| Literature DB >> 25550410 |
Brett Newswanger1, Steve Ammons1, Neelima Phadnis1, W Kenneth Ward2, Jessica Castle3, Robert W Campbell4, Steven J Prestrelski5.
Abstract
Despite a vigorous research effort, to date, the development of systems that achieve glucagon stability in aqueous formulations (without reconstitution) has failed to produce any clinical candidates. We have developed a novel, nonaqueous glucagon formulation based on a biocompatible pharmaceutical solvent, dimethyl sulfoxide, which demonstrates excellent physical and chemical stability at relatively high concentrations and at high temperatures. This article reports the development of a novel, biocompatible, nonaqueous native human glucagon formulation for potential use in subcutaneous infusion pump systems. Data are presented that demonstrate physical and chemical stability under presumed storage conditions (>2 years at room temperature) as well as "in use" stability and compatibility in an Insulet's OmniPod(®) infusion pump. Also presented are results of a skin irritation study in a rabbit model and pharmacokinetics/pharmacodynamics data following pump administration of glucagon in a diabetic swine model. This nonaqueous glucagon formulation is suitable for further clinical development in pump systems.Entities:
Keywords: glucagon; hypoglycemia; infusion; nonaqueous
Mesh:
Substances:
Year: 2015 PMID: 25550410 PMCID: PMC4495537 DOI: 10.1177/1932296814565131
Source DB: PubMed Journal: J Diabetes Sci Technol ISSN: 1932-2968