Frank Vandekerckhove1, Victoria Bracke1, Petra De Sutter1.
Abstract
BACKGROUND/AIMS: The objective of this literature study is to investigate the place of recent software technology sonography-based automated volume count (SonoAVC) for the automatic measurement of follicular volumes in IVF/ICSI. Its advantages and disadvantages and potential future developments are evaluated.Entities:
Keywords: SonoAVC; automated follicle measurement; in vitro fertilization; ovarian stimulation; three-dimensional ultrasound
Year: 2014 PMID: 25593942 PMCID: PMC4286967 DOI: 10.3389/fsurg.2014.00018
Source DB: PubMed Journal: Front Surg ISSN: 2296-875X
Figure 1Follicular measurement using sonography-based automated volume count (SonoAVC).
Figure 2The use of volume measurements compared to classical two-dimensional diameters of follicles. v, follicular volume; dx, dy, dz, follicular diameters in three planes; R, follicular radius (diameter/2); dv, reconstructed follicular diameter.
Figure 3Flow chart of the review process. ESHRE, European Society of Human Reproduction and Embryology; ASRM, American Society for Reproductive Medicine.
Systematic review of the literature.
| Study | Year | Type of study | Aims | Variables | Level of evidence |
|---|---|---|---|---|---|
| Ata and Tulandi ( | 2011 | Review | Automated volume measurements in reproduction and pregnancy | B | |
| Ata et al. ( | 2011 | Prospective | Follicles were measured with 2D and SonoAVC in 100 patients | AFC | B |
| Comparison of both techniques | |||||
| Reproducibility | |||||
| Deb et al. ( | 2009 | Prospective | AFC was measured with manual 2D, manual 3D, and SonoAVC in 55 patients | AFC | B |
| Reproducibility | |||||
| Intra- and interobserver reliability | |||||
| Work flow | |||||
| Deb et al. ( | 2010 | Prospective | AFC was measured with 2D and SonoAVC in 24 patients | AFC | B |
| Work flow | |||||
| Comparison of both techniques | |||||
| Deb et al. ( | 2013 | Prospective | To evaluate intracycle variation in AFC | AFC of follicles ≤6 mm | B |
| Deutch et al. ( | 2009 | Prospective | Evaluation of SonoAVC on phantom | Accuracy | B |
| Offline analysis of 3D volumes with SonoAVC | Work flow | ||||
| Lamazou et al. ( | 2009 | Review | Ultrasound in the evaluation of ovarian reserve | B | |
| Lamazou et al. ( | 2010 | Prospective | Aspirated follicle volumes of 72 follicles in 27 patients were compared with 3D volume measurements (VOCAL and SonoAVC) | Accuracy | B |
| Murtinger et al. ( | 2009 | Prospective RCT | The use of SonoAVC in the timing of oocyte maturation (hCG) | Number of (mature) oocytes Number of fertilizations | A2 |
| Comparison of 2D and SonoAVC | |||||
| Pregnancy rates | |||||
| Raine-Fenning et al. ( | 2007 | Conclusion | Evaluation of early studies | Accuracy | D |
| Work flow | |||||
| The importance of image quality | |||||
| Raine-Fenning et al. ( | 2008 | Prospective | The aspirated volume of 224 follicles from 51 patients is compared with ultrasound measurements (manual 2D, VOCAL, SonoAVC) | Accuracy | B |
| Raine-Fenning et al. ( | 2009 | Prospective | Follicle measurements with 2D and SonoAVC on the day of hCG triggering in 89 patients are compared | Comparison of techniques | B |
| Work flow | |||||
| Raine-Fenning et al. ( | 2009 | Prospective | The aspirated volume of 200 follicles is compared with ultrasound measurements (manual 2D, VOCAL, SonoAVC) | Accuracy | B |
| Comparison of techniques | |||||
| Work flow | |||||
| Raine-Fenning et al. ( | 2010 | Prospective RCT | Comparing 2D and SonoAVC to define the moment of hCG triggering | Comparison of techniques | A2 |
| Number of (mature) oocytes | |||||
| Number of fertilizations | |||||
| Number of embryos | |||||
| Rodriguez-Fuentes et al. ( | 2010 | Prospective | Follicle measurements with 2D and SonoAVC in 58 patients on the day of hCG triggering | The importance of image quality | B |
| Work flow | |||||
| The number of mature oocytes | |||||
| Comparison of both techniques | |||||
| Rodriguez et al. ( | 2014 | Prospective | To evaluate the number of procedures to reach proficiency in SonoAVC folliculometry | Learning curve–cumulative summation (LC–CUSUM) | B |
| Rousian et al. ( | 2009 | Prospective | 10 Phantoms studies with VOCAL, inversion mode, SonoAVC, and V-scope | Accuracy | B |
| 24 Gestational sacs in early pregnancy | |||||
| Reproducibility | |||||
| Salama et al. ( | 2010 | Prospective | Comparison of the aspirated volume with VOCAL and SonoAVC measurements in monofollicular cycles in 15 patients | Accuracy | B |
| Comparison of techniques | |||||
| Reproducibility | |||||
| Ata et al. ( | 2010 | Prospective (OCA) | Comparing monitoring of COH with 2D and SonoAVC in 100 patients | Comparison of techniques | – |
| Bouhanna et al. ( | 2012 | Prospective (abstract from ESHRE) | Comparing monitoring of monofollicular cycles with 2D, VOCAL, and SonoAVC in 22 patients | Comparison of techniques Reproducibility | – |
| Deutch et al. ( | 2007 | Prospective (OCA) | Follicle measurements with 2D and SonoAVC of 10 follicles in three saved volume files by two observers | Comparison of techniques | – |
| Intra- and interobserver reliability | |||||
| Work flow | |||||
| Deutch et al. ( | 2007 | Prospective (OCA) | Analysis of 347 follicles from 31 saved volume files in 14 patients | Work flow | – |
| Reproducibility | |||||
| Hernandez et al. ( | 2009 | Prospective (abstract from ESHRE) | Comparing monitoring with 2D, VOCAL, and SonoAVC on the day of hCG triggering in 27 patients | Association between follicle volume and number of mature oocytes | – |
| Comparison of techniques | |||||
| Hernandez et al. ( | 2009 | Prospective (OCA) | Comparing 2D and SonoAVC for monitoring oocyte donors | Comparison of techniques | – |
| Number of (mature) oocytes | |||||
| Number of fertilizations | |||||
| Number of embryos | |||||
| Implantation ratio | |||||
| Pregnancy rate | |||||
| Hernandez et al. ( | 2011 | Prospective (abstract from ESHRE) | Follicle measurements with 2D and SonoAVC in 100 patients | Work flow The importance of image quality | – |
| Raine-Fenning et al. ( | 2007 | OCA | Follicles ranging 11 to 22 mm diameter were measured with 2D, manual 3D, and SonoAVC | Comparison of techniques | – |
| Raine-Fenning et al. ( | 2007 | OCA | The aspirated volume of 100 follicles was compared with 2D, manual 3D, and SonoAVC measurements | Accuracy | – |
| Comparison of techniques | |||||
| Rodriguez-Fuentes et al. ( | 2008 | Prospective (OCA) | Monitoring with 2D, VOCAL, and SonoAVC in 46 patients | Comparison of techniques | – |
| Association between follicle volume and number of mature oocytes | |||||
| Salama et al. ( | 2008 | Prospective (OCA) | Comparing the aspirated volume with VOCAL and SonoAVC measurements in 15 patients with a monodominant follicle | Accuracy | – |
| Comparison of techniques | |||||
| Reproducibility | |||||
| Sanabria et al. ( | 2009 | Prospective (abstract from ESHRE) | Monitoring with 2D or SonoAVC in 42 oocyte donor cycles | Comparison of techniques | – |
| Number of (mature) oocytes | |||||
| Number of fertilizations | |||||
| Number of embryos | |||||
| Implantation ratio | |||||
| Pregnancy rate | |||||
| Saumet et al. ( | 2012 | Prospective (OCA) | Comparison of different ultrasound techniques with AMH to measure ovarian reserve in 174 patients | Comparison of techniques | – |
| Sherbahn et al. ( | 2009 | Retrospective (OCA) | Retrospective study comparing monitoring with 2D and SonoAVC | Comparison of techniques | – |
| Work flow |
OCA, oral communication abstract; ESHRE, European Society of Human Reproduction and Embryology.
Evaluation of work flow.
| Manual 2D | SonoAVC | SonoAVC with post-processing (pp.) | 2D versus SonoAVC with pp. | |
|---|---|---|---|---|
| Raine-Fenning et al. ( | 236.10 ± 57.07 s | 39.00 ± 6.00 s | 180.50 ± 63.55 s | |
| Rodriguez-Fuentes et al. ( | 9.6 min | – | 5.6 min | 4 min time gain per patient |
| Deb et al. ( | 324.47 ± 162.22 s | 41.06 ± 11.12 s | 132.05 ± 56.23 s | |
| Deutch et al. ( | 361 s (129–878) | – | 133 s (73–271) | |
| Deutch et al. ( | 56.8 s | 5.8 s | ||
| Sherbahn et al. ( | 6.2 min | – | 5.6 min | |
| Raine-Fenning et al. ( | 26.36 ± 4.35 s | 9.34 ± 0.67 s | – | |
| Deb et al. ( | 71 ± 8.48 s | 26 ± 2.67 s | 112 ± 29.86 s | |
Investigation of operator independence.
| Manual 2D | SonoAVC | SonoAVC with pp. | |||
|---|---|---|---|---|---|
| Deb et al. ( | 19.26 ± 10.55* | 6.51 ± 4.79* | 18.42 ± 10.53* | 0.006** | |
| ICC = 0.979 | ICC = 0.997 | ||||
| Salama et al. ( | ICC < 0.08 | ICC = 0.97 | |||
| Bouhanna et al. ( | Intra observer variability | ICC = 0.62 | ICC = 0.96 | ||
| Inter observer variability | ICC = 0.83 | ICC = 0.94 | |||
ICC, intra class correlation coefficient; *Mean ± SD; **Comparison of manual 2D and SonoAVC with pp.