; Haldun Müderrisoğlu; Bülent Ozin
Journal:  Am J Cardiol       Date:  2009-01-12       Impact factor: 2.778

3.  Long-term impact of right ventricular septal versus apical pacing on left ventricular synchrony and function in patients with second- or third-degree heart block.

Authors:  Arnold C T Ng; Christine Allman; Jane Vidaic; Hui Tie; Andrew P Hopkins; Dominic Y Leung
Journal:  Am J Cardiol       Date:  2009-02-21       Impact factor: 2.778

4.  Pacing from the right ventricular septum: is there a danger to the coronary arteries?

Authors:  Andrew W Teh; Caroline Medi; Raphael Rosso; Geoffrey Lee; Ronen Gurvitch; Harry G Mond
Journal:  Pacing Clin Electrophysiol       Date:  2009-07       Impact factor: 1.976

5.  Pacing of the interventricular septum versus the right ventricular apex: a prospective, randomized study.

Authors:  Giulia Domenichini; Henri Sunthorn; Eric Fleury; Huberdine Foulkes; Carine Stettler; Haran Burri
Journal:  Eur J Intern Med       Date:  2012-04-11       Impact factor: 4.487

6.  Comparison of tools and techniques for implanting pacemaker leads on the ventricular mid-septum.

Authors:  Haran Burri; Giulia Domenichini; Henri Sunthorn; Vincent Ganière; Carine Stettler
Journal:  Europace       Date:  2011-12-26       Impact factor: 5.214

7.  Long-term effect of right ventricular pacing on myocardial perfusion and function.

Authors:  H F Tse; C P Lau
Journal:  J Am Coll Cardiol       Date:  1997-03-15       Impact factor: 24.094

8.  Assessment of the post-implant final left ventricular lead position: a comparative study between radiographic and angiographic modalities.

Authors:  Prabhat Kumar; Dan Blendea; Veena Nandigam; Stephanie A Moore; E Kevin Heist; Jagmeet P Singh
Journal:  J Interv Card Electrophysiol       Date:  2010-06-17       Impact factor: 1.900

9.  Dual-chamber pacing or ventricular backup pacing in patients with an implantable defibrillator: the Dual Chamber and VVI Implantable Defibrillator (DAVID) Trial.

Authors:  Bruce L Wilkoff; James R Cook; Andrew E Epstein; H Leon Greene; Alfred P Hallstrom; Henry Hsia; Steven P Kutalek; Arjun Sharma
Journal:  JAMA       Date:  2002-12-25       Impact factor: 56.272

10.  Advanced image fusion to overlay coronary sinus anatomy with real-time fluoroscopy to facilitate left ventricular lead implantation in CRT.

Authors:  Simon G Duckett; Matthew R Ginks; Benjamin R Knowles; Yingliang Ma; Anoop Shetty; Julian Bostock; Michael Cooklin; Jas S Gill; Gerry S Carr-White; Reza Razavi; Tobias Schaeffter; Kawal S Rhode; C Aldo Rinaldi
Journal:  Pacing Clin Electrophysiol       Date:  2010-10-28       Impact factor: 1.976

View more
  1 in total

1.  Common sense or evidence: an optimal place for right and left ventricular leads?

Authors:  I Eli Ovsyshcher
Journal:  J Interv Card Electrophysiol       Date:  2015-04-16       Impact factor: 1.900

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.