Literature DB >> 25648870

[Intraoperative virtual implant planning for volar plate osteosynthesis of distal radius fractures].

J Franke1, S Y Vetter2, K Reising3, S Herrmann3, N P Südkamp3, P A Grützner2, J von Recum2.   

Abstract

BACKGROUND: Digital planning of implants is in most cases conducted prior to surgery. The virtual implant planning system (VIPS) is an application developed for mobile C-arms, which assists the virtual planning of screws close to the joint line during surgery for treatment of distal radius fractures with volar plate osteosynthesis. The aim of this prospective randomized study was to acquire initial clinical experiences and to compare the VIPS method with the conventional technique.
METHOD: The study included 10 patients for primary testing and 30 patients with distal radius fractures of types A3, C1 and C2, divided in 2 groups. In the VIPS group, after placement of the plate and fracture reduction, a virtual 3D model of the plate was matched with the image of the plate from the fluoroscopic acquisition. Next, the length and position of the screws close to the joint line were planned on the virtual plate. The control group was treated with the same implant in the conventional way. Data were collected regarding screw replacement, fluoroscopy and operating room (OR) times.
RESULTS: The VIPS group included six A3, one C1 and eight C2 fractures, while the control group consisted of six A3 and nine C2 fractures. Three screws were replaced in the VIPS group and two in the control group (p = 0.24). The mean intraoperative fluoroscopy time of the VIPS group amounted to 2.58 ± 1.38 min, whereas it was 2.12 ± 0.73 min in the control group (p = 0.26). The mean OR time in the VIPS group was 53.3 ± 34.5 minutes and 42.3 ± 8.8 min (p = 0.23) in the control group.
CONCLUSION: The VIPS enables a precise positioning of screws close to joint line in the treatment of distal radius fractures; however, for routine use, further development of the system is necessary.

Entities:  

Keywords:  Intraoperative 3D imaging; Intraoperative fluoroscopy time; Matching; Operating time; Screw misplacement

Mesh:

Year:  2016        PMID: 25648870     DOI: 10.1007/s00113-014-2715-z

Source DB:  PubMed          Journal:  Unfallchirurg        ISSN: 0177-5537            Impact factor:   1.000


  13 in total

1.  Osteosynthesis of distal radial fractures with a volar locking screw plate system.

Authors:  H Drobetz; E Kutscha-Lissberg
Journal:  Int Orthop       Date:  2002-08-21       Impact factor: 3.075

2.  Screw placement for acetabular fractures: which navigation modality (2-dimensional vs. 3-dimensional) should be used? An experimental study.

Authors:  Florian Gras; Ivan Marintschev; Kajetan Klos; Thomas Mückley; Gunther O Hofmann; David M Kahler
Journal:  J Orthop Trauma       Date:  2012-08       Impact factor: 2.512

3.  Computer-assisted LISS plate osteosynthesis of proximal tibia fractures: feasibility study and first clinical results.

Authors:  Paul Alfred Grützner; Frank Langlotz; Guoyan Zheng; Jan von Recum; Christina Keil; Lutz P Nolte; Andreas Wentzensen; Klaus Wendl
Journal:  Comput Aided Surg       Date:  2005-05

4.  Fluoroscopy-based navigation system in spine surgery.

Authors:  P Merloz; J Troccaz; H Vouaillat; C Vasile; J Tonetti; A Eid; S Plaweski
Journal:  Proc Inst Mech Eng H       Date:  2007-10       Impact factor: 1.617

5.  [Evaluation of a 2D fluoroscopy-based navigation system for insertion of femoral neck screws. An experimental study].

Authors:  M C Müller; P Belei; M de la Fuente; M Strake; K Kabir; O Weber; C Burger; K Radermacher; D C Wirtz
Journal:  Unfallchirurg       Date:  2012-09       Impact factor: 1.000

6.  Percutaneous screw fixation of acetabular fractures with 2D fluoroscopy-based computerized navigation.

Authors:  Gao Hong; Luo Cong-Feng; Hu Cheng-Fang; Zhang Chang-Qing; Zeng Bing-Fang
Journal:  Arch Orthop Trauma Surg       Date:  2010-04-03       Impact factor: 3.067

7.  [Early results after palmar multi-axial plate osteosynthesis for treatment of distal radius fractures].

Authors:  T Geyer; K Hefele; J Gülke; F Gebhard; M Mentzel
Journal:  Unfallchirurg       Date:  2011-10       Impact factor: 1.000

8.  Fracture of the distal radius: risk factors for complications after locked volar plate fixation.

Authors:  Maximillian Soong; Roderick van Leerdam; Thierry G Guitton; Christopher Got; Julia Katarincic; David Ring
Journal:  J Hand Surg Am       Date:  2011-01       Impact factor: 2.230

9.  Delayed ruptures of the extensor tendon secondary to the use of volar locking compression plates for distal radial fractures.

Authors:  M Al-Rashid; K Theivendran; M A C Craigen
Journal:  J Bone Joint Surg Br       Date:  2006-12

10.  2D-fluoroscopic based navigation for Gamma 3 nail insertion versus conventional procedure- a feasibility study.

Authors:  A Wilharm; I Marintschev; G O Hofmann; F Gras
Journal:  BMC Musculoskelet Disord       Date:  2013-02-28       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.