| Literature DB >> 25479830 |
Jingsheng Shi1, Jie Chen1, Jianguo Wu1, Feiyan Chen1, Guangyong Huang1, Zhan Wang1, Guanglei Zhao1, Yibing Wei1, Siqun Wang1.
Abstract
BACKGROUND: The aim of this study was to contrast the collapse values of the postoperative weight-bearing areas of different tantalum rod implant positions, fibula implantation, and core decompression model and to investigate the advantages and disadvantages of tantalum rod implantation in different ranges of osteonecrosis in comparison with other methods.Entities:
Mesh:
Substances:
Year: 2014 PMID: 25479830 PMCID: PMC4266390 DOI: 10.12659/MSM.890920
Source DB: PubMed Journal: Med Sci Monit ISSN: 1234-1010
Figure 13D finite element model of the upper femur. (A) Frontal and rear images of the 3D finite element model in the Abaqus software. Upper: Displacement cloud map; Lower: Stress cloud map. (B) Schematic of tantalum rod implantation into the different positions of the osteonecrosis area (from left to right: outside, middle, and inside) for ONFH in the cone-shaped range of 60°, 90°, and 120°.
Figure 2Tantalum rod assembly model for ONFH in the range of 60°. (A) Cancellous bone and tantalum rod assembly model for ONFH in the range of 60° (B) Cancellous bone, tantalum rod, and artificial bone assembly model for ONFH in the 60° range.
Mean displacement in the weight-bearing area of the femoral head of the 3D finite element model (mm).
| 60° | 90° | 120° | Normal | |
|---|---|---|---|---|
| Without treatment | 0.870 | 1.109 | 1.547 | 0.053 |
| Simple core decompression | 1.896 | 1.926 | 1.995 | |
| Tantalum rod implantation inside of the osteonecrosis area | 0.082 | 0.076 | 0.076 | |
| Tantalum rod implantation in the middle of the osteonecrosis area | 0.087 | 0.085 | 0.081 | |
| Tantalum rod implantation outside of the osteonecrosis area | 0.088 | 0.087 | 0.095 | |
| 5-mm shortening implant | 0.084 | 0.072 | 0.063 | |
| Fibula implant | 0.083 | 0.082 | 0.080 |
Figure 3Collapse values in the weight-bearing area of the femoral head of the different osteonecrosis ranges of finite element models and different tantalum rod positions.
Von Mises stress mean on the implant surface of artificial 3D finite element model (MPa).
| 60°osteonecrosis | 90°osteonecrosis | 120°osteonecrosis | |
|---|---|---|---|
| Tantalum rod implant inside of the osteonecrosis area | 0.336 | 0.471 | 0.532 |
| Tantalum rod implant in the middle of the osteonecrosis area | 0.390 | 0.509 | 0.684 |
| Tantalum rod implant outside of the osteonecrosis area | 0.372 | 0.457 | 0.634 |
| Tantalum rod shortening 5-mm implant | 0.509 | 0.527 | 0.706 |
| Fibula implant | 0.372 | 0.517 | 0.741 |
Figure 4Stress on the surface of artificial bone implant area of the 3D finite element model and stress on the surface of artificial bone implant area in different positions.