Development of a 6DOF robotic motion phantom for radiation therapy.

Andrew H Belcher1, Xinmin Liu1, Zachary Grelewicz1, Erik Pearson1, Rodney D Wiersma1.   

Abstract

PURPOSE: The use of medical technology capable of tracking patient motion or positioning patients along 6 degree-of-freedom (6DOF) has steadily increased in the field of radiation therapy. However, due to the complex nature of tracking and performing 6DOF motion, it is critical that such technology is properly verified to be operating within specifications in order to ensure patient safety. In this study, a robotic motion phantom is presented that can be programmed to perform highly accurate motion along any X (left-right), Y (superior-inferior), Z (anterior-posterior), pitch (around X), roll (around Y), and yaw (around Z) axes. In addition, highly synchronized motion along all axes can be performed in order to simulate the dynamic motion of a tumor in 6D. The accuracy and reproducibility of this 6D motion were characterized.
METHODS: An in-house designed and built 6D robotic motion phantom was constructed following the Stewart-Gough parallel kinematics platform archetype. The device was controlled using an inverse kinematics formulation, and precise movements in all 6 degrees-of-freedom (X, Y, Z, pitch, roll, and yaw) were performed, both simultaneously and separately for each degree-of-freedom. Additionally, previously recorded 6D cranial and prostate motions were effectively executed. The robotic phantom movements were verified using a 15 fps 6D infrared marker tracking system and the measured trajectories were compared quantitatively to the intended input trajectories. The workspace, maximum 6D velocity, backlash, and weight load capabilities of the system were also established.
RESULTS: Evaluation of the 6D platform demonstrated translational root mean square error (RMSE) values of 0.14, 0.22, and 0.08 mm over 20 mm in X and Y and 10 mm in Z, respectively, and rotational RMSE values of 0.16°, 0.06°, and 0.08° over 10° of pitch, roll, and yaw, respectively. The robotic stage also effectively performed controlled 6D motions, as well as reproduced cranial trajectories over 15 min, with a maximal RMSE of 0.04 mm translationally and 0.04° rotationally, and a prostate trajectory over 2 min, with a maximal RMSE of 0.06 mm translationally and 0.04° rotationally.
CONCLUSIONS: This 6D robotic phantom has proven to be accurate under clinical standards and capable of reproducing tumor motion in 6D. Such functionality makes the robotic phantom usable for either quality assurance or research purposes.

Entities:  

Mesh:

Year:  2014        PMID: 25471951      PMCID: PMC4235649          DOI: 10.1118/1.4900828

Source DB:  PubMed          Journal:  Med Phys        ISSN: 0094-2405            Impact factor:   4.071


  28 in total

1.  Inclusion of geometrical uncertainties in radiotherapy treatment planning by means of coverage probability.

Authors:  J C Stroom; H C de Boer; H Huizenga; A G Visser
Journal:  Int J Radiat Oncol Biol Phys       Date:  1999-03-01       Impact factor: 7.038

2.  A breathing thorax phantom with independently programmable 6D tumour motion for dosimetric measurements in radiation therapy.

Authors:  P Steidl; D Richter; C Schuy; E Schubert; Th Haberer; M Durante; C Bert
Journal:  Phys Med Biol       Date:  2012-03-29       Impact factor: 3.609

3.  Evaluation of a new six degrees of freedom couch for radiation therapy.

Authors:  D Schmidhalter; M K Fix; M Wyss; N Schaer; P Munro; S Scheib; P Kunz; P Manser
Journal:  Med Phys       Date:  2013-11       Impact factor: 4.071

4.  Use of the BrainLAB ExacTrac X-Ray 6D system in image-guided radiotherapy.

Authors:  Jian-Yue Jin; Fang-Fang Yin; Stephen E Tenn; Paul M Medin; Timothy D Solberg
Journal:  Med Dosim       Date:  2008-04-01       Impact factor: 1.482

5.  Clinical evaluation of a robotic 6-degree of freedom treatment couch for frameless radiosurgery.

Authors:  Thierry Gevaert; Dirk Verellen; Benedikt Engels; Tom Depuydt; Karina Heuninckx; Koen Tournel; Michael Duchateau; Truus Reynders; Mark De Ridder
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-09-22       Impact factor: 7.038

6.  Comparison between infrared optical and stereoscopic X-ray technologies for patient setup in image guided stereotactic radiotherapy.

Authors:  Barbara Tagaste; Marco Riboldi; Maria F Spadea; Simone Bellante; Guido Baroni; Raffaella Cambria; Cristina Garibaldi; Mario Ciocca; Gianpiero Catalano; Daniela Alterio; Roberto Orecchia
Journal:  Int J Radiat Oncol Biol Phys       Date:  2011-05-24       Impact factor: 7.038

7.  The dosimetric impact of prostate rotations during electromagnetically guided external-beam radiation therapy.

Authors:  Hanan Amro; Daniel A Hamstra; Daniel L Mcshan; Howard Sandler; Karen Vineberg; Scott Hadley; Dale Litzenberg
Journal:  Int J Radiat Oncol Biol Phys       Date:  2012-05-02       Impact factor: 7.038

8.  Development of a frameless stereotactic radiosurgery system based on real-time 6D position monitoring and adaptive head motion compensation.

Authors:  Rodney D Wiersma; Zhifei Wen; Meredith Sadinski; Karl Farrey; Kamil M Yenice
Journal:  Phys Med Biol       Date:  2009-12-17       Impact factor: 3.609

9.  Accuracy of a commercial optical 3D surface imaging system for realignment of patients for radiotherapy of the thorax.

Authors:  Philipp J Schöffel; Wolfgang Harms; Gabriele Sroka-Perez; Wolfgang Schlegel; Christian P Karger
Journal:  Phys Med Biol       Date:  2007-06-06       Impact factor: 3.609

10.  Prospective real-time correction for arbitrary head motion using active markers.

Authors:  Melvyn B Ooi; Sascha Krueger; William J Thomas; Srirama V Swaminathan; Truman R Brown
Journal:  Magn Reson Med       Date:  2009-10       Impact factor: 4.668

View more
  7 in total

1.  Robotic real-time translational and rotational head motion correction during frameless stereotactic radiosurgery.

Authors:  Xinmin Liu; Andrew H Belcher; Zachary Grelewicz; Rodney D Wiersma
Journal:  Med Phys       Date:  2015-06       Impact factor: 4.071

2.  Towards frameless maskless SRS through real-time 6DoF robotic motion compensation.

Authors:  Andrew H Belcher; Xinmin Liu; Steven Chmura; Kamil Yenice; Rodney D Wiersma
Journal:  Phys Med Biol       Date:  2017-11-13       Impact factor: 3.609

3.  Spatial and rotational quality assurance of 6DOF patient tracking systems.

Authors:  Andrew H Belcher; Xinmin Liu; Zachary Grelewicz; Rodney D Wiersma
Journal:  Med Phys       Date:  2016-06       Impact factor: 4.071

4.  Technical Note: High temporal resolution characterization of gating response time.

Authors:  Rodney D Wiersma; Bradley P McCabe; Andrew H Belcher; Patrick J Jensen; Brett Smith; Bulent Aydogan
Journal:  Med Phys       Date:  2016-06       Impact factor: 4.071

5.  Development of a four-axis moving phantom for patient-specific QA of surrogate signal-based tracking IMRT.

Authors:  Nobutaka Mukumoto; Mitsuhiro Nakamura; Masahiro Yamada; Kunio Takahashi; Mami Akimoto; Yuki Miyabe; Kenji Yokota; Shuji Kaneko; Akira Nakamura; Satoshi Itasaka; Yukinori Matsuo; Takashi Mizowaki; Masaki Kokubo; Masahiro Hiraoka
Journal:  Med Phys       Date:  2016-12       Impact factor: 4.071

6.  Optimization based trajectory planning for real-time 6DoF robotic patient motion compensation systems.

Authors:  Xinmin Liu; Rodney D Wiersma
Journal:  PLoS One       Date:  2019-01-11       Impact factor: 3.240

7.  A novel dynamic robotic moving phantom system for patient-specific quality assurance in real-time tumor-tracking radiotherapy.

Authors:  Takehiro Shiinoki; Fumitake Fujii; Koya Fujimoto; Yuki Yuasa; Tatsuhiro Sera
Journal:  J Appl Clin Med Phys       Date:  2020-04-13       Impact factor: 2.102

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.