Literature DB >> 25506113

A New Time Calibration Method for Switched-capacitor-array-based Waveform Samplers.

H Kim1, C-T Chen1, N Eclov1, A Ronzhin2, P Murat2, E Ramberg2, S Los2, W Moses3, W-S Choong3, C-M Kao1.   

Abstract

We have developed a new time calibration method for the DRS4 waveform sampler that enables us to precisely measure the non-uniform sampling interval inherent in the switched-capacitor cells of the DRS4. The method uses the proportionality between the differential amplitude and sampling interval of adjacent switched-capacitor cells responding to a sawtooth-shape pulse. In the experiment, a sawtooth-shape pulse with a 40 ns period generated by a Tektronix AWG7102 is fed to a DRS4 evaluation board for calibrating the sampling intervals of all 1024 cells individually. The electronic time resolution of the DRS4 evaluation board with the new time calibration is measured to be ~2.4 ps RMS by using two simultaneous Gaussian pulses with 2.35 ns full-width at half-maximum and applying a Gaussian fit. The time resolution dependencies on the time difference with the new time calibration are measured and compared to results obtained by another method. The new method could be applicable for other switched-capacitor-array technology-based waveform samplers for precise time calibration.

Entities:  

Keywords:  Switched-Capacitors; Time Calibration; Waveform Sampling

Year:  2014        PMID: 25506113      PMCID: PMC4260358          DOI: 10.1016/j.nima.2014.08.025

Source DB:  PubMed          Journal:  Nucl Instrum Methods Phys Res A        ISSN: 0168-9002            Impact factor:   1.455


  1 in total

1.  A Prototype TOF PET Detector Module Using a Micro-Channel Plate Photomultiplier Tube with Waveform Sampling.

Authors:  H Kim; C-T Chen; H Frisch; F Tang; C-M Kao
Journal:  Nucl Instrum Methods Phys Res A       Date:  2012-01-11       Impact factor: 1.455

  1 in total
  3 in total

1.  A feasibility study of a PET/MRI insert detector using strip-line and waveform sampling data acquisition.

Authors:  H Kim; C-T Chen; N Eclov; A Ronzhin; P Murat; E Ramberg; S Los; Alice M Wyrwicz; Limin Li; C-M Kao
Journal:  Nucl Instrum Methods Phys Res A       Date:  2015-06-01       Impact factor: 1.455

2.  Note: Rapid measurement of fluorescence lifetimes using SiPM detection and waveform sampling.

Authors:  H-M Tsai; J S Souris; H-J Kim; S-H Cheng; L Chen; L-W Lo; C-T Chen; C-M Kao
Journal:  Rev Sci Instrum       Date:  2017-09       Impact factor: 1.523

3.  Multiplexing Readout for Time-of-Flight (TOF) PET Detectors Using Striplines.

Authors:  Heejong Kim; Chien-Min Kao; Yuexuan Hua; Qingguo Xie; Chin-Tu Chen
Journal:  IEEE Trans Radiat Plasma Med Sci       Date:  2021-01-13
  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.