Literature DB >> 21208234

The impact of the right ventricular lead position on response to cardiac resynchronization therapy.

Fakhar Z Khan1, Pegah Salahshouri, Rudy Duehmke, Philip A Read, Peter J Pugh, Maros Elsik, David Begley, Simon P Fynn, David P Dutka, Munmohan S Virdee.   

Abstract

INTRODUCTION: Left ventricular (LV) lead placement to the latest contracting area (concordant LV lead) is associated with better response to cardiac resynchronization therapy (CRT) compared to a discordant LV lead. However, the effect of the right ventricular (RV) lead site on CRT response is unclear. We investigated the relationship of the RV and LV lead positions on CRT response.
METHODS: In 131 CRT patients, the LV lead was positioned preferentially in a lateral or posterolateral vein and the RV lead to either the RV septum (RVS, n = 55) or RV apex (RVA, n = 76). The latest site of contraction was determined with two-dimensional speckle tracking radial strain imaging and patients had a concordant LV lead position if pacing the latest segment, and discordant if not. Response was defined as ≥15% reduction in LV end systolic volume (LVESV) at 6-month follow-up.
RESULTS: There were no significant differences in mean reduction of LVESV at follow-up (RVS vs RVA: -23.3 ± 16% vs 22.1 ± 18%, P = 0.70) or rate of responders (58.2% vs 57.9%, P = 0.97) between the two groups. In patients with a concordant LV lead (n = 71), the response rate was significantly higher than those with a discordant lead (76.1% vs 36.7%, P < 0.001). There were no differences in outcomes in patients with a concordant or discordant LV lead according to the RV lead location.
CONCLUSION: The extent of LV reverse remodeling following CRT is not related to the RV lead position, but is significantly higher in patients with a concordant LV lead. ©2010, The Authors. Journal compilation ©2010 Wiley Periodicals, Inc.

Entities:  

Mesh:

Year:  2011        PMID: 21208234     DOI: 10.1111/j.1540-8159.2010.02995.x

Source DB:  PubMed          Journal:  Pacing Clin Electrophysiol        ISSN: 0147-8389            Impact factor:   1.976


  8 in total

1.  ICD lead type and RV lead position in CRT-D recipients.

Authors:  Alexander P Benz; Mate Vamos; Julia W Erath; Peter Bogyi; Gabor Z Duray; Stefan H Hohnloser
Journal:  Clin Res Cardiol       Date:  2018-05-24       Impact factor: 5.460

2.  Right ventricular lead location, right-left ventricular lead interaction, and long-term outcomes in cardiac resynchronization therapy patients.

Authors:  Usama A Daimee; Helmut U Klein; Michael C Giudici; Wojciech Zareba; Scott McNitt; Bronislava Polonsky; Arthur J Moss; Valentina Kutyifa
Journal:  J Interv Card Electrophysiol       Date:  2018-03-23       Impact factor: 1.900

3.  Localization of pacing and defibrillator leads using standard x-ray views is frequently inaccurate and is not reproducible.

Authors:  Larry R Jackson; Jonathan P Piccini; James P Daubert; Lynne M Hurwitz Koweek; Brett D Atwater
Journal:  J Interv Card Electrophysiol       Date:  2015-02-27       Impact factor: 1.900

4.  Biophysical modeling to simulate the response to multisite left ventricular stimulation using a quadripolar pacing lead.

Authors:  Steven A Niederer; A K Shetty; G Plank; J Bostock; R Razavi; N P Smith; C A Rinaldi
Journal:  Pacing Clin Electrophysiol       Date:  2011-10-31       Impact factor: 1.976

5.  Adverse response to cardiac resynchronisation therapy in patients with septal scar on cardiac MRI preventing a septal right ventricular lead position.

Authors:  Simon G Duckett; Matthew Ginks; Anoop Shetty; Senthil Kirubakaran; Julian Bostock; Stam Kapetanakis; Jaswinder Gill; Gerry Carr-White; Reza Razavi; C Aldo Rinaldi
Journal:  J Interv Card Electrophysiol       Date:  2011-11-30       Impact factor: 1.900

6.  Right ventricular lead location and outcomes among patients with cardiac resynchronization therapy: A meta-analysis.

Authors:  Fatima Ali-Ahmed; Frederik Dalgaard; Nancy M Allen Lapointe; Andrzej S Kosinski; Vanessa Blumer; Daniel P Morin; Gillian D Sanders; Sana M Al-Khatib
Journal:  Prog Cardiovasc Dis       Date:  2021-04-20       Impact factor: 8.194

7.  Right ventricular pacing and sensing function in high posterior septal and apical lead placement in cardiac resynchronization therapy.

Authors:  Hm Kristiansen; T Hovstad; G Vollan; S Faerestrand
Journal:  Indian Pacing Electrophysiol J       Date:  2012-01-31

Review 8.  Optimal site selection and image fusion guidance technology to facilitate cardiac resynchronization therapy.

Authors:  Benjamin J Sieniewicz; Justin Gould; Bradley Porter; Baldeep S Sidhu; Jonathan M Behar; Simon Claridge; Steve Niederer; Christopher A Rinaldi
Journal:  Expert Rev Med Devices       Date:  2018-07-30       Impact factor: 3.166

  8 in total