| Literature DB >> 25675888 |
M A Elbashier Abdelbasit1, Mazeni Alwi1, Geetha Kandavello1, Marhisham Che Mood1, Hasri Samion1, Ziyad M Hijazi2.
Abstract
OBJECTIVES: To evaluate the feasibility, safety, and efficacy of the Occlutech® PDA occluder for closure of patent ductus arteriosus (PDA).Entities:
Keywords: Occlutech® PDA occluder; patent ductus arteriosus
Mesh:
Year: 2015 PMID: 25675888 PMCID: PMC6680177 DOI: 10.1002/ccd.25878
Source DB: PubMed Journal: Catheter Cardiovasc Interv ISSN: 1522-1946 Impact factor: 2.692
Figure 1(a and b) Schematic representation and actual photograph of the Occlutech® PDA occluder design. (D1) The aortic disc diameter. (D2) The distal (aortic) end diameter. (D3) The proximal (pulmonary) end diameter. (L) The shank length, note the long tether between the aortic disc and the shank.
Occlutech Ductal Occluder Sizes and Lengths
| Ductal occluder short shank | Size | 3.5/5 | 4/6 | 5/7 | 6/8 | 8/10 | 10/12 | 12/15 | 14/18 |
| Length | 4.25 | 5 | 6.05 | 6.3 | 7 | 12 | 14 | 16 | |
| Ductal occluder longer shank | Size | 3.5/5 | 4/6L | 5/7L | 6/8L | 8/10L | |||
| Length | 7 | 7.5 | 8.5 | 9 | 10.5 |
Figure 2(a) Descending aortic angiography in lateral projection. Large type A PDA measuring 4.8 mm at its narrowest diameter in an asymptomatic 20‐month old, 7.8 kg Down syndrome female. (b) The 8/10 short shank version 3 Occlutech® PDA Occluder was implanted. Small residual shunt shown on angiography, and significant residual shunt 24‐hr post implant was shown by echocardiography together with audible murmur but achieved complete closure in 30 days. (c) Descending aortic angiography in lateral projection. Small type C PDA measuring 2.1 mm in a 4‐year‐old male. Note mild coarctation. (d) The 4/6 short shank Occlutech® PDA Occluder was implanted, no residual shunt with no increase in descending aorta velocity pre and post‐procedure. (e) Type E PDA measured 2.75 mm at its narrowest diameter in a 3‐years old, 11.2 kg, female. (f) 4/6 short shank Occlutech® PDA occluder was implanted with no residual shunt.
Demographic and Hemodynamic Data of All Patients Underwent Attempted PDA Closure
| No. of patients | 33 patients |
| Female 20(60.6%)/male 13 (39.4%) | |
| Age (years) | Median = 2 (0.5–38) years |
| Weight (kg) | Median = 9.3 (6–69.2) kg |
| PDA size (narrowest diameter) (mm) | Mean = 2.97 ± 1.03 |
|
| Mean = 2±1.1 |
| PVR | Mean = 2.06 ± 0.89 |
| MEAN PAP (mm Hg) | Median = 21 (12–52) |
| Mean PA//AO ratio | Median = 0.33 (0.17–0.81) |
| Large | 9 (27.3%) |
| Moderate | 18 (54.5%) |
| Small | 6 (18.1%) |
| PDA morphology | |
| Type A | 30 (90.9%) |
| Type E | 2 (6.1%) |
| Type C | 1 (3%) |
| Fluoroscopy time (min) | 8.76 ± 3.81 |
| Procedure time (min) | Median = 60 (20–165) |