Literature DB >> 25690710

Reproducibility of retrobulbar blood flow velocity measurements in normal subjects using two different CDI devices.

Carolina Dellafiore1, Massimiliano Lava, Alice Chandra Verticchio Vercellin, Carlo Alberto Cutolo, Carmine Tinelli, Annalisa De Silvestri, Fabrizio Calliada, Giovanni Milano.   

Abstract

PURPOSE: This study was done to evaluate the intra- and inter-operator reproducibility of colour Doppler imaging (CDI) in assessing blood flow velocity in the ophthalmic (OA), central retinal (CRA) and short posterior ciliary arteries (SPCA) in healthy subjects.
MATERIALS AND METHODS: The right eye of two groups of eight healthy volunteers was examined. Two radiologists and two ophthalmologists, divided into pairs, measured peak systolic velocity (PSV), end-diastolic velocity (EDV) and resistivity index (RI) of each vessel using a different CDI device for each group. The concordance between two measurements was evaluated with Lin's concordance correlation coefficient (CCC).
RESULTS: Globally, very good degrees of intra-operator concordance were obtained for the PSV (0.859 cm/s), EDV (0.834 cm/s) and RI (0.859) of the OA. There was moderate concordance for PSV (0.574 cm/s) and EDV (0.594 cm/s) and good concordance for RI (0.694) for the CRA. Good degrees of concordance were obtained for the SPCA measurements. However, inter-operator concordance was found globally poor.
CONCLUSIONS: These data show that CDI measurements in retrobulbar vessels are operator dependent. To increase the intra-operator and inter-operator concordance, rules should be adopted for timing of the examination and positioning of the probe to minimise the pressure applied on the eye.

Entities:  

Mesh:

Year:  2015        PMID: 25690710     DOI: 10.1007/s11547-015-0522-4

Source DB:  PubMed          Journal:  Radiol Med        ISSN: 0033-8362            Impact factor:   3.469


  16 in total

1.  Agreement among three examiners of colour Doppler imaging retrobulbar blood flow velocity measurements.

Authors:  Panayiota Founti; Alon Harris; Domniki Papadopoulou; Petros Emmanouilidis; Brent Siesky; Vassilis Kilintzis; Eleftherios Anastasopoulos; Angeliki Salonikiou; Theofanis Pappas; Fotis Topouzis
Journal:  Acta Ophthalmol       Date:  2011-09-20       Impact factor: 3.761

2.  Observer experience improves reproducibility of color Doppler sonography of orbital blood vessels.

Authors:  János Németh; Rita Kovács; Zoltán Harkányi; Krisztina Knézy; Katalin Sényi; István Marsovszky
Journal:  J Clin Ultrasound       Date:  2002 Jul-Aug       Impact factor: 0.910

Review 3.  Color Doppler imaging of orbital vessels: personal experience and literature review.

Authors:  François Tranquart; Olivier Bergès; Patricia Koskas; Sophie Arsene; Christian Rossazza; Pierre-Jean Pisella; Léandre Pourcelot
Journal:  J Clin Ultrasound       Date:  2003-06       Impact factor: 0.910

4.  Color Doppler imaging of ophthalmic artery blood flow velocity: a study of repeatability and agreement.

Authors:  L Quaranta; A Harris; F Donato; M Cassamali; F Semeraro; G Nascimbeni; E Gandolfo; C A Quaranta
Journal:  Ophthalmology       Date:  1997-04       Impact factor: 12.079

5.  Systemic vascular dysregulation and retrobulbar hemodynamics in normal-tension glaucoma.

Authors:  Fernando Galassi; Barbara Giambene; Roberta Varriale
Journal:  Invest Ophthalmol Vis Sci       Date:  2011-06-23       Impact factor: 4.799

6.  Effect of body position on intraocular pressure and aqueous flow.

Authors:  K H Carlson; J W McLaren; J E Topper; R F Brubaker
Journal:  Invest Ophthalmol Vis Sci       Date:  1987-08       Impact factor: 4.799

7.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

8.  Repeatability of retrobulbar blood flow velocity measured using color Doppler imaging in the Indianapolis Glaucoma Progression Study.

Authors:  Rita Ehrlich; Alon Harris; Brent A Siesky; Adam M Moss; Meera Ramanathan; Melanie A Pickett; Darrell Wudunn; Michael Hawkes; Yochai Z Shoshani
Journal:  J Glaucoma       Date:  2011-12       Impact factor: 2.503

Review 9.  Timing of blood pressure measurement related to caffeine consumption.

Authors:  Jane R Mort; Heather R Kruse
Journal:  Ann Pharmacother       Date:  2007-12-19       Impact factor: 3.154

10.  Ocular perfusion pressure and retrobulbar haemodynamics in pseudoexfoliative glaucoma.

Authors:  Fernando Galassi; Barbara Giambene; Ugo Menchini
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2007-10-31       Impact factor: 3.117

View more
  4 in total

1.  Inter-device reproducibility of retrobulbar blood flow velocity measurements in healthy subjects using color Doppler imaging.

Authors:  Verticchio Vercellin Alice C; Carlo A Cutolo; Carolina Dellafiore; Massimiliano Lava; Carmine Tinelli; Annalisa De Silvestri; Fabrizio Calliada; Giovanni Milano
Journal:  J Ultrasound       Date:  2016-02-01

2.  Usefulness of colour Doppler flow imaging in the management of lacrimal gland lesions.

Authors:  A Lecler; M Boucenna; F Lafitte; P Koskas; E Nau; P V Jacomet; O Galatoire; S Morax; M Putterman; F Mann; F Héran; J C Sadik; H Picard; O Bergès
Journal:  Eur Radiol       Date:  2016-06-07       Impact factor: 5.315

3.  Higher glucose level and systemic oxidative stress decrease the mean velocity index of the retinal artery during flickering light stimulation in type 1 diabetes.

Authors:  Vladimir Debelić; Brigita Drnovšek Olup; Bogomir Žižek; Milan Skitek; Aleš Jerin
Journal:  Croat Med J       Date:  2016-10-31       Impact factor: 1.351

Review 4.  Neurodegenerative Disorders of the Eye and of the Brain: A Perspective on Their Fluid-Dynamical Connections and the Potential of Mechanism-Driven Modeling.

Authors:  Giovanna Guidoboni; Riccardo Sacco; Marcela Szopos; Lorenzo Sala; Alice Chandra Verticchio Vercellin; Brent Siesky; Alon Harris
Journal:  Front Neurosci       Date:  2020-11-12       Impact factor: 4.677