Literature DB >> 25683995

How to minimize rotational conflict between femoral & tibial component in total knee arthroplasty: the use of femoro-tibial axial synchronizer (Linker).

Jai-Gon Seo1, Young-Wan Moon1, Sang-Min Kim1, Sang-Hoon Park2.   

Abstract

PURPOSE: The purpose of this study was to investigate the correlation between rotational axes of femur and tibia with the use of Linker.
MATERIALS AND METHODS: This study was carried out from August 2009 to February 2010 on 54 patients (106 knees), who were diagnosed with simultaneous bilateral total knee arthroplasty. With the use of postoperative computed tomography scans, it was investigated how much the rotational angle of femoral and tibial components matched.
RESULTS: The tibial component was internally rotated for the femoral component at an angle of 0.8°. The femoral component was externally rotated for the surgical transepicondylar axis (TEA) at an angle of 1.6 (range: from 4.8° of internal rotation to 7.9° of external rotation, SD=2.2°), and the tibial component was externally rotated for the surgical TEA at an average angle of 0.9 (range: from 5.1° of internal rotation to 8.3° of external rotation, SD=3.1°).
CONCLUSION: The femoro-tibial synchronizer helped to improve the orientation and positioning of both femoral component and tibial component, and also increase the correlation of the rotational axes of the two components.

Entities:  

Keywords:  Rotational axis; extramedullary reference; femoro-tibial synchronizer; total knee arthroplasty

Mesh:

Year:  2015        PMID: 25683995      PMCID: PMC4329358          DOI: 10.3349/ymj.2015.56.2.454

Source DB:  PubMed          Journal:  Yonsei Med J        ISSN: 0513-5796            Impact factor:   2.759


  20 in total

1.  The Ranawat Award. Femoral component rotation during total knee arthroplasty.

Authors:  C W Olcott; R D Scott
Journal:  Clin Orthop Relat Res       Date:  1999-10       Impact factor: 4.176

2.  Reproducibility of measurement of femoral component rotation after total knee arthroplasty using computer tomography.

Authors:  Thomas Suter; Marco Zanetti; Marius Schmid; José Romero
Journal:  J Arthroplasty       Date:  2006-08       Impact factor: 4.757

3.  Synchronisation of tibial rotational alignment with femoral component in total knee arthroplasty.

Authors:  Dong-Hoon Lee; Jai-Gon Seo; Young-Wan Moon
Journal:  Int Orthop       Date:  2007-01-12       Impact factor: 3.075

4.  Determining the rotational alignment of the tibial component at total knee replacement: a comparison of two techniques.

Authors:  M Ikeuchi; N Yamanaka; Y Okanoue; E Ueta; T Tani
Journal:  J Bone Joint Surg Br       Date:  2007-01

5.  A prospective, randomized study of computer-assisted and conventional total knee arthroplasty. Three-dimensional evaluation of implant alignment and rotation.

Authors:  Georg Matziolis; Doerte Krocker; Ulrike Weiss; Stephan Tohtz; Carsten Perka
Journal:  J Bone Joint Surg Am       Date:  2007-02       Impact factor: 5.284

6.  Rotational position of femoral and tibial components in TKA using the femoral transepicondylar axis.

Authors:  Paolo Aglietti; Lorenzo Sensi; Pierluigi Cuomo; Antonio Ciardullo
Journal:  Clin Orthop Relat Res       Date:  2008-09-30       Impact factor: 4.176

7.  An additional reference axis for determining rotational alignment of the femoral component in total knee arthroplasty.

Authors:  Ye-Yeon Won; Wen-Quan Cui; Myong-Hyun Baek; Tae-Bong Yun; Seung-Ho Han
Journal:  J Arthroplasty       Date:  2007-03-23       Impact factor: 4.757

8.  Screw home motion after total knee replacement.

Authors:  Y Ishii; K Terajima; Y Koga; J E Bechtold
Journal:  Clin Orthop Relat Res       Date:  1999-01       Impact factor: 4.176

9.  Mechanical axis-derived femoral component rotation in extramedullary total knee arthroplasty: a comparison between femoral transverse axis and transepicondylar axis.

Authors:  Jai-Gon Seo; Young-Wan Moon; Ji-Soon Lim; Se-Jun Park; Sang-Min Kim
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2011-07-13       Impact factor: 4.342

10.  Variability of extraarticular tibial rotation references for total knee arthroplasty.

Authors:  Masao Akagi; Shigeshi Mori; Shunji Nishimura; Akio Nishimura; Taiyo Asano; Chiaki Hamanishi
Journal:  Clin Orthop Relat Res       Date:  2005-07       Impact factor: 4.176

View more
  4 in total

1.  Optimal rotational positioning of tibial component in total knee arthroplasty: determined by linker surgical technique using a high definition CT.

Authors:  Woon-Hwa Jung; Jai-Gon Seo; Dong Hyun Kim; Suryamanikanta Balabadra; Utkrisht Mandot; Dinesh Kumar
Journal:  Arch Orthop Trauma Surg       Date:  2020-01-21       Impact factor: 3.067

2.  Finite element analysis of the tibial component alignment in a transverse plane in total knee arthroplasty.

Authors:  Roman Popescu; Stefan Cristea; Valentin Oleksik; Adrian Marius Pascu; Emil George Haritinian
Journal:  J Appl Biomed       Date:  2021-09-20       Impact factor: 1.797

3.  Revision Arthroplasty Using a MUTARS® Prosthesis in Comminuted Periprosthetic Fracture of the Distal Femur.

Authors:  Hyung Suk Choi; Jae Hwi Nho; Chung Hyun Kim; Sai Won Kwon; Jong Seok Park; You Sung Suh
Journal:  Yonsei Med J       Date:  2016-11       Impact factor: 2.759

4.  The impact of generalized joint laxity on clinical outcomes of total knee arthroplasty.

Authors:  Sae Kwang Kwon; Hyuck Min Kwon; Youngho Kong; Kwan Kyu Park
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2017-03-06       Impact factor: 4.342

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.