Reliability of Tonolab measurements in rats.

Li-Fang Liu1, Chu-Kai Huang1, Ming-Zhi Zhang1.   

Abstract

AIM: To assess the repeatability and reproducibility of Tonolab tonometer in rats with high intraocular pressure (IOP) and evaluate its ability to detect IOP changes in rats with general anaesthesia.
METHODS: Left eyes of adult Fischer rats (F344) were photocoagulated by 532 nm diode laser to induce high IOP. Hypertensive eyes of 30 conscious rats were randomly chosen to measure IOP on a single occasion. Two observers independently and alternately undertook IOP measurements consecutively for three times using the same Tonolab tonometer blind to the other observer's IOP measurements. The within subject standard deviation (Sw), coefficient of variation (CVw) (100×Sw/overall mean), and intraclass correlation coefficient (ICC) were calculated to evaluate intra-observer repeatability. Inter-observer difference was analysed by using 95% limits of agreement described by Bland-Altman and paired sample t-test. Also, another 13 normal F344 rats were intraperitoneally administrated with ketamine/xylazine or chloral hydrate, and IOPs of both eyes were measured by a single operator once every 5min until animals came to conscious. IOPs at various time points were compared by using one-way ANOVAs.
RESULTS: Mean IOP was 35.58 mm Hg (range 17.33 to 65.33 mm Hg). For intraobserver repeatability, the Sw, CVw and ICC of high IOP for two observers were 5.20 mm Hg/3.41 mm Hg, 9.98%/8.08% and 0.820/0.928 respectively. The inter-observer difference was 14.76%±19.76% of the mean IOP of two observers, with a 95% limits of agreement -23.97% to 53.50%, and the difference between mean IOP of these two observers was statistically significant (P=0.001). IOPs dropped slightly during the first 15min post-aneathesia, with a IOP change between 0.17 and 1.17 mm Hg. IOPs changed from basline of 11.75±2.05 mm Hg (n=12) to 8.75±1.06 mm Hg 20min post-anesthesia (P=0.001), and this hypotensive condition persisted until 80min post-anesthesia.
CONCLUSION: In this sample of hypertensive rats, Tonolab measurements demonstrated high levels of intraobserver repeatability, however, its interobserver reproducibility was poor. Longitudinal changes of IOP caused by genral anaesthesia can be sensitively detected by Tonolab. So we suggested that measurements of IOP using Tonolab are best measured by a single observer, and it could be included in experimental glaucoma.

Entities:  

Keywords:  Tonolab; rats; repeatability; reproducibility

Year:  2014        PMID: 25540741      PMCID: PMC4270983          DOI: 10.3980/j.issn.2222-3959.2014.06.03

Source DB:  PubMed          Journal:  Int J Ophthalmol        ISSN: 2222-3959            Impact factor:   1.779


  27 in total

1.  Noninvasive determination of intraocular pressure (IOP) in nonsedated mice of 5 different inbred strains.

Authors:  Nicholas Nissirios; David Goldblum; Kaspar Rohrer; Thomas Mittag; John Danias
Journal:  J Glaucoma       Date:  2007-01       Impact factor: 2.503

2.  Noninvasive measurement of rodent intraocular pressure with a rebound tonometer.

Authors:  Wan-Heng Wang; J Cameron Millar; Iok-Hou Pang; Martin B Wax; Abbot F Clark
Journal:  Invest Ophthalmol Vis Sci       Date:  2005-12       Impact factor: 4.799

3.  Measurement error.

Authors:  J M Bland; D G Altman
Journal:  BMJ       Date:  1996-09-21

4.  The induction/impact tonometer: a new instrument to measure intraocular pressure in the rat.

Authors:  A I Kontiola; D Goldblum; T Mittag; J Danias
Journal:  Exp Eye Res       Date:  2001-12       Impact factor: 3.467

5.  Manometric calibration and comparison of TonoLab and TonoPen tonometers in rats with experimental glaucoma and in normal mice.

Authors:  Mary E Pease; John C Hammond; Harry A Quigley
Journal:  J Glaucoma       Date:  2006-12       Impact factor: 2.503

Review 6.  Sleep circuitry and the hypnotic mechanism of GABAA drugs.

Authors:  Jun Lu; Mary Ann Greco
Journal:  J Clin Sleep Med       Date:  2006-04-15       Impact factor: 4.062

7.  Tonopen measurement of intraocular pressure in mice.

Authors:  Herbert A Reitsamer; Jeffrey W Kiel; Joseph M Harrison; Nancy L Ransom; Stuart J McKinnon
Journal:  Exp Eye Res       Date:  2004-04       Impact factor: 3.467

8.  Intertester reliability of the talk test in a cardiac rehabilitation population.

Authors:  Annemette Krintel Petersen; Thomas Maribo; Vibeke Elisabeth Hjortdal; Susanne Laustsen
Journal:  J Cardiopulm Rehabil Prev       Date:  2014 Jan-Feb       Impact factor: 2.081

9.  A comparison of medetomidine and its active enantiomer dexmedetomidine when administered with ketamine in mice.

Authors:  Wesley M Burnside; Paul A Flecknell; Angus I Cameron; Aurélie A Thomas
Journal:  BMC Vet Res       Date:  2013-03-13       Impact factor: 2.741

10.  Repeatability and reproducibility of measurements using a NT-530P noncontact tono/pachymeter and correlation of central corneal thickness with intraocular pressure.

Authors:  Fusako Fujimura; Kazutaka Kamiya; Kazuko Fujiwara; Nobuyuki Shoji; Kimiya Shimizu
Journal:  Biomed Res Int       Date:  2013-10-08       Impact factor: 3.411

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.