Literature DB >> 1631716

Complications of lumbar spinal fusion with transpedicular instrumentation.

S H Davne1, D L Myers.   

Abstract

Transpedicular screw fixation systems are coming into wide use as an adjunct to lumbar spinal fusion procedures. This 5-year series included 486 patients who underwent 533 variable screw placement procedures for discal, degenerative, and spondylolytic problems. The wound infection rate was 2.6%: 0.6% deep, 0.9% graft, and 1.1% superficial. The neural injury rate was 1.1% overall: 0.6% related to posterior lumbar interbody fusion and 0.6% related to instrumentation. Technical problems per procedure included 8.1% difficult screw placement, 5.6% nut loosening, and 4.3% screw breakage (1.1% per screw placed). Device modifications have reduced the incidence of screw breakage and nut loosening. No device-related neural injuries occurred in the last 333 procedures. With experience, the device can be applied safely without significantly increasing the risk of neural injury or morbidity.

Entities:  

Mesh:

Year:  1992        PMID: 1631716     DOI: 10.1097/00007632-199206001-00021

Source DB:  PubMed          Journal:  Spine (Phila Pa 1976)        ISSN: 0362-2436            Impact factor:   3.468


  45 in total

1.  Mechanical implant failure in posterior cervical spine fusion.

Authors:  Takeshi Okamoto; Masashi Neo; Shunsuke Fujibayashi; Hiromu Ito; Mitsuru Takemoto; Takashi Nakamura
Journal:  Eur Spine J       Date:  2011-10-16       Impact factor: 3.134

2.  A preliminary study of reliability of impedance measurement to detect iatrogenic initial pedicle perforation (in the porcine model).

Authors:  Ciaran Bolger; C Carozzo; T Roger; Linda McEvoy; Jabir Nagaria; Gerard Vanacker; Maurice Bourlion
Journal:  Eur Spine J       Date:  2005-10-13       Impact factor: 3.134

Review 3.  Fusion for low-grade adult isthmic spondylolisthesis: a systematic review of the literature.

Authors:  Wilco C H Jacobs; Arnold Vreeling; Marinus De Kleuver
Journal:  Eur Spine J       Date:  2005-10-11       Impact factor: 3.134

4.  Long-term results of ventro-dorsal versus ventral instrumentation fusion in the treatment of spondylitis.

Authors:  O Linhardt; J Matussek; H J Refior; A Krödel
Journal:  Int Orthop       Date:  2006-05-17       Impact factor: 3.075

5.  Facet joint violation during pedicle screw insertion: a cadaveric study of the adult lumbosacral spine comparing the two pedicle screw insertion techniques.

Authors:  Kook Jin Chung; Seung Woo Suh; Keny Swapnil; Jae Hyuk Yang; Hae Ryong Song
Journal:  Int Orthop       Date:  2006-09-30       Impact factor: 3.075

6.  Technical factors related to the incidence of adjacent superior segment facet joint violation after transpedicular instrumentation in the lumbar spine.

Authors:  Zhiming Chen; Jie Zhao; Hao Xu; Aigang Liu; Jiandong Yuan; Cong Wang
Journal:  Eur Spine J       Date:  2008-09-16       Impact factor: 3.134

7.  Early complications of spinal pedicle screw.

Authors:  A A Faraj; J K Webb
Journal:  Eur Spine J       Date:  1997       Impact factor: 3.134

8.  Verification of the position of pedicle screws in lumbar spinal fusion.

Authors:  R G Haaker; U Eickhoff; E Schopphoff; R Steffen; M Jergas; J Krämer
Journal:  Eur Spine J       Date:  1997       Impact factor: 3.134

9.  Spinal fusion: a combined anterior and supplementary interspinous technique.

Authors:  M W Fidler
Journal:  Eur Spine J       Date:  1997       Impact factor: 3.134

10.  Expandable Interbody Spacers: A Two-Year Study Evaluating Radiologic and Clinical Outcomes With Patient-Reported Outcomes.

Authors:  Graham Mulvaney; Steve Monk; Jonathan D Clemente; Deborah Pfortmiller; Domagoj Coric
Journal:  Int J Spine Surg       Date:  2020-10-29
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.