Amiran S Revishvili1, Erik Wissner2, Dmitry S Lebedev3, Christine Lemes4, Sebastian Deiss4, Andreaas Metzner4, Vitaly V Kalinin1, Oleg V Sopov1, Eugeny Z Labartkava1, Alexander V Kalinin5, Michail Chmelevsky3, Stephan V Zubarev3, Maria K Chaykovskaya6, Mikhail G Tsiklauri6, Karl-Heinz Kuck4.
Abstract
AIMS: Use of a non-invasive electrocardiographic mapping system may aid in rapid diagnosis of atrial or ventricular arrhythmias or the detection of ventricular dyssynchrony. The aim of the present study was to validate the mapping accuracy of a novel non-invasive epi- and endocardial electrophysiology system (NEEES). METHODS ANDEntities:
Keywords: Arrhythmias; Endocardial mapping; Epicardial mapping; Non-invasive mapping system
Mesh:
Year: 2015 PMID: 25643987 PMCID: PMC4535554 DOI: 10.1093/europace/euu339
Source DB: PubMed Journal: Europace ISSN: 1099-5129 Impact factor: 5.214
Accuracy according to paced cardiac chamber and number of pacing attempts analysed in implanted device study cohort
| Heart chamber | RA, endocardium ( | RV, endocardium ( | LV, epicardium ( | RA + RV + LV ( |
|---|---|---|---|---|
| Mean ± SD (mm) | 10.8 ± 5.4 | 7.7 ± 5.8 | 7.9 ± 5.7 | 8.6 ± 5.8 |
Mean distance between reference pacing site and reconstructed pacing site and standard deviation for N pacing events.
RA, right atrium; RV, right ventricle; LV, left ventricle.
Effect of breathing on the accuracy of the NEEES
| Heart chamber | RA ( | RV ( | LV ( | RA + RV + LV ( |
|---|---|---|---|---|
| Spontaneous breathing, mean ± SD (mm) | 8.0 ± 1.8 | 6.3 ± 1.8 | 6.4 ± 1.5 | 6.8 ± 1.9 ( |
| Inspiration breath hold, mean ± SD (mm) | 5.9 ± 1.3 | 4.7 ± 1.3 | 5.9 ± 1.6 | 5.5 ± 1.6 ( |
| Expiration breath hold, mean ± SD (mm) | 9.5 ± 1.9 | 8.1 ± 1.7 | 8.5 ± 1.1 | 8.6 ± 1.6 ( |
Pacing during spontaneous breathing and during breath hold on inspiration and expiration. Mean distance between reference pacing site and reconstructed pacing site and standard deviation for N pacing events.
RA, right atrium; RV, right ventricle; LV, left ventricle.
Results for the CARTO pacing study according to anatomical site paced
| Heart chamber | RA ( | LA ( | RV ( | LV ( | RA + LA + RV + LV ( |
|---|---|---|---|---|---|
| Mean distance ± SD (mm) | 7.4 ± 2.7 | 6.9 ± 2.3 | 6.5 ± 2.1 | 6.4 ± 2.2 | 6.8 ± 2.4 |
Mean distance between reference pacing site and reconstructed pacing site and standard deviation for N pacing events.
RA, right atrium; LA, left atrium; RV, right ventricle; LV, left ventricle; SD, standard deviation.