Effect of secondary particles on image quality of dynamic flat panels in carbon ion scanning beam treatment.

S Mori1, S Amano, T Furukawa, T Shirai, K Noda.   

Abstract

OBJECTIVE: Real-time markerless tumour tracking using radiographic fluoroscopic imaging is one of the better solutions to improving respiratory-gated radiotherapy. However, particle beams cause secondary particles from patients, which could affect radiographs. Here, we evaluated the quality of radiographs during carbon ion pencil beam scanning (CPBS) irradiation for respiratory gating.
METHODS: A water phantom and chest phantom were used. The phantoms were irradiated with CPBS at 290 MeV n(-1) from orthogonal directions. Dose rates were 3.4 × 10(8), 1.14 × 10(8) and 3.79 × 10(7) particles per second. A dynamic flat panel detector (DFPD) was installed on the upstream (DFPD1) or downstream (DFPD2) side of the vertical irradiation port. DFPD images were acquired during CPBS at 15.00, 7.50 and 3.75 frames per second (fps). Charge on the DFPD was cleaned using fast readout technique every 30 fps. DFPD images were acquired during CPBS with radiographic exposure, and results with and without fast readout technique were compared.
RESULTS: Secondary particles were visualized as spots or streak-like shapes. Capture of secondary particles from the horizontal beam direction was lower with fast readout technique than without it. With regard to beam irradiation direction dependency, CPBS from the horizontal direction resulted in a greater magnitude of secondary particles reaching DFPD2 than reaching DFPD1. When CPBS was delivered from the vertical direction, however, the magnitude of secondary particles on both DFPDs was very similar.
CONCLUSION: Fast readout technique minimized the effect of secondary particles on DFPD images during CPBS. ADVANCES IN KNOWLEDGE: This technique may be useful for markerless tumour tracking for respiratory gating.

Entities:  

Mesh:

Substances:

Year:  2014        PMID: 25536444      PMCID: PMC4651186          DOI: 10.1259/bjr.20140567

Source DB:  PubMed          Journal:  Br J Radiol        ISSN: 0007-1285            Impact factor:   3.039


  22 in total

1.  Tracking moving organs in real time.

Authors:  Martin J Murphy
Journal:  Semin Radiat Oncol       Date:  2004-01       Impact factor: 5.934

2.  Prediction of respiratory tumour motion for real-time image-guided radiotherapy.

Authors:  Gregory C Sharp; Steve B Jiang; Shinichi Shimizu; Hiroki Shirato
Journal:  Phys Med Biol       Date:  2004-02-07       Impact factor: 3.609

3.  Non-stationarity of breath-by-breath ventilation and approaches to modelling the phenomenon.

Authors:  P J Liang; J J Pandit; P A Robbins
Journal:  Adv Exp Med Biol       Date:  1995       Impact factor: 2.622

4.  Towards fluoroscopic respiratory gating for lung tumours without radiopaque markers.

Authors:  Ross I Berbeco; Hassan Mostafavi; Gregory C Sharp; Steve B Jiang
Journal:  Phys Med Biol       Date:  2005-09-13       Impact factor: 3.609

5.  Effects of a difference in respiratory cycle between treatment planning and irradiation for phase-controlled rescanning and carbon pencil beam scanning.

Authors:  S Mori; T Inaniwa; T Furukawa; S Zenklusen; T Shirai; K Noda
Journal:  Br J Radiol       Date:  2013-07-05       Impact factor: 3.039

6.  Lung tumor tracking during stereotactic radiotherapy treatment with the CyberKnife: Marker placement and early results.

Authors:  J J Nuyttens; J-B Prévost; J Praag; M Hoogeman; R J Van Klaveren; P C Levendag; P M T Pattynama
Journal:  Acta Oncol       Date:  2006       Impact factor: 4.089

Review 7.  Breathing pattern in humans: diversity and individuality.

Authors:  G Benchetrit
Journal:  Respir Physiol       Date:  2000-09

8.  Integrated radiotherapy imaging system (IRIS): design considerations of tumour tracking with linac gantry-mounted diagnostic x-ray systems with flat-panel detectors.

Authors:  Ross I Berbeco; Steve B Jiang; Gregory C Sharp; George T Chen; Hassan Mostafavi; Hiroki Shirato
Journal:  Phys Med Biol       Date:  2004-01-21       Impact factor: 3.609

9.  Multiple template-based fluoroscopic tracking of lung tumor mass without implanted fiducial markers.

Authors:  Ying Cui; Jennifer G Dy; Gregory C Sharp; Brian Alexander; Steve B Jiang
Journal:  Phys Med Biol       Date:  2007-10-01       Impact factor: 3.609

10.  Comparison of carbon ion radiotherapy to photon radiation alone or in combination with temozolomide in patients with high-grade gliomas: explorative hypothesis-generating retrospective analysis.

Authors:  Stephanie E Combs; Thomas Bruckner; Jun-Etso Mizoe; Tadashi Kamada; Hirohiko Tsujii; Meinhard Kieser; Jürgen Debus
Journal:  Radiother Oncol       Date:  2013-08-06       Impact factor: 6.280

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.