Literature DB >> 23427043

Comparison of the results of transcatheter closure of patent ductus arteriosus with newer Amplatzer devices.

Stephen Liddy1, Paul Oslizlok, Kevin P Walsh.   

Abstract

OBJECTIVES: To compare the success and complication rates for the different devices and patent ductus arteriosus (PDA) types.
BACKGROUND: There are numerous devices available for transcatheter closure of PDA but data comparing the success and complication rates of the newer devices are lacking.
METHODS: A retrospective review was undertaken of all cases of transcatheter PDA closure at a tertiary referral center.
RESULTS: Transcatheter PDA closure was achieved in 176 of 177 cases. Mean age, weight, and minimum ductal diameter were 3.3 years, 15.0 kg, and 2.5 mm, respectively. Devices used were Amplatzer Duct Occluder I (ADO I, 50.3%), Amplatzer Duct Occluder II (ADO II, 28.2%), Amplatzer Duct Occluder II Additional Sizes (ADO II AS, 5.1%), flipper coils (14.1%), and "other" devices (2.3%). Regarding Amplatzer devices, arterial delivery had a significantly shorter fluoroscopy time as compared to venous delivery (3.7 vs. 5.0 min, P = 0.0068), with no significant difference in outcomes. Immediate residual shunt rates were 7.9% (7/89) for ADO I, 4.1% (2/49) for ADO II, 11.1% (1/9) for ADO II AS, and 0% (0/25) for flipper coils. Aortic protrusion rates were 9% (8/89) for ADO I, 14.3% (7/49) for ADO II, and O% for both ADO II AS (0/9) and flipper coils (0/25).
CONCLUSIONS: All devices achieved excellent occlusion rates with low complication rates, regardless of PDA type. Device delivery via arterial access was associated with a shorter fluoroscopy time as compared to venous delivery, with no significant difference in outcomes. ADO II AS may be preferable to ADO II in the younger infant.
© 2013 Wiley Periodicals, Inc.

Entities:  

Keywords:  amplatzer duct occluder; congenital heart disease; flipper coil; paediatric interventions

Mesh:

Year:  2013        PMID: 23427043     DOI: 10.1002/ccd.24768

Source DB:  PubMed          Journal:  Catheter Cardiovasc Interv        ISSN: 1522-1946            Impact factor:   2.692


  10 in total

1.  Transcatheter Closure of Patent Ductus Arteriosus in Children with the Occlutech Duct Occluder.

Authors:  Meki Bilici; Fikri Demir; Alper Akın; Mehmet Türe; Hasan Balık; Mahir Kuyumcu
Journal:  Pediatr Cardiol       Date:  2017-08-21       Impact factor: 1.655

2.  Multicenter Off-Label Use of Nit-Occlud Coil in Retrograde Closure of Small Patent Ductus Arteriosus.

Authors:  Keyhan Sayadpour Zanjani; Rodina Sobhy; Rania El-Kaffas; Amal El-Sisi
Journal:  Pediatr Cardiol       Date:  2017-02-21       Impact factor: 1.655

3.  Transcatheter Closure of Arterial Duct in Infants < 6 kg: Amplatzer Duct Occluder Type I vs Amplatzer Duct Occluder II Additional Sizes.

Authors:  Giuseppe Santoro; Mario Giordano; Gianpiero Gaio; Maria Teresa Palladino; Giovanbattista Capozzi; Carola Iacono; Maria Giovanna Russo
Journal:  Pediatr Cardiol       Date:  2018-02-17       Impact factor: 1.655

4.  Initial experience with the 3.3 Fr Mongoose® pigtail catheter for aortic angiography during patent ductus arteriosus closure in small patients.

Authors:  Zachary Hena; Nicole J Sutton; Gregory J Gates; Benjamin H Taragin; Robert H Pass
Journal:  Ann Pediatr Cardiol       Date:  2017 Sep-Dec

5.  Cost-effectiveness analysis of devices for closure of patent ductus arteriosus.

Authors:  K M Krishnamoorthy; Arun Gopalakrishnan
Journal:  Ann Pediatr Cardiol       Date:  2018 Jan-Apr

6.  The new Occlutech® PDA occluder: Initial human experience.

Authors:  M A Elbashier Abdelbasit; Mazeni Alwi; Geetha Kandavello; Marhisham Che Mood; Hasri Samion; Ziyad M Hijazi
Journal:  Catheter Cardiovasc Interv       Date:  2015-04-02       Impact factor: 2.692

7.  Percutaneous closure of patent ductus arteriosus with the Nit-Occlud® patent ductus arteriosus device in 268 consecutive cases.

Authors:  Andrii V Maksymenko; Yulia L Kuzmenko; Arkadii A Dovhaliuk; Oleksandra O Motrechko; Florian E Herrmann; Nikolaus A Haas; Anja Lehner
Journal:  Ann Pediatr Cardiol       Date:  2019 Sep-Dec

8.  Trans-Catheter Closure of Patent Ductus Arteriosus, which device should we use?

Authors:  Rym Gribaa; Sami Ouannes; Heithem Tangour; Sameh Ben Farhat; Houssem Thabet; Mehdi Slim; Aymen Elheraiche; Elyes Neffati
Journal:  Tunis Med       Date:  2021 Novembre

9.  Cost-effectiveness analysis of different devices used for the closure of small-to-medium-sized patent ductus arteriosus in pediatric patients.

Authors:  Sonia A El-Saiedi; Amal M El Sisi; Rodina Sobhy Mandour; Doaa M Abdel-Aziz; Wael A Attia
Journal:  Ann Pediatr Cardiol       Date:  2017 May-Aug

Review 10.  Recent advances in cardiac catheterization for congenital heart disease.

Authors:  Sok-Leng Kang; Lee Benson
Journal:  F1000Res       Date:  2018-03-26
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.