| Literature DB >> 25677938 |
Pregash Ellapparadja1, Vivek Mahajan1, Angela H Deakin1, Kamal Deep1.
Abstract
This study assesses how accurately we can restore hip offset and leg length in navigated total hip arthroplasty (THA). 152 consecutive patients with navigated THA formed the study group. The contra-lateral hip formed control for measuring hip offset and leg length. All radiological measurements were made using Orthoview digital software. In the normal hip offset group, the mean is 75.73 (SD- 8.61). In the reconstructed hip offset group, the mean is 75.35 (SD - 7.48). 95.39% had hip offset within 6 mm of opposite side while 96.04% had leg length restored within 6 mm of contra-lateral side. Equivalence test revealed that the two groups of hip offsets were essentially the same. We conclude that computer navigation can successfully reproduce hip offset and leg length accurately.Entities:
Keywords: Orthopilot; THR; hip offset; leg length; navigation
Mesh:
Year: 2015 PMID: 25677938 DOI: 10.1016/j.arth.2015.01.027
Source DB: PubMed Journal: J Arthroplasty ISSN: 0883-5403 Impact factor: 4.757