Set screw fracture with cage dislocation after two-level transforaminal lumbar interbody fusion (TLIF): a case report.

Philip Johannes Felix Leute1, Ahmed Hammad, Isabel Hoffmann, Sebastian Hoppe, Hans-Michael Klinger, Stefan Lakemeier.   

Abstract

INTRODUCTION: Transforaminal lumbar interbody fusion is a popular procedure used to achieve spondylodesis in patients with degenerative lumbar spinal diseases. We present a rare case of a patient with a set screw fracture with cage dislocation after an open transforaminal lumbar interbody fusion procedure. To the best of our knowledge, this case is the first of its kind to be reported. CASE
PRESENTATION: A 44-year-old Caucasian woman attended a follow-up appointment at our hospital 3 months after treatment for second-degree lumbar spondylolisthesis (L4/L5) and osteochondrosis (L5/S1) with transforaminal lumbar interbody fusion and dorsal spondylodesis. She complained of severe leg pain on the left side. Her physical examination revealed a normal neurological status, except for paresthesia of the entire left lower limb and at the ball of the left foot. Radiological imaging showed breaking of the set screws with cage dislocation. Surgical revision was then performed with exchange of the whole dorsal instrumentation and the dislocated cage. Six weeks post-operatively, the patient was seen again at our clinic without neurological complaints, except for decreased sensitivity on the dorsum of her left foot. The wound healing and radiological follow-up were uneventful.
CONCLUSIONS: Hardware-related complications are rarely seen in patients with open transforaminal lumbar interbody fusion, but must be kept in mind and can potentially cause severe neurological deficits.

Entities:  

Mesh:

Year:  2015        PMID: 25609204      PMCID: PMC4333885          DOI: 10.1186/1752-1947-9-22

Source DB:  PubMed          Journal:  J Med Case Rep        ISSN: 1752-1947


Introduction

Transforaminal lumbar interbody fusion (TLIF) is a popular procedure used to achieve spondylodesis in patients with degenerative lumbar spinal diseases [1]. Neurological, wound-healing and hardware problems have been reported in these patients [2]. Nevertheless, hardware problems with implant fractures are a rare complication of the procedure [3]. We present a case of implant fracture with cage dislocation after an open TLIF procedure. To the best of our knowledge, this is a complication after open TLIF that has not been reported in the literature to date.

Case presentation

A 44-year-old Caucasian woman came to our outpatient clinic for a 3-month follow-up appointment after undergoing an open two-level (L4-S1) TLIF procedure with dorsal spondylodesis. She had been treated for spondylolysis with second-degree lumbar spondylolisthesis (L4/L5) and for osteochondrosis (L5/S1) (Figures  1 and 2). At her follow-up visit, she complained of pain in her whole left leg.
Figure 1

X-ray taken before the patient’s first surgery.

Figure 2

X-ray obtained after the patient’s first surgery.

X-ray taken before the patient’s first surgery. X-ray obtained after the patient’s first surgery. A physical examination revealed pain upon palpation of the lumbar spine. The patient’s spinal muscles were tense. Intermittent paresthesia of her left lower limb and constant paresthesia at the ball of the left foot were present. There was no saddle block paresthesia. Her motor function was MRC (Medical Research Council) Scale 5/5 in all muscles of both legs. Her Lasègue and Bragard signs were positive on the left side. Her bilateral patellar and Achilles tendon reflexes were regular, as was her perianal sensitivity and sphincter muscle tone. No urinary or defecation problems were observed. The patient was in good general condition, but overweight (108 kg weight, 170 cm height). She had sustained no trauma. She had been compliant with the given instructions for sitting and working in a manner that is easy on the back. However, she had not lost any body weight as we had advised. Her past medical history included arterial hypertonus, restless legs syndrome and an axial esophageal hernia. Her past surgical history consisted of a hysterectomy and tonsillectomy, 6 and 31 years ago, respectively. Upon admission to the hospital, she was taking several painkillers, blood pressure medication, a dopamine agonist and a proton pump inhibitor. Her social history revealed that she had consumed nicotine since she was 14 years old (46.5 pack-years). Her family history and review of her organ systems were non-contributory, except for the above-mentioned diseases.Radiography and computed tomography were performed (Figures  3 and 4). A diagnosis of implant failure with a combination of cage dislocation left dorsally and breaking of the set screws on both sides with dislocation of the rods was made. All polyaxial screws were intact and in place, except for the S1 screws, which had become loose and dislocated dorsally on both sides.
Figure 3

X-ray taken at first follow-up examination.

Figure 4

Computed tomographic scan taken at first follow-up examination.

X-ray taken at first follow-up examination. Computed tomographic scan taken at first follow-up examination. A revision surgery was performed by the same surgical team that had done the primary operation. The patient was placed in prone position. A midline incision was made through the existing scar. Removal of cicatricial tissue was performed until all of the screws could be exposed. The S1 polyaxial screws were loose and dislocated on both sides. Metallosis was observed in the surrounding tissue. The S1 set screws were then inspected and removed. They showed a breaking of the worm. The other polyaxial screws were in place and intact. The L5 nerve root was prepared, followed by exposure of the dura. From there, preparation toward the cage in the L5/S1 intervertebral disc space was done. Palpation showed that it had become completely loose. The cage was removed. Next, expansion of the intervertebral disc space, rasping of the edges and insertion of a new cage was done. After that, the whole dorsal instrumentation system was exchanged. Pedicle screws with a 0.5-mm-larger outer diameter than those previously used were placed. Finally, the wound was closed layerwise, and two drains were placed. No intra-operative complications occurred.A post-operative X-ray showed a regular position of the implants (Figure  5). The drains were removed on the second post-operative day. The wound healing was primary and without signs of infection. The patient’s neurological complaints (pain and paresthesia) disappeared completely. Again, physiotherapists instructed the patient to sit and work in a manner that is easy on the back, and she was advised to lose weight and quit smoking. Twelve days after the operation, the patient was discharged to home. From there, she went into a rehabilitation clinic. No peri-operative complications occurred.Six weeks post-operatively, the patient was seen at our outpatient clinic for a follow-up examination (Figure  6). She complained of back pain without radiation to the legs, which was controllable with pain medication. In addition to that, she complained that the dorsum of her left foot had decreased sensitivity. A physical examination revealed a regular scarring of the wound. Paresthesia was present on the dorsum of her left foot. Apart from that, sensitivity was the same bilaterally. Her motor function was MRC (Medical Research Council) Scale 5/5 in all muscles of the legs bilaterally, and she had regular reflexes. Her Lasègue and Bragard signs were negative bilaterally. No post-operative complications were observed.
Figure 5

X-ray taken after the patient’s second surgery.

Figure 6

X-ray taken at the patient’s second follow-up examination.

X-ray taken after the patient’s second surgery. X-ray taken at the patient’s second follow-up examination.

Discussion

To identify similar cases, we reviewed the online databases PubMed, MEDLINE, CareLit and MEDPILOT for the keywords "TLIF," "transforaminal lumbar interbody fusion" and "complication." The search revealed 244 publications. We excluded 204 studies that covered inappropriate topics. This left us with 40 studies with comparable patients and treatments. Implant-related complications were found in 22 of these studies and were present in 76 cases (Table  1) [1-22].
Table 1

Review of the literature

YearAuthorHardware failure not specifiedCage migrationScrew loosening or backing outScrew or cage malpositionCage subsidenceRod fracture
2004Castro et al. [4]000600
2005Hackenberg et al. [5]001000
2006Lauber et al. [6]001000
2006Weiner et al. [7]100000
2006Taneichi et al. [8]011000
2008Yan et al. [9]001000
2009Xu et al. [10]021000
2009Faundez et al. [11]000200
2010Gong et al. [12]001000
2010Fujibayashi et al. [13]000100
2010Wang et al. [14]000100
2011Mura et al. [15]030530
2011Takahashi et al. [2]040000
2012Burneikiene et al. [3]003002
2012Aoki et al. [16]010000
2012Kim et al. [17]010000
2012Lee et al. [18]060000
2012Tormenti et al. [1]01001100
2013Høy et al. [19]000200
2013Lau et al. [20]000200
2013Seng et al. [21]020000
2014Zhang et al. [22]000100
Review of the literature Of these hardware-related complications, 75 were further specified and 1 was unspecified. There were 31 cases with cage or pedicle screw malpositioning, 30 with cage migration, 3 with cage subsidence, 9 with pedicle screw loosening or backing out and 2 with a rod fracture. However, none of these cases were comparable to the one described in the present report, which uniquely consisted of loosening of the polyaxial sacral screws, fracturing of the set screws and cage dislocation. Although our report provides no causal explanation for this kind of implant failure, we assume that the size and placement of the implant did matter in our patient. Furthermore, we believe that the dorsal fixation device may have been too small for our patient’s biomechanical needs. Cho and coworkers reported that pull-out strength of the pedicle screws increased with the outer diameter of the screws [23]. Therefore, we assume that our primary choice to use pedicle screws with a smaller outer diameter led to early loosening of the sacral screws in our patient. Only after exchanging the primary system for a larger one did we achieve good biomechanical stability. The broken set screws make the present case unique. We could not identify any matching cases that have been reported to date. However, in a similar report, Agrawal described a case of nut loosening, and thus a loosened set screw after dorsal instrumentation, but without breakage of the worm [24]. Moreover, nut loosening was reported by Davne and Myers for dorsal instrumentation and posterior lumbar interbody fusion with a variable screw placement system, which uses a plate over which screws are placed into the bone [25]. In their large series of 486 patients, they attributed a certain amount of surgical inexperience to most cases of nut loosening. In our case, we assume that it was not inexperience of the operating surgeons that led to breakage of these screws. However, we acknowledge that, in general, malpositioning of the set screws or flaws in their production can lead to a certain predisposition for the worm to break. With regard to fixation of the pedicle screws, Ha and coworkers showed that side fixation of the pedicle screws led to greater stability than head fixation with a set screw [26]. We were unable to identify any literature addressing whether an increase of the contact face of the set screws also leads to greater stability. Nevertheless, in our patient, the exchange of the internal fixation system led to the implantation of a set screw with a larger contact face. Whether an increase of the contact face of the set screws leads to greater stability remains unclear. Cage migration typically occurs dorsally, as it did in our patient [1]. It is also associated with cage placement in a central position as compared to cages placed in anterior positions [27]. In our patient, the migrated cage between L5 and S1 lay rather dorsally. This may have also predisposed the cage to migrate in this direction. With regard to our patient’s obesity, the available literature shows no difference in operative time, length of hospitalization or complication rates in general between obese and non-obese patients [28]. Nevertheless, this remains a controversial topic, and therefore obesity must be kept in mind as a possible negative factor. The fact that our patient has many pack-years of smoking is unlikely to have contributed to her outcome. Our review of the literature showed that negative effects of smoking in TLIF have not been observed and can probably be excluded as a factor related to poor outcome [1, 2, 20].

Conclusions

The complications we describe in this report are particularly of interest for spinal surgeons. To the best of our knowledge, this report is currently the only available one of its kind in the literature. The following take-home messages must be taken into account. (1) Hardware-related problems are a rare complication in open TLIF, but must be kept in mind. (2) Hardware-related problems must be mentioned when obtaining informed consent from a patient prior to performing open TLIF. (3) Hardware-related complications can potentially cause severe neurological deficits. (4) Lifestyle modifications should be advised for all patients, if appropriate.

Consent

Written informed consent was obtained from the patient for publication of this case report and any accompanying images. A copy of the written consent is available for review by the Editor-in-Chief of this journal.
  28 in total

1.  Reduction and transforaminal lumbar interbody fusion with posterior fixation versus transsacral cage fusion in situ with posterior fixation in the treatment of Grade 2 adult isthmic spondylolisthesis in the lumbosacral spine.

Authors:  Kai Gong; Zhe Wang; Zhuojing Luo
Journal:  J Neurosurg Spine       Date:  2010-09

2.  Paraspinal-approach transforaminal lumbar interbody fusion for the treatment of lumbar foraminal stenosis.

Authors:  Shunsuke Fujibayashi; Masashi Neo; Mitsuru Takemoto; Masato Ota; Takashi Nakamura
Journal:  J Neurosurg Spine       Date:  2010-10

3.  Unilateral versus bilateral instrumented transforaminal lumbar interbody fusion in two-level degenerative lumbar disorders: a prospective randomised study.

Authors:  Kai Zhang; Wei Sun; Chang-qing Zhao; Hua Li; Wei Ding; You-zhuan Xie; Xiao-jiang Sun; Jie Zhao
Journal:  Int Orthop       Date:  2013-08-06       Impact factor: 3.075

4.  Transforaminal lumbar interbody fusion: an independent assessment of outcomes in a difficult patient population.

Authors:  Bradley K Weiner; Hoan-Vu Nguyen; S Will Hazard
Journal:  Med Sci Monit       Date:  2006-02-23

5.  Biomechanical stability according to different configurations of screws and rods.

Authors:  Kee-Yong Ha; Sung-Chul Hwang; Tae-Hyuk Whang
Journal:  J Spinal Disord Tech       Date:  2013-05

6.  Clinical and radiological outcomes of open versus minimally invasive transforaminal lumbar interbody fusion.

Authors:  Kong Hwee Lee; Wai Mun Yue; William Yeo; Henry Soeharno; Seang Beng Tan
Journal:  Eur Spine J       Date:  2012-03-28       Impact factor: 3.134

7.  TLIF for symptomatic disc degeneration: a retrospective study of 100 patients.

Authors:  Pier Paolo Mura; Mauro Costaglioli; Maurizio Piredda; Silvia Caboni; Silvia Casula
Journal:  Eur Spine J       Date:  2011-04-02       Impact factor: 3.134

8.  Comparison of one-level minimally invasive and open transforaminal lumbar interbody fusion in degenerative and isthmic spondylolisthesis grades 1 and 2.

Authors:  Jian Wang; Yue Zhou; Zheng Feng Zhang; Chang Qing Li; Wen Jie Zheng; Jie Liu
Journal:  Eur Spine J       Date:  2010-04-22       Impact factor: 3.134

9.  Transforaminal lumbar interbody fusion (TLIF) versus posterolateral instrumented fusion (PLF) in degenerative lumbar disorders: a randomized clinical trial with 2-year follow-up.

Authors:  Kristian Høy; Cody Bünger; Bent Niederman; Peter Helmig; Ebbe Stender Hansen; Haisheng Li; Thomas Andersen
Journal:  Eur Spine J       Date:  2013-04-13       Impact factor: 3.134

10.  Surgical treatment of adult degenerative spondylolisthesis by instrumented transforaminal lumbar interbody fusion in the Han nationality.

Authors:  Hao Xu; Hao Tang; Zhonghai Li
Journal:  J Neurosurg Spine       Date:  2009-05
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.