Reliability of magnetic resonance elastography using multislice two-dimensional spin-echo echo-planar imaging (SE-EPI) and three-dimensional inversion reconstruction for assessing renal stiffness.

Gavin Low1,2, Nicola E Owen3, Ilse Joubert1, Andrew J Patterson1, Martin J Graves1, Kevin J Glaser4, Graeme J M Alexander3, David J Lomas1.   

Abstract

BACKGROUND: To evaluate the reliability of MRE using a spin-echo echo-planar imaging (SE-EPI) renal MRE technique in healthy volunteers.
METHODS: Institutional review board approved prospective study in which all participants provided written informed consent. Sixteen healthy volunteers comprising seven males and nine females with a median age of 35 years (age range: 23 to 59 years) were included. Coronal 90 Hz and 60 Hz MRE acquisitions were performed twice within a 30-min interval between examinations. Renal MRE reliability was assessed by (i) test-retest repeatability, and (ii) inter-rater agreement between two independent readers. The MRE-measured averaged renal stiffness values were evaluated using: intraclass correlation coefficient (ICC), Bland-Altman and the within-subject coefficient of variation (COV).
RESULTS: For test-retest repeatability, Bland-Altman showed a mean stiffness difference between examinations of 0.07 kPa (95% limits of agreement: -1.41, 1.54) at 90 Hz and 0.01 kPa (95% limits of agreement: -0.51, 0.53) at 60 Hz. Coefficient of repeatability was 1.47 kPa and 0.52 kPa at 90 Hz and 60 Hz, respectively. The within-subject COV was 13.6% and 7.7% at 90 Hz and 60 Hz, respectively. ICC values were 0.922 and 0.907 for test-retest repeatability and 0.998 and 0.989 for inter-rater agreement, respectively (P < 0.001).
CONCLUSION: SE-EPI renal MRE is a reliable technique.
© 2014 Wiley Periodicals, Inc.

Entities:  

Keywords:  diagnostic reliability; kidneys; magnetic resonance elastography

Mesh:

Year:  2014        PMID: 25537823      PMCID: PMC4560097          DOI: 10.1002/jmri.24826

Source DB:  PubMed          Journal:  J Magn Reson Imaging        ISSN: 1053-1807            Impact factor:   4.813


  33 in total

Review 1.  Measuring agreement in method comparison studies.

Authors:  J M Bland; D G Altman
Journal:  Stat Methods Med Res       Date:  1999-06       Impact factor: 3.021

2.  ACR guidance document for safe MR practices: 2007.

Authors:  Emanuel Kanal; A James Barkovich; Charlotte Bell; James P Borgstede; William G Bradley; Jerry W Froelich; Tobias Gilk; J Rod Gimbel; John Gosbee; Ellisa Kuhni-Kaminski; James W Lester; John Nyenhuis; Yoav Parag; Daniel J Schaefer; Elizabeth A Sebek-Scoumis; Jeffrey Weinreb; Loren A Zaremba; Pamela Wilcox; Leonard Lucey; Nancy Sass
Journal:  AJR Am J Roentgenol       Date:  2007-06       Impact factor: 3.959

3.  Nephrogenic systemic fibrosis, gadolinium, and iron mobilization.

Authors:  Sundararaman Swaminathan; Thomas D Horn; Donna Pellowski; Sameh Abul-Ezz; Joshua A Bornhorst; Sanjaya Viswamitra; Sudhir V Shah
Journal:  N Engl J Med       Date:  2007-08-16       Impact factor: 91.245

Review 4.  MRI safety update 2008: part 1, MRI contrast agents and nephrogenic systemic fibrosis.

Authors:  Frank G Shellock; Alberto Spinazzi
Journal:  AJR Am J Roentgenol       Date:  2008-10       Impact factor: 3.959

5.  Magnetic resonance elastography for the noninvasive staging of liver fibrosis.

Authors:  Laurent Huwart; Christine Sempoux; Eric Vicaut; Najat Salameh; Laurence Annet; Etienne Danse; Frank Peeters; Leon C ter Beek; Jacques Rahier; Ralph Sinkus; Yves Horsmans; Bernard E Van Beers
Journal:  Gastroenterology       Date:  2008-04-04       Impact factor: 22.682

6.  Magnetic resonance elastography by direct visualization of propagating acoustic strain waves.

Authors:  R Muthupillai; D J Lomas; P J Rossman; J F Greenleaf; A Manduca; R L Ehman
Journal:  Science       Date:  1995-09-29       Impact factor: 47.728

7.  MR elastography of the liver: preliminary results.

Authors:  Olivier Rouvière; Meng Yin; M Alex Dresner; Phillip J Rossman; Lawrence J Burgart; Jeff L Fidler; Richard L Ehman
Journal:  Radiology       Date:  2006-08       Impact factor: 11.105

8.  Assessment of hepatic fibrosis with magnetic resonance elastography.

Authors:  Meng Yin; Jayant A Talwalkar; Kevin J Glaser; Armando Manduca; Roger C Grimm; Phillip J Rossman; Jeff L Fidler; Richard L Ehman
Journal:  Clin Gastroenterol Hepatol       Date:  2007-10       Impact factor: 11.382

9.  Liver fibrosis: noninvasive assessment with MR elastography versus aspartate aminotransferase-to-platelet ratio index.

Authors:  Laurent Huwart; Christine Sempoux; Najat Salameh; Jacques Jamart; Laurence Annet; Ralph Sinkus; Frank Peeters; Leon C ter Beek; Yves Horsmans; Bernard E Van Beers
Journal:  Radiology       Date:  2007-11       Impact factor: 11.105

10.  Liver fibrosis: non-invasive assessment with MR elastography.

Authors:  Laurent Huwart; Frank Peeters; Ralph Sinkus; Laurence Annet; Najat Salameh; Leon C ter Beek; Yves Horsmans; Bernard E Van Beers
Journal:  NMR Biomed       Date:  2006-04       Impact factor: 4.044

View more
  12 in total

Review 1.  General review of magnetic resonance elastography.

Authors:  Gavin Low; Scott A Kruse; David J Lomas
Journal:  World J Radiol       Date:  2016-01-28

2.  Imaging brain function with simultaneous BOLD and viscoelasticity contrast: fMRI/fMRE.

Authors:  Patricia S Lan; Kevin J Glaser; Richard L Ehman; Gary H Glover
Journal:  Neuroimage       Date:  2020-02-01       Impact factor: 6.556

Review 3.  Magnetic Resonance Imaging of the Fibrotic Kidney.

Authors:  Glen R Morrell; Jeff L Zhang; Vivian S Lee
Journal:  J Am Soc Nephrol       Date:  2017-08-07       Impact factor: 10.121

4.  Magnetic Resonance Elastography of the Liver: Qualitative and Quantitative Comparison of Gradient Echo and Spin Echo Echoplanar Imaging Sequences.

Authors:  Mathilde Wagner; Cecilia Besa; Jad Bou Ayache; Temel Kaya Yasar; Octavia Bane; Maggie Fung; Richard L Ehman; Bachir Taouli
Journal:  Invest Radiol       Date:  2016-09       Impact factor: 6.016

5.  Magnetic resonance elastography can monitor changes in medullary stiffness in response to treatment in the swine ischemic kidney.

Authors:  Xin Zhang; Xiangyang Zhu; Christopher Martyn Ferguson; Kai Jiang; Tyson Burningham; Amir Lerman; Lilach Orly Lerman
Journal:  MAGMA       Date:  2017-12-30       Impact factor: 2.310

6.  Magnetic Resonance Elastography of kidneys: SE-EPI MRE reproducibility and its comparison to GRE MRE.

Authors:  Deep Gandhi; Prateek Kalra; Brian Raterman; Xiaokui Mo; Huiming Dong; Arunark Kolipaka
Journal:  NMR Biomed       Date:  2019-07-22       Impact factor: 4.044

7.  MR Elastography of the Abdomen: Experimental Protocols.

Authors:  Suraj D Serai; Meng Yin
Journal:  Methods Mol Biol       Date:  2021

8.  Magnetic resonance elastography-derived stiffness of the kidneys and its correlation with water perfusion.

Authors:  Deep Gandhi; Prateek Kalra; Brian Raterman; Xiaokui Mo; Huiming Dong; Arunark Kolipaka
Journal:  NMR Biomed       Date:  2019-12-30       Impact factor: 4.044

9.  The utility of magnetic resonance imaging for noninvasive evaluation of diabetic nephropathy.

Authors:  Robert S Brown; Maryellen R M Sun; Isaac E Stillman; Teresa L Russell; Sylvia E Rosas; Jesse L Wei
Journal:  Nephrol Dial Transplant       Date:  2020-06-01       Impact factor: 7.186

Review 10.  Innovative Perspective: Gadolinium-Free Magnetic Resonance Imaging in Long-Term Follow-Up after Kidney Transplantation.

Authors:  Mick J M van Eijs; Arjan D van Zuilen; Anneloes de Boer; Martijn Froeling; Tri Q Nguyen; Jaap A Joles; Tim Leiner; Marianne C Verhaar
Journal:  Front Physiol       Date:  2017-05-16       Impact factor: 4.566

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.