Validation of a conical cuff on the forearm for estimating radial artery blood pressure.

Lawrence T Hersh1, Jean C Sesing, William J Luczyk, Bruce A Friedman, Shanggen Zhou, Paul B Batchelder.   

Abstract

INTRODUCTION: There has been a growing need for and interest in measuring noninvasive blood pressure (NIBP) in obese patients. In many situations, available rectangular upper arm blood pressure cuffs do not fit properly, closing in a crisscross manner or overlapping the elbow. To address this issue, GE Healthcare has designed a conically shaped cuff for use on the forearm to estimate radial arterial blood pressure. This study evaluated using this forearm cuff with an oscillometric NIBP algorithm compared with an invasive radial arterial blood pressure reference. PATIENTS AND METHODS: Thirty-four patients were enrolled in the study with an aim to acquire a minimum of 150 paired measurements. Blood pressure was measured and recorded invasively from the radial artery of each patient, while noninvasive oscillometric measurements were acquired from the same limb using a conically shaped cuff placed on the patient's forearm. ANALYSIS: NIBP values were compared with the invasive blood pressure values acquired in the combined 30-s period preceding cuff inflation and the 30-s period following cuff deflation. The acceptance criteria for measurement accuracy were defined in accordance with the ANSI/AAMI/ISO 81060-2:2009 standard, which requires an absolute average error of 5 mmHg or less, with an SD of 8 mmHg or less.
RESULTS: The systolic mean error (-0.82 mmHg) and SD (6.08 mmHg) and the diastolic mean error (1.53 mmHg) and SD (3.83 mmHg) were within the 81060-2 acceptance criteria.
CONCLUSION: The statistical results show that oscillometric NIBP measurements taken with the conically shaped cuff placed on the forearm give an accurate estimation of radial arterial blood pressure.

Entities:  

Mesh:

Year:  2014        PMID: 24217368     DOI: 10.1097/MBP.0000000000000011

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  6 in total

1.  Evaluation of filtering methods for acquiring radial intra-artery blood pressure waveforms.

Authors:  Lawrence T Hersh; Bruce Friedman; William Luczyk; Jean Sesing
Journal:  J Clin Monit Comput       Date:  2014-12-17       Impact factor: 2.502

2.  Prognostic Value of Ambulatory Blood Pressure Monitoring in Obese Patients.

Authors:  Eoin O'Brien
Journal:  J Clin Hypertens (Greenwich)       Date:  2015-10-05       Impact factor: 3.738

Review 3.  Anesthetic management of obese and morbidly obese parturients.

Authors:  Sang Tae Kim
Journal:  Anesth Pain Med (Seoul)       Date:  2021-10-29

4.  Cuff challenges in blood pressure measurement.

Authors:  Paolo Palatini; Roland Asmar
Journal:  J Clin Hypertens (Greenwich)       Date:  2018-07       Impact factor: 3.738

Review 5.  Managing anesthesia for cesarean section in obese patients: current perspectives.

Authors:  Agnes M Lamon; Ashraf S Habib
Journal:  Local Reg Anesth       Date:  2016-08-16

6.  Comparison of the non-invasive Nexfin® monitor with conventional methods for the measurement of arterial blood pressure in moderate risk orthopaedic surgery patients.

Authors:  Felix Balzer; Marit Habicher; Michael Sander; Julian Sterr; Stephanie Scholz; Aarne Feldheiser; Michael Müller; Carsten Perka; Sascha Treskatsch
Journal:  J Int Med Res       Date:  2016-05-02       Impact factor: 1.671

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.