| Literature DB >> 25469097 |
Eun Ju Ha1, Jung Hwan Baek2, Jeong Hyun Lee2.
Abstract
OBJECTIVE: To compare the ablation characteristics of the moving-shot technique (MST) and the fixed electrode technique (FET) for radiofrequency (RF) ablation in an ex-vivo bovine liver tissue model.Entities:
Keywords: Fixed electrode technique; Intervention; Moving shot technique; Radiofrequency ablation; Thyroid nodule
Mesh:
Year: 2014 PMID: 25469097 PMCID: PMC4248641 DOI: 10.3348/kjr.2014.15.6.836
Source DB: PubMed Journal: Korean J Radiol ISSN: 1229-6929 Impact factor: 3.500
Fig. 1Measurement of ablation zone.
A. On longitudinal plane, we measured vertical diameter (DV) along electrode. B. On transverse plane, we measured transverse diameter (DT1) perpendicular to DV in L-plane and another transverse diameter (DT2) in T-plane.
Variables Before, During, and After Ablation According to Serial Ablation Time Using Fixed Electrode Technique
Note.- *Impedance before ablation, †Impedance during ablation (mean ± standard deviation). DT1 = transverse diameter, DT2 = another transverse diameter, DV = vertical diameter
Results of Post-Hoc Test for Variables during Ablation According to Serial Ablation Time
Note.- *Impedance during ablation, p < 0.005. DT1 = transverse diameter, DT2 = another transverse diameter, DV = vertical diameter
Fig. 2Relationship between ablation volume and time using fixed electrode technique. Ablation volume gradually increased with increasing ablation time; however, pair-wise statistical comparison between 2 neighboring ablation time lengths was not significant after 30 seconds.
Comparison of Variables of Moving Shot and Fixed Electrode Technique
Note.- *Total ablation time-cut-off time. DT1 = transverse diameter, DT2 = another transverse diameter, DV = vertical diameter
Fig. 3Graphs of radiofrequency (RF) power, current, impedance, and temperature during 2-minute ablation using fixed electrode and moving-shot techniques.
A. With fixed electrode technique, power graphed over time demonstrates that first RF cut-off appears at 33 seconds (arrow) and 6 RF cut-offs times (arrows) were detected during 2-minute ablation procedure. Drops of RF current and sudden increases of impedance were detected in corresponding area. These RF cut-offs denote cooling period that serves to prevent tissue carbonization near electrode tip. Total energy deposition was 0.68 kcal. B. With moving shot technique, RF power is maintained at approximately 50 watts during entire ablation procedure (arrows). Moving electrode tip could prevent RF cut-off, and total energy deposition was 1.32 kcal.