Finite element simulation of ultrasonic wave propagation in a dental implant for biomechanical stability assessment.

Romain Vayron1, Vu-Hieu Nguyen, Romain Bosc, Salah Naili, Guillaume Haïat.   

Abstract

Dental implant stability, which is an important parameter for the surgical outcome, can now be assessed using quantitative ultrasound. However, the acoustical propagation in dental implants remains poorly understood. The objective of this numerical study was to understand the propagation phenomena of ultrasonic waves in cylindrically shaped prototype dental implants and to investigate the sensitivity of the ultrasonic response to the surrounding bone quantity and quality. The 10-MHz ultrasonic response of the implant was calculated using an axisymetric 3D finite element model, which was validated by comparison with results obtained experimentally and using a 2D finite difference numerical model. The results show that the implant ultrasonic response changes significantly when a liquid layer is located at the implant interface compared to the case of an interface fully bounded with bone tissue. A dedicated model based on experimental measurements was developed in order to account for the evolution of the bone biomechanical properties at the implant interface. The effect of a gradient of material properties on the implant ultrasonic response is determined. Based on the reproducibility of the measurement, the results indicate that the device should be sensitive to the effects of a healing duration of less than one week. In all cases, the amplitude of the implant response is shown to decrease when the dental implant primary and secondary stability increase, which is consistent with the experimental results. This study paves the way for the development of a quantitative ultrasound method to evaluate dental implant stability.

Mesh:

Substances:

Year:  2015        PMID: 25619479     DOI: 10.1007/s10237-015-0651-7

Source DB:  PubMed          Journal:  Biomech Model Mechanobiol        ISSN: 1617-7940


  10 in total

Review 1.  Biomechanical behaviours of the bone-implant interface: a review.

Authors:  Xing Gao; Manon Fraulob; Guillaume Haïat
Journal:  J R Soc Interface       Date:  2019-07-31       Impact factor: 4.118

2.  Updates on ultrasound research in implant dentistry: a systematic review of potential clinical indications.

Authors:  Vaishnavi Bhaskar; Hsun-Liang Chan; Mark MacEachern; Oliver D Kripfgans
Journal:  Dentomaxillofac Radiol       Date:  2018-06-06       Impact factor: 2.419

3.  Ultrasonic Evaluation of the Bone-Implant Interface.

Authors:  Yoann Hériveaux; Vu-Hieu Nguyen; Guillaume Haïat
Journal:  Adv Exp Med Biol       Date:  2022       Impact factor: 2.622

4.  Evaluating the dynamic behaviour of bone anchored hearing aids using a finite element model and its applications to implant stability assessment.

Authors:  Mostafa Mohamed; Lindsey Westover
Journal:  Med Biol Eng Comput       Date:  2022-07-20       Impact factor: 3.079

Review 5.  Finite element analysis of dental implants with validation: to what extent can we expect the model to predict biological phenomena? A literature review and proposal for classification of a validation process.

Authors:  Yuanhan Chang; Abhijit Anil Tambe; Yoshinobu Maeda; Masahiro Wada; Tomoya Gonda
Journal:  Int J Implant Dent       Date:  2018-03-08

6.  Elastography of the bone-implant interface.

Authors:  Yoann Hériveaux; Vu-Hieu Nguyen; Didier Geiger; Guillaume Haïat
Journal:  Sci Rep       Date:  2019-10-02       Impact factor: 4.379

7.  Quantitative ultrasound assessment of the influence of roughness and healing time on osseointegration phenomena.

Authors:  M Fraulob; R Vayron; S Le Cann; B Lecuelle; Y Hériveaux; H Albini Lomami; C H Flouzat Lachaniette; G Haïat
Journal:  Sci Rep       Date:  2020-12-15       Impact factor: 4.379

8.  Comparison of Resonance Frequency Analysis and of Quantitative Ultrasound to Assess Dental Implant Osseointegration.

Authors:  Romain Vayron; Vu-Hieu Nguyen; Benoît Lecuelle; Hugues Albini Lomami; Jean-Paul Meningaud; Romain Bosc; Guillaume Haiat
Journal:  Sensors (Basel)       Date:  2018-05-02       Impact factor: 3.576

Review 9.  Clinical Assessment of Dental Implant Stability During Follow-Up: What Is Actually Measured, and Perspectives.

Authors:  Elisabetta M Zanetti; Giulia Pascoletti; Michele Calì; Cristina Bignardi; Giordano Franceschini
Journal:  Biosensors (Basel)       Date:  2018-07-13

10.  Contribution to COVID-19 spread modelling: a physical phenomenological dissipative formalism.

Authors:  Oualid Limam; Mohamed Limam
Journal:  Biomech Model Mechanobiol       Date:  2020-09-25
  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.