Inter-device reproducibility of retrobulbar blood flow velocity measurements in healthy subjects using color Doppler imaging.

Verticchio Vercellin Alice C1, Carlo A Cutolo2, Carolina Dellafiore3, Massimiliano Lava3, Carmine Tinelli4, Annalisa De Silvestri4, Fabrizio Calliada3, Giovanni Milano1.   

Abstract

PURPOSE: To evaluate the inter-device reproducibility of retrobulbar blood flow measurements obtained by two commercially available CDI (color Doppler imaging) devices.
METHODS: The right eyes of 10 healthy volunteers were investigated. Four examiners, namely two ophthalmologists and two radiologists, performed CDI examination of the ophthalmic artery, central retinal artery and temporal short posterior ciliary arteries using both CDI devices: ESAOTE MYLAB™ and SIEMENS ANTARES STELLAR PLUS™. The peak systolic velocity (PSV), the end-diastolic velocity (EDV) and the resulting resistivity index (RI) were averaged for 3 cardiac cycles. To evaluate the reproducibility between both device measurements, the Lin's concordance correlation coefficient (CCC) was used. CCC can be expressed as the product of Pearson's r (the measure of precision) and C_b (the measure of accuracy).
RESULTS: Results show that the inter-device reproducibility for CDI measurements is not acceptable since a poor degree of overall concordance (0.15<CCC<0.37) was obtained: accuracy was high (C_b > 0.71) but overall precision low (0.18< Pearson's r <0.47). Ophthalmologists and radiologists obtained similar results.
CONCLUSIONS: To evaluate the causal role of blood flow abnormalities in glaucoma, CDI analysis using different devices seems unreliable. CDI inter-device reproducibility seems unrelated to medical speciality of the examiners. However, to improve present results, the use of similar probes and standardized CDI instrument settings as well as a CDI images analysis by a single grader, might possibly improve the inter-device reproducibility when testing the retrobulbar blood flow velocity.

Entities:  

Keywords:  Blood flow velocity; Doppler ultrasound imaging; Glaucoma; Healthy volunteers; Reproducibility of results

Mesh:

Year:  2016        PMID: 27298650      PMCID: PMC4879017          DOI: 10.1007/s40477-016-0194-5

Source DB:  PubMed          Journal:  J Ultrasound        ISSN: 1876-7931


  12 in total

1.  Agreement among three examiners of colour Doppler imaging retrobulbar blood flow velocity measurements.

Authors:  Panayiota Founti; Alon Harris; Domniki Papadopoulou; Petros Emmanouilidis; Brent Siesky; Vassilis Kilintzis; Eleftherios Anastasopoulos; Angeliki Salonikiou; Theofanis Pappas; Fotis Topouzis
Journal:  Acta Ophthalmol       Date:  2011-09-20       Impact factor: 3.761

2.  Reproducibility of retrobulbar blood flow velocity measurements in normal subjects using two different CDI devices.

Authors:  Carolina Dellafiore; Massimiliano Lava; Alice Chandra Verticchio Vercellin; Carlo Alberto Cutolo; Carmine Tinelli; Annalisa De Silvestri; Fabrizio Calliada; Giovanni Milano
Journal:  Radiol Med       Date:  2015-02-19       Impact factor: 3.469

3.  A concordance correlation coefficient to evaluate reproducibility.

Authors:  L I Lin
Journal:  Biometrics       Date:  1989-03       Impact factor: 2.571

4.  Retrobulbar blood flow velocities in open angle glaucoma and their association with mean arterial blood pressure.

Authors:  Gerhard Garhöfer; Gabriele Fuchsjäger-Mayrl; Clemens Vass; Berthold Pemp; Anton Hommer; Leopold Schmetterer
Journal:  Invest Ophthalmol Vis Sci       Date:  2010-08-04       Impact factor: 4.799

5.  Norepinephrine and epinephrine release and adrenergic mediation of smoking-associated hemodynamic and metabolic events.

Authors:  P E Cryer; M W Haymond; J V Santiago; S D Shah
Journal:  N Engl J Med       Date:  1976-09-09       Impact factor: 91.245

6.  Test/Retest reproducibility of color Doppler imaging assessment of blood flow velocity in orbital vessels.

Authors:  A Harris; T H Williamson; B Martin; J A Shoemaker; R C Sergott; G L Spaeth; J L Katz
Journal:  J Glaucoma       Date:  1995-08       Impact factor: 2.503

7.  Color Doppler imaging in evaluation of optic nerve blood supply in normal and glaucomatous subjects.

Authors:  F Galassi; G Nuzzaci; A Sodi; P Casi; A Vielmo
Journal:  Int Ophthalmol       Date:  1992-09       Impact factor: 2.031

8.  Reduction of intraocular pressure and glaucoma progression: results from the Early Manifest Glaucoma Trial.

Authors:  Anders Heijl; M Cristina Leske; Bo Bengtsson; Leslie Hyman; Boel Bengtsson; Mohamed Hussein
Journal:  Arch Ophthalmol       Date:  2002-10

9.  Reproducibility of blood flow velocity measurements using colour decoded Doppler imaging.

Authors:  E T Matthiessen; O Zeitz; G Richard; M Klemm
Journal:  Eye (Lond)       Date:  2004-04       Impact factor: 3.775

10.  Current primary open-angle glaucoma treatments and future directions.

Authors:  Gabriel Beidoe; Shaker A Mousa
Journal:  Clin Ophthalmol       Date:  2012-10-23
View more
  2 in total

Review 1.  Neurodegenerative Disorders of the Eye and of the Brain: A Perspective on Their Fluid-Dynamical Connections and the Potential of Mechanism-Driven Modeling.

Authors:  Giovanna Guidoboni; Riccardo Sacco; Marcela Szopos; Lorenzo Sala; Alice Chandra Verticchio Vercellin; Brent Siesky; Alon Harris
Journal:  Front Neurosci       Date:  2020-11-12       Impact factor: 4.677

2.  Anti-atherosclerotic effects between a combined treatment with simvastatin plus hirudin and single simvastatin therapy in patients with early type 2 diabetes mellitus.

Authors:  De-Qiang Li; Fei-Fei Lv; Zhong-Chun Li; Zhi-Yuan Dai; Hong-Xia Wang; Yang Han
Journal:  Ann Transl Med       Date:  2019-07
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.